CLASS II ONGOING Device recall · Reported 11 Jan 2023 · FDA Enforcement Report

Verify Steam Test Pack (20 Tests and 20 controls) recalled by Steris

Steris is recalling Verify Steam Test Pack (20 Tests and 20 controls), distributed nationwide in the US. The recall was initiated on 12 Dec 2022 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesUDI: 10724995023864 lot #230613
Quantity recalled105 boxes

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-0781-2023
FDA event ID
91315
Recalling firm
Steris, Mentor, OH
Classification
Class II
Status
Ongoing
Recall initiated
12 Dec 2022
FDA classified
3 Jan 2023
Reported
11 Jan 2023
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Sterility
Device type
Patient Monitors

Timeline

12 Dec 2022Recall initiated by company
3 Jan 2023Classified by FDA+22 days
11 Jan 2023Published in enforcement report+8 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

Biological indicator (BI) used to promote growth of the G. stearothermophilus and B. atrophaeus spores was inconsistent in promoting growth throughout lot , some BIs within the lot performed as intended while others did not. When used to monitor steam and ethylene oxide sterilization cycles, the potential hazard is that the BI will show no growth indicating that the cycle was effective, even if it may not have been.

A sterility recall means a product that must be sterile, such as an injection or eye drop, may be contaminated. Using a non-sterile product can cause serious infection.

Product as listed by the FDA

VERIFY Steam Test Pack (20 Tests and 20 controls) Item Number: S3069 Intended for use in installation testing and routine monitoring of steam and ethylene oxide sterilization processes

Where it was distributed

Nationwide Foreign: AE AT CA CO ES FI GU IN IT JP MX SG TH TW ZA

Nationwide CaliforniaColorado

Questions about this recall

Why was it recalled?

Biological indicator (BI) used to promote growth of the G. stearothermophilus and B. atrophaeus spores was inconsistent in promoting growth throughout lot , some BIs within the lot performed as intended while others did not. When used to monitor steam and ethylene oxide sterilization cycles, the potential hazard is that the BI will show no growth indicating that the cycle was effective, even if it may not have been.

Which lots are affected?

UDI: 10724995023864 lot #230613

Where was it sold?

Nationwide Foreign: AE AT CA CO ES FI GU IN IT JP MX SG TH TW ZA

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

Other products in this recall event

3 other products · FDA event 91315

More recalls from Steris

All 120

Other Patient Monitors recalls

All 1,022