CLASS II ONGOING Device recall · Reported 17 Dec 2025 · FDA Enforcement Report

Access 2 Reaction Vessels, individual, disposable recalled

Beckman Coulter is recalling Access 2 Reaction Vessels, individual, disposable, plastic tubes used to process patient, distributed nationwide in the US. The recall was initiated on 7 Nov 2025 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesUDI-DI: 15099590180249; Lot Numbers: P243943, P243971, P244202 P244344, P244345, P250001, P250002, P250003, P250004 and P250005
Quantity recalled38,225 untis

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-0889-2026
FDA event ID
97932
Recalling firm
Beckman Coulter, Chaska, MN
Classification
Class II
Status
Ongoing
Recall initiated
7 Nov 2025
FDA classified
9 Dec 2025
Reported
17 Dec 2025
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Foreign Material

Timeline

7 Nov 2025Recall initiated by company
9 Dec 2025Classified by FDA+32 days
17 Dec 2025Published in enforcement report+8 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

It has been determined that certain lots of Access 2 Reaction Vessels may contain manufacturing deformities. The bottom (round end) of the Reaction Vessel may have a small protrusion or hair like deformity. This could cause causing instrument errors (e.g., pipetting, rake, wash carousel errors) that could delay patient results.

Foreign material recalls cover pieces of metal, plastic, glass, wood or other objects that got into the product during processing and could cause injury if eaten.

Product as listed by the FDA

Access 2 Reaction Vessels, individual, disposable, plastic tubes used to process patient, quality control and calibration samples on the Access 2 Immunoassay system, Catalog Number 81901

Where it was distributed

US Nationwide. Global Distribution.

Nationwide

Questions about this recall

Why was it recalled?

It has been determined that certain lots of Access 2 Reaction Vessels may contain manufacturing deformities. The bottom (round end) of the Reaction Vessel may have a small protrusion or hair like deformity. This could cause causing instrument errors (e.g., pipetting, rake, wash carousel errors) that could delay patient results.

Which lots are affected?

UDI-DI: 15099590180249; Lot Numbers: P243943, P243971, P244202 P244344, P244345, P250001, P250002, P250003, P250004 and P250005

Where was it sold?

US Nationwide. Global Distribution.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

More recalls from Beckman Coulter

All 334

Other Test Kits & Diagnostics recalls

All 2,656