CLASS I ONGOING Device recall · Reported 26 Jun 2024 · FDA Enforcement Report
Ventilator Hamilton-C6, PN recalled by Hamilton Medical AG
Hamilton Medical is recalling Ventilator Hamilton-C6, PN: 160021, distributed nationwide in the US. The recall was initiated on 15 May 2024 and the FDA has classified it as Class I.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-2020-2024
- FDA event ID
- 94671
- Recalling firm
- Hamilton Medical, Bonaduz, N/A
- Classification
- Class I
- Status
- Ongoing
- Recall initiated
- 15 May 2024
- FDA classified
- 18 Jun 2024
- Reported
- 26 Jun 2024
- Type
- Voluntary: Firm initiated
- Customers notified by
- Letter
- Reason category
- Device Malfunction
- Device type
- CPAP & Ventilators
Timeline
| 15 May 2024 | Recall initiated by company | |
| 18 Jun 2024 | Classified by FDA | +34 days |
| 26 Jun 2024 | Published in enforcement report | +8 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
Ventilator may enter sensor fail mode, ventilation may not re-initiate, after patient is reconnected, which may cause hypoxia, if this sequence occurs: 1)User presses O2 enrichment, 2)User disconnects endotracheal tube for open suctioning, 3)Sensor error initiated, sensor fail mode occurs, 4)Patient is re-connected with sensor fail mode active, 5) Ventilation is not re-initiated by the ventilator.
Device recalls are issued when a component, software or design fault could cause the device to fail or give wrong readings. Many are corrections applied in the field rather than physical returns.
Product as listed by the FDA
Ventilator HAMILTON-C6, PN: 160021
Where it was distributed
US Nationwide distribution in the states of GA, KY, FL, IN, NV, MI, VA, OR, TX, SC, CA, PA.
Nationwide CaliforniaFloridaGeorgiaIndianaKentuckyMichiganNevadaOregonPennsylvaniaSouth CarolinaTexasVirginia
Questions about this recall
Why was it recalled?
Ventilator may enter sensor fail mode, ventilation may not re-initiate, after patient is reconnected, which may cause hypoxia, if this sequence occurs: 1)User presses O2 enrichment, 2)User disconnects endotracheal tube for open suctioning, 3)Sensor error initiated, sensor fail mode occurs, 4)Patient is re-connected with sensor fail mode active, 5) Ventilation is not re-initiated by the ventilator.
Which lots are affected?
UDI: 07630002808590, Software Versions: SW v1.1.4, SW v1.1.5, SW v1.1.6
Where was it sold?
US Nationwide distribution in the states of GA, KY, FL, IN, NV, MI, VA, OR, TX, SC, CA, PA.
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class I mean?
Class I is the most serious FDA recall category: there is a reasonable probability that using the product will cause serious health consequences or death.
Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
More recalls from Hamilton Medical
All 34| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | IntelliCuff, Portable automatic cuff pressure controller recalled | Hamilton Medical | Packaging Defects | Nationwide |
| Device | CLASS I | Breathing circuit set, coaxial recalled by Hamilton Medical AG | Hamilton Medical | Device Malfunction | Nationwide |
| Device | CLASS I | Hamilton C6 recalled by Hamilton Medical AG | Hamilton Medical | Device Malfunction | Nationwide |
| Device | CLASS I | Breathing circuit set, coaxial, 1.80 m, with flow sensor recalled | Hamilton Medical | Device Malfunction | Nationwide |
| Device | CLASS I | Hamilton-T1, REF: 161009, Intensive Care and Transport Ventilator recalled | Hamilton Medical | Packaging Defects | Nationwide |
Other CPAP & Ventilators recalls
All 568| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS I | Vocsn+Pro recalled by Ventec Life Systems | Ventec Life Systems | Packaging Defects | Nationwide |
| Device | CLASS I | VC+Pro recalled by Ventec Life Systems | Ventec Life Systems | Packaging Defects | Nationwide |
| Device | CLASS I | V*Home recalled by Ventec Life Systems | Ventec Life Systems | Packaging Defects | Nationwide |
| Device | CLASS I | V+Pro recalled by Ventec Life Systems | Ventec Life Systems | Packaging Defects | Nationwide |
| Device | CLASS I | VC Model Number: Prt-01098-000, Package PN:Prt-01125-000. used recalled | Ventec Life Systems | Packaging Defects | Nationwide |