CLASS II ONGOING Device recall · Reported 8 Oct 2025 · FDA Enforcement Report

Venclose digiRF Generators recalled by Bard Peripheral Vascular

Bard Peripheral Vascular is recalling Venclose digiRF Generators, distributed nationwide in the US. The recall was initiated on 21 Aug 2025 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesUDI-DI: 00858254006022. Generator Software Version 3.35
Quantity recalled1,725

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-2638-2025
FDA event ID
97312
Recalling firm
Bard Peripheral Vascular, Tempe, AZ
Classification
Class II
Status
Ongoing
Recall initiated
21 Aug 2025
FDA classified
26 Sep 2025
Reported
8 Oct 2025
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Temperature Abuse

Timeline

21 Aug 2025Recall initiated by company
26 Sep 2025Classified by FDA+36 days
8 Oct 2025Published in enforcement report+12 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

Software version 3.35 of the Venclose digiRF Generator incorporates a catheter verification feature specifically designed to detect internal wiring anomalies in Venclose EVSRF Ablation Catheters prior to clinical use. This automated diagnostic check is executed immediately upon catheter connection to the generator, occurring before the procedure interface becomes accessible. When a catheter fails this verification process, the generator displays a Red X indicator on the screen without accompanying error codes, effectively disabling catheter functionality. However, BD has determined that software version 3.35 generates false positive failures due to temperature-dependent verification parameters, incorrectly flagging properly functioning Venclose EVSRF Ablation Catheters as defective when the catheter temperature is not at steady state during the check initiation.

Foods that need refrigeration and were stored or shipped too warm can grow harmful bacteria without any change in look or smell.

Product as listed by the FDA

Venclose digiRF Generators, Model VCRFG1, with software 3.35 designed to identify internal wiring anomalies in the Venclose EVSRF Ablation Catheter prior to use. The Venclose digiRF Generator is a sophisticated multi-voltage energy delivery system featuring touchscreen control that automatically configures non-adjustable treatment parameters (including time and temperature) for compatible catheters connected via a triaxial connector port. Designed for use with Venclose RF Catheters (either the EVSRF or Maven variants) as an integrated system, the device employs resistive radiofrequency ablation to deliver temperature-controlled RF energy that heats incompetent vein walls, inducing irreversible luminal occlusion followed by progressive fibrosis and eventual vein resorption. The 510(k) submission K250068 encompasses solely the Venclose digiRF Generator due to software modifications with no alterations to the cleared Venclose Catheters (EVSRF and Maven variants, cleared under K160754 and K211806, respectively).

Where it was distributed

Worldwide - US Nationwide distribution including in the states of NJ, NY, FL, LA, VA, CA, TX, NV, PA, NE, TN, IL, AZ, SC, AK, GA, CO, HI, OH, NC, WI, WV, MD, MO, AL, OK, MS, PR, KS, UT, CT, IA, MI, IN, MA, KY, WA, MT, MN, SD, ME, DE, NM, AR and the country of Malaysia.

Nationwide AlaskaAlabamaArkansasArizonaCaliforniaColoradoConnecticutDelawareFloridaGeorgiaHawaiiIowaIllinoisIndianaKansasKentuckyLouisianaMassachusettsMarylandMaine

Questions about this recall

Why was it recalled?

Software version 3.35 of the Venclose digiRF Generator incorporates a catheter verification feature specifically designed to detect internal wiring anomalies in Venclose EVSRF Ablation Catheters prior to clinical use. This automated diagnostic check is executed immediately upon catheter connection to the generator, occurring before the procedure interface becomes accessible. When a catheter fails this verification process, the generator displays a Red X indicator on the screen without accompanying error codes, effectively disabling catheter functionality. However, BD has determined that software version 3.35 generates false positive failures due to temperature-dependent verification parameters, incorrectly flagging properly functioning Venclose EVSRF Ablation Catheters as defective when the catheter temperature is not at steady state during the check initiation.

Which lots are affected?

UDI-DI: 00858254006022. Generator Software Version 3.35

Where was it sold?

Worldwide - US Nationwide distribution including in the states of NJ, NY, FL, LA, VA, CA, TX, NV, PA, NE, TN, IL, AZ, SC, AK, GA, CO, HI, OH, NC, WI, WV, MD, MO, AL, OK, MS, PR, KS, UT, CT, IA, MI, IN, MA, KY, WA, MT, MN, SD, ME, DE, NM, AR and the country of Malaysia.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.