CLASS II ONGOING Device recall · Reported 6 Aug 2025 · FDA Enforcement Report

Visions Digital IVUS Catheter: PV.014P (Platinum) recalled by Volcano

Volcano is recalling Visions Digital IVUS Catheter: PV.014P (Platinum), distributed nationwide in the US. The recall was initiated on 13 Jun 2025 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesUDI-DI: 00845225002848, 00184360000037. Instructions for use: 300004875732_A Revision Date: 03/2021, 300004410212.A Release Date: 03/2021
Quantity recalled233,817

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-2221-2025
FDA event ID
97192
Recalling firm
Volcano, San Diego, CA
Classification
Class II
Status
Ongoing
Recall initiated
13 Jun 2025
FDA classified
30 Jul 2025
Reported
6 Aug 2025
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Other Reasons

Timeline

13 Jun 2025Recall initiated by company
30 Jul 2025Classified by FDA+47 days
6 Aug 2025Published in enforcement report+7 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

Reports of IVUS catheter and guide wire entanglement during radial-to-peripheral procedures, due to catheter use without appropriate sheath and/or guide catheter, which may require further intervention, such as surgical removal; Firm is restating instructions for use and adding: "Use a guide sheath of appropriate length to provide adequate support to the rapid exchange IVUS catheter and guidewire"

Product as listed by the FDA

Volcano Visions Digital IVUS Catheter: PV.014P (Platinum), REF:85910P; PV.014P RX, REF: 014R; and PV.018, REF: 86700, 86700J used with Instructions for use.

Where it was distributed

Worldwide distribution - US Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY and the countries of BR, PL, SK, IT, DK, DE, GB, AT, KW, CR, ES, NL, LV, MT, IE, CH, ZA, RO, FI, PR, FR, CA, PA, IL, SE, PT, HR, GR, NZ, HK, TW, MY, TH, MX, AR, PE, AE, BG, AU, JP, ID, RS, CO, SG, BE.

Nationwide AlaskaAlabamaArkansasArizonaCaliforniaColoradoConnecticutDistrict of ColumbiaDelawareFloridaGeorgiaHawaiiIowaIdahoIllinoisIndianaKansasKentuckyLouisianaMassachusetts

Questions about this recall

Why was it recalled?

Reports of IVUS catheter and guide wire entanglement during radial-to-peripheral procedures, due to catheter use without appropriate sheath and/or guide catheter, which may require further intervention, such as surgical removal; Firm is restating instructions for use and adding: "Use a guide sheath of appropriate length to provide adequate support to the rapid exchange IVUS catheter and guidewire"

Which lots are affected?

UDI-DI: 00845225002848, 00184360000037. Instructions for use: 300004875732_A Revision Date: 03/2021, 300004410212.A Release Date: 03/2021

Where was it sold?

Worldwide distribution - US Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY and the countries of BR, PL, SK, IT, DK, DE, GB, AT, KW, CR, ES, NL, LV, MT, IE, CH, ZA, RO, FI, PR, FR, CA, PA, IL, SE, PT, HR, GR, NZ, HK, TW, MY, TH, MX, AR, PE, AE, BG, AU, JP, ID, RS, CO, SG, BE.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

More recalls from Volcano

All 32

Other Heart & Vascular Devices recalls

All 6,220