CLASS II ONGOING Device recall · Reported 30 Apr 2025 · FDA Enforcement Report
The Alinity m System is designed to automate the steps recalled
Abbott Molecular is recalling The Alinity m System is designed to automate the steps for nucleic acid testing, distributed nationwide in the US. The recall was initiated on 20 Mar 2025 and the FDA has classified it as Class II.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-1665-2025
- FDA event ID
- 96559
- Recalling firm
- Abbott Molecular, Des Plaines, IL
- Classification
- Class II
- Status
- Ongoing
- Recall initiated
- 20 Mar 2025
- FDA classified
- 24 Apr 2025
- Reported
- 30 Apr 2025
- Type
- Voluntary: Firm initiated
- Customers notified by
- Letter
- Reason category
- Other Reasons
- Device type
- Other Medical Devices
Timeline
| 20 Mar 2025 | Recall initiated by company | |
| 24 Apr 2025 | Classified by FDA | +35 days |
| 30 Apr 2025 | Published in enforcement report | +6 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
The design of the Alinity m System Solutions drawer enclosure may allow liquid (liquid waste or system solutions) originating in the System Solutions drawer to flow beyond the footprint of the instrument and into the walking-path of the user.
Product as listed by the FDA
The Alinity m System is designed to automate the steps for nucleic acid testing. Model Numbers included in this recall are 08N53-001, 08N53-002, and 08N53-032 (Refurbished).
Where it was distributed
Worldwide - US Nationwide distribution.
Questions about this recall
Why was it recalled?
The design of the Alinity m System Solutions drawer enclosure may allow liquid (liquid waste or system solutions) originating in the System Solutions drawer to flow beyond the footprint of the instrument and into the walking-path of the user.
Which lots are affected?
All Serial Numbers. 08N53-001, UDI-DI: 00884999047389; 08N53-002, UDI-DI: 00884999048034; 08N53-032, UDI-DI: 00884999047587
Where was it sold?
Worldwide - US Nationwide distribution.
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class II mean?
Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.
Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
More recalls from Abbott Molecular
All 36| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Alinity m Resp-4-Plex Amplification (AMP) Kit (CE) recalled | Abbott Molecular | Other | Nationwide |
| Device | CLASS II | Alinity m Resp-4-Plex AMP Kit, List Number 09N79 recalled | Abbott Molecular | Other | Nationwide |
| Device | CLASS II | Abbott Alinity m STI AMP Kit, used with the Alinity m System recalled | Abbott Molecular | Potency & Assay | Nationwide |
| Device | CLASS II | Abbott Alinity m HPV AMP Kit, used with the Alinity m System recalled | Abbott Molecular | Potency & Assay | Nationwide |
| Device | CLASS II | Alinity m System recalled by Abbott Molecular | Abbott Molecular | Other | Nationwide |
Other Other Medical Devices recalls
All 13,986| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | ANNOUNCED | Luna Bilevel Positive Airway Pressure (BPAP) System recalledLuna | React Health | Other | VT |
| Device | ANNOUNCED | Beds, Rails, Mattresses, and accessories recalled by MedlineMedline | Medline | Other | |
| Device | CLASS II | Hemochron Jr., Whole Blood Microcoagulation System recalled | Accriva Diagnostics | Other | Nationwide |
| Device | CLASS II | Senographe Pristina 3D recalled by GE Medical Systems | GE Medical Systems | Other | WI |
| Device | CLASS II | Mobile column TruSystem recalled by Baxter Healthcare | Baxter Healthcare | Device Malfunction | Nationwide |