CLASS II Device recall · Reported 21 Mar 2018 · FDA Enforcement Report

Versys Beaded FC 14X160MM Std Body Ext & Std Neck recalled

Zimmer Biomet is recalling Versys Beaded FC 14X160MM Std Body Ext & Std Neck Prosthesis, Hip, Semi-Constrained, distributed nationwide in the US. The recall was initiated on 29 Nov 2017 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesItem Number: 00-7843-014-26 Lot Number: 61741806 Item Number: 00-7843-014-06 Lot Numbers: 61886028 61752736 62120676
Quantity recalled13,227 in total

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-1024-2018
FDA event ID
78706
Recalling firm
Zimmer Biomet, Warsaw, IN
Classification
Class II
Status
Terminated
Recall initiated
29 Nov 2017
FDA classified
14 Mar 2018
Reported
21 Mar 2018
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Foreign Material

Timeline

29 Nov 2017Recall initiated by company
14 Mar 2018Classified by FDA+105 days
21 Mar 2018Published in enforcement report+7 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

The LDPE bag packaging for various highly polished hip and knee implants may adhere to the highly polished surface, leaving residue or material from the LDPE bag on the implant after it is removed from the bag.

Foreign material recalls cover pieces of metal, plastic, glass, wood or other objects that got into the product during processing and could cause injury if eaten.

Product as listed by the FDA

VERSYS 6 INCH BEADED FC 14X160MM STD BODY EXT & STD NECK Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Porous Uncemented

Where it was distributed

Worldwide Distribution: US (nationwide) and countries of: Argentina, Australia, Brazil, Bolivia, Canada, Chile, China, Colombia, Ecuador, El Salvador, India, Japan, Malaysia, Mexico, Netherlands, New Zealand, Nicaragua, Panama, Seoul-Korea, Singapore, Taiwan, Thailand, and UAE.

Nationwide

Questions about this recall

Is Versys Beaded FC 14X160MM Std Body Ext & Std Neck Prosthesis, Hip, Semi-Constrained recalled?

Yes. Zimmer Biomet recalled Versys Beaded FC 14X160MM Std Body Ext & Std Neck Prosthesis, Hip, Semi-Constrained on 29 Nov 2017. The recall is Class II and its status is Terminated. Recall number Z-1024-2018.

Why was it recalled?

The LDPE bag packaging for various highly polished hip and knee implants may adhere to the highly polished surface, leaving residue or material from the LDPE bag on the implant after it is removed from the bag.

Which lots are affected?

Item Number: 00-7843-014-26 Lot Number: 61741806 Item Number: 00-7843-014-06 Lot Numbers: 61886028 61752736 62120676

Where was it sold?

Worldwide Distribution: US (nationwide) and countries of: Argentina, Australia, Brazil, Bolivia, Canada, Chile, China, Colombia, Ecuador, El Salvador, India, Japan, Malaysia, Mexico, Netherlands, New Zealand, Nicaragua, Panama, Seoul-Korea, Singapore, Taiwan, Thailand, and UAE.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 21 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

More recalls from Zimmer Biomet

All 873

Other Hip & Knee Implants recalls

All 3,097