CLASS II ONGOING Device recall · Reported 27 Dec 2023 · FDA Enforcement Report

Venous probe connection cable recalled by Maquet Medical Systems U

Maquet Medical Systems U is recalling Venous probe connection cable (, distributed nationwide in the US. The recall was initiated on 13 Nov 2023 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesUDI-DI 04058863074863 and 04037691658384 All serial numbers up to and including 90413927
Quantity recalled1213 US, 2589 OUS

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-0605-2024
FDA event ID
93563
Recalling firm
Maquet Medical Systems U, Wayne, NJ
Classification
Class II
Status
Ongoing
Recall initiated
13 Nov 2023
FDA classified
21 Dec 2023
Reported
27 Dec 2023
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Temperature Abuse

Timeline

13 Nov 2023Recall initiated by company
21 Dec 2023Classified by FDA+38 days
27 Dec 2023Published in enforcement report+6 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

Firm received complaints of breaks in the insulation, shielding, or wires of the venous probe connection cable. Fracture of the venous probe cable may influence the management of extracorporeal support. Patient may be exposed to inappropriate blood flow, blood temperature, or hematocrit/hemoglobin levels as a result of the CARDIOHELP-i not receiving information from the venous probe.

Foods that need refrigeration and were stored or shipped too warm can grow harmful bacteria without any change in look or smell.

Product as listed by the FDA

Venous probe connection cable (REF 701069333 and 701048804) used with the Cardiohelp-i System, part numbers 701072780 and 701041802

Where it was distributed

Domestic distribution Nationwide. Foreign distribution to Argentina, Australia, Austria, Azerbaijan, Bahrain, Belgium, Canada, Cayman Islands, China, Colombia, Croatia, Czech Republic, Denmark, Egypt, Estonia, Finland, France, French Polynesia, Germany, Greece, Hong Kong, Hungary, Iceland, India, Indonesia, Iran, Ireland, Israel, Islamic Republic Of Iraq, Italy, Japan, Kazakhstan, Korea, Republic Of Kuwait, Latvia, Lebanon, Luxembourg, Macao, Malaysia, Maldives, Martinique, Mauritius, Mayotte, Mexico, Netherlands, New Zealand, Norway, Oman, Pakistan, Philippines, Poland, Portugal, Puerto Rico, Qatar, R¿union, Romania, Russian Federation, Saudi Arabia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Thailand, Taiwan, Turkey, Ukraine, United Arab Emirates, United Kingdom, Vietnam.

Nationwide Puerto Rico

Questions about this recall

Is Venous probe connection cable ( recalled?

Yes. Maquet Medical Systems U recalled Venous probe connection cable ( on 13 Nov 2023. The recall is Class II and its status is Ongoing. Recall number Z-0605-2024.

Why was it recalled?

Firm received complaints of breaks in the insulation, shielding, or wires of the venous probe connection cable. Fracture of the venous probe cable may influence the management of extracorporeal support. Patient may be exposed to inappropriate blood flow, blood temperature, or hematocrit/hemoglobin levels as a result of the CARDIOHELP-i not receiving information from the venous probe.

Which lots are affected?

UDI-DI 04058863074863 and 04037691658384 All serial numbers up to and including 90413927

Where was it sold?

Domestic distribution Nationwide. Foreign distribution to Argentina, Australia, Austria, Azerbaijan, Bahrain, Belgium, Canada, Cayman Islands, China, Colombia, Croatia, Czech Republic, Denmark, Egypt, Estonia, Finland, France, French Polynesia, Germany, Greece, Hong Kong, Hungary, Iceland, India, Indonesia, Iran, Ireland, Israel, Islamic Republic Of Iraq, Italy, Japan, Kazakhstan, Korea, Republic Of Kuwait, Latvia, Lebanon, Luxembourg, Macao, Malaysia, Maldives, Martinique, Mauritius, Mayotte, Mexico, Netherlands, New Zealand, Norway, Oman, Pakistan, Philippines, Poland, Portugal, Puerto Rico, Qatar, R¿union, Romania, Russian Federation, Saudi Arabia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Thailand, Taiwan, Turkey, Ukraine, United Arab Emirates, United Kingdom, Vietnam.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 21 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

More recalls from Maquet Medical Systems U

All 41

Other Other Medical Devices recalls

All 13,771