CLASS II ONGOING Device recall · Reported 12 Aug 2026 · FDA Enforcement Report

REF # D23352, MicroScan LabPro V5.03 Panel Update 11 (US) recalled

Beckman Coulter is recalling REF # D23352, MicroScan LabPro V5.03 Panel Update 11 (US), Electronic, distributed nationwide in the US. The recall was initiated on 26 Jun 2026 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesUDI: 15099590799243/ V5.03 Panel Update 11
Quantity recalled158 instruments

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-2888-2026
FDA event ID
99349
Recalling firm
Beckman Coulter, Brea, CA
Classification
Class II
Status
Ongoing
Recall initiated
26 Jun 2026
FDA classified
5 Aug 2026
Reported
12 Aug 2026
Type
Voluntary: Firm initiated
Reason category
Other Reasons

Timeline

26 Jun 2026Recall initiated by company
5 Aug 2026Classified by FDA+40 days
12 Aug 2026Published in enforcement report+7 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

Analyzers contain incorrect algorithms which may affect antimicrobial susceptibility test results, leading to false susceptibility classifications."

Product as listed by the FDA

REF # D23352, MicroScan LabPro V5.03 Panel Update 11 (US), Electronic installed on the following instruments: - B1018-280 MicroScan autoSCAN-4 Instrument - B1018-283 MicroScan WalkAway-40 plus Instrument - B1018-284 MicroScan WalkAway-96 plus Instrument - B1018-440 DxM-1040 MicroScan WalkAway Instrument - B1018-496 DxM-1096 MicroScan WalkAway Instrument

Where it was distributed

US Nationwide distribution including in the states of AK, AL, AR, CA, CO, DE, FL, GA, IA, IL, IN, KY, LA, MA, MD, ME, MI, MN, MO, MS, NC, ND, NE, NJ, NV, NY, OH, OK, OR, PA, SC, TN, TX, VA, WI, WV, WY.

Nationwide AlaskaAlabamaArkansasCaliforniaColoradoDelawareFloridaGeorgiaIowaIllinoisIndianaKentuckyLouisianaMassachusettsMarylandMaineMichiganMinnesotaMissouriMississippi

Questions about this recall

Why was it recalled?

Analyzers contain incorrect algorithms which may affect antimicrobial susceptibility test results, leading to false susceptibility classifications."

Which lots are affected?

UDI: 15099590799243/ V5.03 Panel Update 11

Where was it sold?

US Nationwide distribution including in the states of AK, AL, AR, CA, CO, DE, FL, GA, IA, IL, IN, KY, LA, MA, MD, ME, MI, MN, MO, MS, NC, ND, NE, NJ, NV, NY, OH, OK, OR, PA, SC, TN, TX, VA, WI, WV, WY.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

Other products in this recall event

1 other product · FDA event 99349

This product is part of a recall event; the main page for the event is REF # D23354, MicroScan LabPro V5.03 Panel Update 11 DVD Kit (US) recalled.

ReportedClassRecallCompanyReasonWhere
CLASS IIREF # D23354, MicroScan LabPro V5.03 Panel Update 11 DVD Kit (US) recalledBeckman CoulterOtherNationwide

More recalls from Beckman Coulter

All 334

Other Other Medical Devices recalls

All 13,884