CLASS II ONGOING Device recall · Reported 12 Apr 2023 · FDA Enforcement Report

Abbott Alinity m System recalled by Abbott Molecular

Abbott Molecular is recalling Abbott Alinity m System, distributed in Pennsylvania. The recall was initiated on 11 Nov 2022 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesAlinity m System Serial #01015, UDI: (01)00884999048034(11)220505(21)01015(240)08N53-002; ADU Serial #AMP05139.
Quantity recalled1

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-1349-2023
FDA event ID
91777
Recalling firm
Abbott Molecular, Des Plaines, IL
Classification
Class II
Status
Ongoing
Recall initiated
11 Nov 2022
FDA classified
6 Apr 2023
Reported
12 Apr 2023
Type
Voluntary: Firm initiated
Customers notified by
Visit
Reason category
Other Reasons

Timeline

11 Nov 2022Recall initiated by company
6 Apr 2023Classified by FDA+146 days
12 Apr 2023Published in enforcement report+6 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

The instrument was found to be installed without plunger assemblies in the clamp bar for an Amplification Detection Unit. After correction of the issue, the instrument was returned for use without calibration. This would result in a potential for delay of results and incorrect results.

Product as listed by the FDA

Abbott Alinity m System, REF 08N53-002, containing Amplification Detetction Unit (ADU) #4.

Where it was distributed

US Distribution to state of: PA, a government account.

Pennsylvania

Questions about this recall

Why was it recalled?

The instrument was found to be installed without plunger assemblies in the clamp bar for an Amplification Detection Unit. After correction of the issue, the instrument was returned for use without calibration. This would result in a potential for delay of results and incorrect results.

Which lots are affected?

Alinity m System Serial #01015, UDI: (01)00884999048034(11)220505(21)01015(240)08N53-002; ADU Serial #AMP05139.

Where was it sold?

US Distribution to state of: PA, a government account.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

More recalls from Abbott Molecular

All 36

Other Other Medical Devices recalls

All 13,908