CLASS III Device recall · Reported 11 Sep 2013 · FDA Enforcement Report

WASP (Walk Away Specimen Process) recalled by Copan Italia

Copan Italia is recalling WASP (Walk Away Specimen Process), distributed in California, Oregon. The recall was initiated on 9 Aug 2013 and the FDA has classified it as Class III.

Check your package

Lot, batch or date codesSerial Numbers: W086-61-084/106/115/119/134/136; W086-050-179; W086-61-087/109/125/127/130/133; W086-050-144/151/153/157/170/171/181/184; W086-61-088/097/100/117; W086-050-177/178/183; W086-61-093; W086-050-143/155; W086-61-096; W086-050-164/169/159/175; W086-61-103/137/141/142; W08661-104/114118/128/138; W086050-182; W08661 111; W08661 113; W08661 124; W08661129/140; W086050150/152/154/185; W08661 139; W086050 147; W08661 110; W086050160/161/162; W086050 180; W086050 187.
Quantity recalled63 units

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-2131-2013
FDA event ID
65923
Recalling firm
Copan Italia, Brescia, N/A
Classification
Class III
Status
Terminated
Recall initiated
9 Aug 2013
FDA classified
30 Aug 2013
Reported
11 Sep 2013
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Device Malfunction

Timeline

9 Aug 2013Recall initiated by company
30 Aug 2013Classified by FDA+21 days
11 Sep 2013Published in enforcement report+12 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

The recall of the WASP (Walk Away Specimen Process) instrument was initiated because Copan Italia SpA has upgraded the software concerning the broth inoculation table feature. The upgrade has been considered necessary from the manufacturer in order to eliminate the very low risk of rotation of the broth table when the door is open.

Device recalls are issued when a component, software or design fault could cause the device to fail or give wrong readings. Many are corrections applied in the field rather than physical returns.

Product as listed by the FDA

WASP (Walk Away Specimen Process). Automatic Planting and Streaking Instrument Catalog number: W086 WASP is a fully automated system used to process microbiological specimens in liquid phase for the investigation of infectious microbes.

Where it was distributed

Worldwide Distribution - USA including Oregon and California., and the countries of France, Belgium, Italy, Denmark, Norway, Switzerland, Romania, Finland, Germany, South Africa, Canada, Ireland, UK, Sweden and Japan

CaliforniaOregon

Questions about this recall

Why was it recalled?

The recall of the WASP (Walk Away Specimen Process) instrument was initiated because Copan Italia SpA has upgraded the software concerning the broth inoculation table feature. The upgrade has been considered necessary from the manufacturer in order to eliminate the very low risk of rotation of the broth table when the door is open.

Which lots are affected?

Serial Numbers: W086-61-084/106/115/119/134/136; W086-050-179; W086-61-087/109/125/127/130/133; W086-050-144/151/153/157/170/171/181/184; W086-61-088/097/100/117; W086-050-177/178/183; W086-61-093; W086-050-143/155; W086-61-096; W086-050-164/169/159/175; W086-61-103/137/141/142; W08661-104/114118/128/138; W086050-182; W08661 111; W08661 113; W08661 124; W08661129/140; W086050150/152/154/185; W08661 139; W086050 147; W08661 110; W086050160/161/162; W086050 180; W086050 187.

Where was it sold?

Worldwide Distribution - USA including Oregon and California., and the countries of France, Belgium, Italy, Denmark, Norway, Switzerland, Romania, Finland, Germany, South Africa, Canada, Ireland, UK, Sweden and Japan

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class III mean?

Class III recalls are for products unlikely to cause any adverse health consequence, usually labelling or quality issues.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

More recalls from Copan Italia

All 5

Other Other Medical Devices recalls

All 13,908