CLASS I Device recall · Reported 21 May 2014 · FDA Enforcement Report

Abacus TPN Calculation recalled by Baxter Corporation Englewood

Baxter Corporation Englewood is recalling Abacus TPN Calculation, distributed nationwide in the US. The recall was initiated on 2 Apr 2014 and the FDA has classified it as Class I.

Check your package

Lot, batch or date codesProduct Code 8300-0045, 8300-0046, 8300-0047, 8300-0156, 8300-0157, 8300-0158, 8300-0166, 8300-0167, 8300-0168, 8300-0169
Quantity recalled2034

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-1566-2014
FDA event ID
67884
Recalling firm
Baxter Corporation Englewood, Englewood, CO
Classification
Class I
Status
Terminated
Recall initiated
2 Apr 2014
FDA classified
12 May 2014
Reported
21 May 2014
Type
Voluntary: Firm initiated
Customers notified by
E-Mail
Reason category
Sterility
Device type
Medical Software

Timeline

2 Apr 2014Recall initiated by company
12 May 2014Classified by FDA+40 days
21 May 2014Published in enforcement report+9 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

1. ABACUS v3.1 may calculate quantities of electrolytes that are double the expected values during the creation of TPN orders. 2. ABACUS v3.1 may automatically add additional sterile water to a formula equal to the volume of a premix resulting in an over dilution. 3. All software versions of ABACUS software display the calcium phosphate curve points for Premasol incorrectly. 4. All software v

A sterility recall means a product that must be sterile, such as an injection or eye drop, may be contaminated. Using a non-sterile product can cause serious infection.

Product as listed by the FDA

ABACUS TPN Calculation Software, Product Codes: 8300-0045 ABACUS CE (Calculator Edition), 8300-0046 ABACUS SE (Single-Workstation Edition), 8300-0047 ABACUS ME (Multi-Workstation Edition), 8300-0156 ABACUS 3.0 CE, 8300-0157 ABACUS 3.0 SE, 8300-0158 ABACUS 3.0 ME, 8300-0166 ABACUS 3.1 DE, 8300-0167 ABACUS 3.1 CE, 8300-0168 ABACUS 3.1 SE, 8300-0169 ABACUS 3.1 ME. ABACUS is a Windows - based order entry TPN software application for comprehensive calculations and label printing. The primary use of Abacus is for Total Parenteral Nutrition (TPN) order calculations.

Where it was distributed

Worldwide Distribution - USA nationwide, Canada, Colombia, Dominican Republic, Panama, Austria, Bosnia Herzegovina, Bulgaria, Bahrain, Estonia, France, Germany, Georgia, Hungary, Ireland, Israel, Jordan, Kuwait, Latvia, Morocco, Oman, Qatar, Saudi Arabia, Slovenia, Slovakia, South Africa, Switzerland, United Arab Emirates, United Kingdom, Hong Kong, and Taiwan.

Nationwide Georgia

Questions about this recall

Why was it recalled?

1. ABACUS v3.1 may calculate quantities of electrolytes that are double the expected values during the creation of TPN orders. 2. ABACUS v3.1 may automatically add additional sterile water to a formula equal to the volume of a premix resulting in an over dilution. 3. All software versions of ABACUS software display the calcium phosphate curve points for Premasol incorrectly. 4. All software v

Which lots are affected?

Product Code 8300-0045, 8300-0046, 8300-0047, 8300-0156, 8300-0157, 8300-0158, 8300-0166, 8300-0167, 8300-0168, 8300-0169

Where was it sold?

Worldwide Distribution - USA nationwide, Canada, Colombia, Dominican Republic, Panama, Austria, Bosnia Herzegovina, Bulgaria, Bahrain, Estonia, France, Germany, Georgia, Hungary, Ireland, Israel, Jordan, Kuwait, Latvia, Morocco, Oman, Qatar, Saudi Arabia, Slovenia, Slovakia, South Africa, Switzerland, United Arab Emirates, United Kingdom, Hong Kong, and Taiwan.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class I mean?

Class I is the most serious FDA recall category: there is a reasonable probability that using the product will cause serious health consequences or death.