CLASS II Device recall · Reported 23 Sep 2015 · FDA Enforcement Report
Voyant Open Fusion Device recalled by Applied Medical Resources
Applied Medical Resources is recalling Voyant Open Fusion Device, distributed nationwide in the US. The recall was initiated on 31 Jul 2015 and the FDA has classified it as Class II.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-2769-2015
- FDA event ID
- 71956
- Recalling firm
- Applied Medical Resources, Rancho Santa Margarita, CA
- Classification
- Class II
- Status
- Terminated
- Recall initiated
- 31 Jul 2015
- FDA classified
- 15 Sep 2015
- Reported
- 23 Sep 2015
- Type
- Voluntary: Firm initiated
- Customers notified by
- Letter
- Reason category
- Packaging Defects
- Device type
- Surgical Instruments
Timeline
| 31 Jul 2015 | Recall initiated by company | |
| 15 Sep 2015 | Classified by FDA | +46 days |
| 23 Sep 2015 | Published in enforcement report | +8 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
Potential to improperly characterize tissue within the jaws, which may result in nonoptimal power delivery.
Packaging recalls include wrong labels, wrong products in the wrong bottle, missing leaflets, short-filled or empty capsules and seals that fail.
Product as listed by the FDA
Voyant Open Fusion Device Models: EB040 and EB040+ Product Usage: The Applied Voyant Open Fusion Device is designed for use with the Voyant Electrosurgical Generator. The device is an advanced bipolar instrument that uses RF energy, provided by the generator, to seal vessels up to and including 5mm (EB040+) or 7mm (EB040) in diameter. The device may also be used to seal tissue bundles that can be captured in the device jaws. The system delivers the energy required to fuse the tissue between the jaws of the device by monitoring voltage and current throughout the seal cycle. The device also features a useractuated blade for the transection of sealed tissue.
Where it was distributed
Worldwide Distribution - U.S Nationwide in the states of:FL, GA, NV. and countries of: France, Netherlands, Spain, and United Kingdom.
Questions about this recall
Why was it recalled?
Potential to improperly characterize tissue within the jaws, which may result in nonoptimal power delivery.
Which lots are affected?
Model Number: EB040 Lot Numbers: 1243964, 1243975, 1244125 Model Number: EB040+ Lot Numbers: 1245512, 1245513, 1246241, 1246508, 1247079, 1247566, 1248102, 1249264
Where was it sold?
Worldwide Distribution - U.S Nationwide in the states of:FL, GA, NV. and countries of: France, Netherlands, Spain, and United Kingdom.
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class II mean?
Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.
Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
More recalls from Applied Medical Resources
All 30| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Kii Low Profile, Bladed Dual Pack , Qty recalled | Applied Medical Resources | Packaging Defects | CA, ME, MI |
| Device | CLASS II | Epix Universal Clip Applier recalled by Applied Medical Resources | Applied Medical Resources | Device Malfunction | Nationwide |
| Device | CLASS II | Epix Universal Clip Applier recalled by Applied Medical Resources | Applied Medical Resources | Device Malfunction | Nationwide |
| Device | CLASS II | Epix latis Grasper recalled by Applied Medical Resources | Applied Medical Resources | Other | 43 states |
| Device | CLASS II | Epix latis Grasper recalled by Applied Medical Resources | Applied Medical Resources | Other | 43 states |
Other Surgical Instruments recalls
All 3,905| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Pentax Medical Ultrasound Endoscope recalled by Pentax of America | Pentax of America | Sterility | Nationwide |
| Device | CLASS II | Pentax Medical Ultrasound Endoscope recalled by Pentax of America | Pentax of America | Sterility | Nationwide |
| Device | CLASS II | Pentax Medical Ultrasound Endoscope recalled by Pentax of America | Pentax of America | Sterility | Nationwide |
| Device | CLASS II | Pentax Medical Ultrasound Endoscope recalled by Pentax of America | Pentax of America | Sterility | Nationwide |
| Device | CLASS II | Ethilon Nylon Suture recalled by Ethicon | Ethicon | Device Malfunction | CA |