CLASS II Device recall · Reported 8 Jul 2020 · FDA Enforcement Report

ABL80-Flex Co-ox 393-841 Box Label recalled by Radiometer Medical ApS

Radiometer Medical ApS is recalling ABL80-Flex Co-ox 393-841 Box Label, distributed nationwide in the US. The recall was initiated on 11 May 2020 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesAll lots
Quantity recalled1981

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-2468-2020
FDA event ID
85634
Recalling firm
Radiometer Medical ApS, Bronshoj, N/A
Classification
Class II
Status
Terminated
Recall initiated
11 May 2020
FDA classified
29 Jun 2020
Reported
8 Jul 2020
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Other Reasons

Timeline

11 May 2020Recall initiated by company
29 Jun 2020Classified by FDA+49 days
8 Jul 2020Published in enforcement report+9 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

Barcode readers on ALB900 and ABL800 analyzers used for measuring pH, blood gases, electrolytes, glucose, lactate, hematocrit and oximetry may misinterpret the contents of some barcode types not using a check digit for patient ID or accession number. Misinterpretation of a barcode, including e.g. patient ID or accession number, may result in patient mix up and/or delayed medical treatment.

Product as listed by the FDA

ABL80-FLEX CO-OX 393-841 Box Label, REF 393-841, IVD, CE, UDI: 05700693938417 - Product Usage: It is intended for use in a laboratory environment, near patient or point-of-care setting.

Where it was distributed

Worldwide distribution - US Nationwide distribution including in the states of AL, AR, AK, AZ, CA, CO, CT, DC, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY and the country of CA.

Nationwide AlaskaAlabamaArkansasArizonaCaliforniaColoradoConnecticutDistrict of ColumbiaDelawareFloridaGeorgiaIowaIdahoIllinoisIndianaKansasKentuckyLouisianaMassachusettsMaryland

Questions about this recall

Is ABL80-Flex Co-ox 393-841 Box Label recalled?

Yes. Radiometer Medical ApS recalled ABL80-Flex Co-ox 393-841 Box Label on 11 May 2020. The recall is Class II and its status is Terminated. Recall number Z-2468-2020.

Why was it recalled?

Barcode readers on ALB900 and ABL800 analyzers used for measuring pH, blood gases, electrolytes, glucose, lactate, hematocrit and oximetry may misinterpret the contents of some barcode types not using a check digit for patient ID or accession number. Misinterpretation of a barcode, including e.g. patient ID or accession number, may result in patient mix up and/or delayed medical treatment.

Which lots are affected?

All lots

Where was it sold?

Worldwide distribution - US Nationwide distribution including in the states of AL, AR, AK, AZ, CA, CO, CT, DC, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY and the country of CA.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 21 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

More recalls from Radiometer Medical ApS

All 24

Other Other Medical Devices recalls

All 13,778