CLASS II Device recall · Reported 8 Aug 2018 · FDA Enforcement Report
Xltek EMU40EX EEG Headbox recalled by Natus Neurology
Natus Neurology is recalling Xltek EMU40EX EEG Headbox, distributed in 34 states. The recall was initiated on 15 May 2018 and the FDA has classified it as Class II.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-2570-2018
- FDA event ID
- 80239
- Recalling firm
- Natus Neurology, Oakville, N/A
- Classification
- Class II
- Status
- Terminated
- Recall initiated
- 15 May 2018
- FDA classified
- 28 Jul 2018
- Reported
- 8 Aug 2018
- Type
- Voluntary: Firm initiated
- Customers notified by
- Letter
- Reason category
- Foreign Material
- Device type
- Other Medical Devices
Timeline
| 15 May 2018 | Recall initiated by company | |
| 28 Jul 2018 | Classified by FDA | +74 days |
| 8 Aug 2018 | Published in enforcement report | +11 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
Possible degradation of an internal electronic component over time which may cause the XLTEK EMU 40EX Breakout box to fail. It may feel hot and plastic case may start melting.
Foreign material recalls cover pieces of metal, plastic, glass, wood or other objects that got into the product during processing and could cause injury if eaten.
Product as listed by the FDA
XLTEK EMU40EX EEG Headbox
Where it was distributed
USA Distribution: AL, CA, CO, CT, FL, GA, IA, ID, IL, IN, KY, LA, MA, MD, ME, MI MN, MO, NC, NE, NJ, NV, NY, OH, OR, PA, ,RI, SC, TN, TX, UT, VA, WA, WI. World wide Distribution: Canada, China, Cyprus, Germany, Hong Kong, Italy, Mexico, Saudi Arabia, Spain, United Kingdom
AlabamaCaliforniaColoradoConnecticutFloridaGeorgiaIowaIdahoIllinoisIndianaKentuckyLouisianaMassachusettsMarylandMaineMichiganMinnesotaMissouriNorth CarolinaNebraska
Questions about this recall
Why was it recalled?
Possible degradation of an internal electronic component over time which may cause the XLTEK EMU 40EX Breakout box to fail. It may feel hot and plastic case may start melting.
Which lots are affected?
Part Numbers: 10406, 002933
Where was it sold?
USA Distribution: AL, CA, CO, CT, FL, GA, IA, ID, IL, IN, KY, LA, MA, MD, ME, MI MN, MO, NC, NE, NJ, NV, NY, OH, OR, PA, ,RI, SC, TN, TX, UT, VA, WA, WI. World wide Distribution: Canada, China, Cyprus, Germany, Hong Kong, Italy, Mexico, Saudi Arabia, Spain, United Kingdom
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class II mean?
Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.
Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
More recalls from Natus Neurology
All 16| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Natus Brain Monitor Breakout recalled by Natus Neurology | Natus Neurology | Device Malfunction | 7 states |
| Device | CLASS II | OBM00002 OBM DAB (Digital Acquisition Box) recalled by Natus Neurology | Natus Neurology | Device Malfunction | Nationwide |
| Device | CLASS II | Sleep Supplies Start Kit with Single Use Supplies recalled | Natus Neurology | Other | Nationwide |
| Device | CLASS II | Tensive Conductive Adhesive Gel recalled by Natus Neurology | Natus Neurology | Other | Nationwide |
| Device | CLASS II | Ipela Fhd Ptz Tile Kits with Axis cameras used recalled | Natus Neurology | Device Malfunction | Nationwide |
Other Other Medical Devices recalls
All 13,778| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Citric Complete Dry Citric Acid Concentrate recalled | Nipro Renal Soultions USA | Other | Nationwide |
| Device | CLASS I | Kit: Epidural/Pain TQC 20/Cs recalled by Medical Action Industries | Medical Action Industries Inc. 306 | Sterility | 6 states |
| Device | CLASS II | Halyard IV Start Kit Non-Sterile recalled by Avid Medical | Avid Medical | Sterility | Nationwide |
| Device | CLASS I | Kit: Single Shot Epidural 20/Cs recalled by Medical Action Industries | Medical Action Industries Inc. 306 | Sterility | 6 states |
| Device | CLASS II | AMS Convenience Kit: Dsaek Pack - RX recalled | Windstone Medical Packaging | Device Malfunction | Nationwide |