CLASS II ONGOING Device recall · Reported 21 Oct 2026 · FDA Enforcement Report

Rapicide PA Part A High-Level Disinfectant recalled by Steris

Steris is recalling Rapicide PA Part A High-Level Disinfectant, distributed nationwide in the US. The recall was initiated on 21 Sep 2026 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesUDI: 00724995245962 Lot numbers with expiration dates: 688019 10/19/2027 688021 10/19/2027 688023 10/19/2027 688025 10/19/2027 688576 10/26/2027 688580 10/26/2027 688582 10/26/2027 688584 10/26/2027 688586 10/26/2027 688588 10/26/2027 689173 11/2/2027 689175 11/2/2027 689177 11/2/2027 689179 11/2/2027 689983 11/10/2027 689985 11/10/2027 689987 11/10/2027 690299 11/16/2027 690301 11/16/2027 690995 11/24/2027 690997 11/24/2027 690999 11/24/2027 691001 11/24/2027 691741 12/1/2027 691743 12/1/2027 691745 12/8/2027 692203 12/8/2027 692206 12/8/2027 692208 12/8/2027 692210 12/8/2027 692213 12/8/2027 692215 12/8/2027 692217 12/9/2027 692572 12/10/2027 692574 12/10/2027 692576 12/10/2027 692578 12/10/2027 692580 12/10/2027 692582 12/10/2027 692730 12/14/2027 692732 12/14/2027 692734 12/14/2027 692736 12/14/2027 692738 12/14/2027 692740 12/14/2027 692742 12/14/2027 692744 12/14/2027 694359 12/30/2027 694361 12/30/2027 694363 12/30/2027
Quantity recalled106758 units

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-0156-2027
FDA event ID
99804
Recalling firm
Steris, Mentor, OH
Classification
Class II
Status
Ongoing
Recall initiated
21 Sep 2026
FDA classified
14 Oct 2026
Reported
21 Oct 2026
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Other Reasons

Timeline

21 Sep 2026Recall initiated by company
14 Oct 2026Classified by FDA+23 days
21 Oct 2026Published in enforcement report+7 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

If the vented cap becomes blocked, gas may accumulate and the bottle may swell due to increased internal pressure. In some cases, excess vapor or liquid chemistry may escape from the bottle when the cap is twisted during opening, which could result in user exposure if required personal protective equipment (PPE) is not properly used. Such exposure may result in injury to the user or others in close proximity.

Product as listed by the FDA

RAPICIDE PA Part A High-Level Disinfectant, Model/Catalog Number: ML020117

Where it was distributed

Worldwide - US Nationwide distribution.

Nationwide

Questions about this recall

Why was it recalled?

If the vented cap becomes blocked, gas may accumulate and the bottle may swell due to increased internal pressure. In some cases, excess vapor or liquid chemistry may escape from the bottle when the cap is twisted during opening, which could result in user exposure if required personal protective equipment (PPE) is not properly used. Such exposure may result in injury to the user or others in close proximity.

Which lots are affected?

UDI: 00724995245962 Lot numbers with expiration dates: 688019 10/19/2027 688021 10/19/2027 688023 10/19/2027 688025 10/19/2027 688576 10/26/2027 688580 10/26/2027 688582 10/26/2027 688584 10/26/2027 688586 10/26/2027 688588 10/26/2027 689173 11/2/2027 689175 11/2/2027 689177 11/2/2027 689179 11/2/2027 689983 11/10/2027 689985 11/10/2027 689987 11/10/2027 690299 11/16/2027 690301 11/16/2027 690995 11/24/2027 690997 11/24/2027 690999 11/24/2027 691001 11/24/2027 691741 12/1/2027 691743 12/1/2027 691745 12/8/2027 692203 12/8/2027 692206 12/8/2027 692208 12/8/2027 692210 12/8/2027 692213 12/8/2027 692215 12/8/2027 692217 12/9/2027 692572 12/10/2027 692574 12/10/2027 692576 12/10/2027 692578 12/10/2027 692580 12/10/2027 692582 12/10/2027 692730 12/14/2027 692732 12/14/2027 692734 12/14/2027 692736 12/14/2027 692738 12/14/2027 692740 12/14/2027 692742 12/14/2027 692744 12/14/2027 694359 12/30/2027 694361 12/30/2027 694363 12/30/2027

Where was it sold?

Worldwide - US Nationwide distribution.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 27 Oct 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

Other products in this recall event

2 other products · FDA event 99804

This product is part of a recall event; the main page for the event is Rapicide PA Part A High-Level Disinfectant recalled by Steris.

ReportedClassRecallCompanyReasonWhere
CLASS IIRapicide PA Part A High-Level Disinfectant recalled by SterisSterisOtherNationwide
CLASS IIRapicide PA Part A High-Level Disinfectant recalled by SterisSterisOtherNationwide

More recalls from Steris

All 123

Other Other Medical Devices recalls

All 13,929