CLASS II ONGOING Device recall · Reported 24 Sep 2025 · FDA Enforcement Report

Vitros Chemistry Products LAC Slides recalled

Ortho-Clinical Diagnostics is recalling Vitros Chemistry Products LAC Slides, distributed nationwide in the US. The recall was initiated on 29 Jul 2025 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesCatalog number 8150112; UDI# 10758750004454; Coating 0130 and above, this includes Lot numbers 3533-0130-2947, 3533-0130-2972, 3533-0130-3001, 3533-0130-3002, 3533-0130-3048, 3533-0130-3084, 3533-0131-5887, 3533-0131-5958, 3533-0131-5962, 3533-0131-6058 ; Expiration date range 01-Oct-2026 to 01-Jan-2027
Quantity recalled2,320 units

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-2560-2025
FDA event ID
97409
Recalling firm
Ortho-Clinical Diagnostics, Rochester, NY
Classification
Class II
Status
Ongoing
Recall initiated
29 Jul 2025
FDA classified
12 Sep 2025
Reported
24 Sep 2025
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Device Malfunction

Timeline

29 Jul 2025Recall initiated by company
12 Sep 2025Classified by FDA+45 days
24 Sep 2025Published in enforcement report+12 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

VITROS Chemistry Products LAC Slides on VITROS XT 3400 and VITROS XT 7600 Systems are experiencing an increased occurrence of calibration failure related to lots from coating 0130 and above with Condition Code TH4-63J during calibration, causing the delay of lac test results and potentially impact patient management. The issue was identified by QuidelOrtho through an increase in customer complaints compared to the historical baseline with average failure rate of 1% for coating 0130 in 2025 while average of failure rate for other coating was 0.06% in 2025 and average of failure rate for all coatings is 0.03% for 2024.

Device recalls are issued when a component, software or design fault could cause the device to fail or give wrong readings. Many are corrections applied in the field rather than physical returns.

Product as listed by the FDA

VITROS Chemistry Products LAC Slides. Catalog Number: 8150112. 1 Sales Unit = contains 5 cartridges of 18 slides, for a total of 90 tests.

Where it was distributed

US Nationwide distribution.

Nationwide

Questions about this recall

Why was it recalled?

VITROS Chemistry Products LAC Slides on VITROS XT 3400 and VITROS XT 7600 Systems are experiencing an increased occurrence of calibration failure related to lots from coating 0130 and above with Condition Code TH4-63J during calibration, causing the delay of lac test results and potentially impact patient management. The issue was identified by QuidelOrtho through an increase in customer complaints compared to the historical baseline with average failure rate of 1% for coating 0130 in 2025 while average of failure rate for other coating was 0.06% in 2025 and average of failure rate for all coatings is 0.03% for 2024.

Which lots are affected?

Catalog number 8150112; UDI# 10758750004454; Coating 0130 and above, this includes Lot numbers 3533-0130-2947, 3533-0130-2972, 3533-0130-3001, 3533-0130-3002, 3533-0130-3048, 3533-0130-3084, 3533-0131-5887, 3533-0131-5958, 3533-0131-5962, 3533-0131-6058 ; Expiration date range 01-Oct-2026 to 01-Jan-2027

Where was it sold?

US Nationwide distribution.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

Other products in this recall event

1 other product · FDA event 97409

More recalls from Ortho-Clinical Diagnostics

All 311

Other Other Medical Devices recalls

All 13,778