CLASS II ONGOING Device recall · Reported 11 Jun 2025 · FDA Enforcement Report

Access Testosterone assay recalled by Beckman Coulter

Beckman Coulter is recalling Access Testosterone assay, distributed in Florida. The recall was initiated on 5 May 2025 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesUDI-DI: 15099590231347; Serial Number: 538081
Quantity recalled70 units

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-1901-2025
FDA event ID
96824
Recalling firm
Beckman Coulter, Chaska, MN
Classification
Class II
Status
Ongoing
Recall initiated
5 May 2025
FDA classified
2 Jun 2025
Reported
11 Jun 2025
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Potency & Assay

Timeline

5 May 2025Recall initiated by company
2 Jun 2025Classified by FDA+28 days
11 Jun 2025Published in enforcement report+9 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

Inadequate preventative maintenance (PM) performed on the scales of the RPL5 fill line, led to splashing (cross-contamination) between the wells of testosterone reagent packs lot 538081, which could potentially result in up to a 34% increase or a 23% decrease in patient results.

Potency and specification recalls happen when a batch tests outside its approved limits for strength, purity or dissolution, which can mean the dose delivered is higher or lower than labelled.

Product as listed by the FDA

Access Testosterone assay, Catalog Number 33560

Where it was distributed

US State: FL. UAE, Vietnam

Florida

Questions about this recall

Why was it recalled?

Inadequate preventative maintenance (PM) performed on the scales of the RPL5 fill line, led to splashing (cross-contamination) between the wells of testosterone reagent packs lot 538081, which could potentially result in up to a 34% increase or a 23% decrease in patient results.

Which lots are affected?

UDI-DI: 15099590231347; Serial Number: 538081

Where was it sold?

US State: FL. UAE, Vietnam

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

More recalls from Beckman Coulter

All 336

Other Test Kits & Diagnostics recalls

All 2,663