CLASS II ONGOING Device recall · Reported 28 Oct 2026 · FDA Enforcement Report

Rotor Assy-SW40, Reference Number recalled by Beckman Coulter

Beckman Coulter is recalling Rotor Assy-SW40, Reference Number 331302, when the rotors are installed in the following, distributed nationwide in the US. The recall was initiated on 21 Sep 2026 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesSerial Numbers: 24U15260 to 26U15764; Optima UDI-DI: A99846 - 15099590416324, A99842 - 15099590415754, A99839 - 15099590415822, A99836 - 15099590415846, A99833 - 15099590415860
Quantity recalled92 units

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-0201-2027
FDA event ID
99882
Recalling firm
Beckman Coulter, Indianapolis, IN
Classification
Class II
Status
Ongoing
Recall initiated
21 Sep 2026
FDA classified
21 Oct 2026
Reported
28 Oct 2026
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Device Malfunction

Timeline

21 Sep 2026Recall initiated by company
21 Oct 2026Classified by FDA+30 days
28 Oct 2026Published in enforcement report+7 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

Optima XE XPN ultracentrifuge rotors may contain a microstructure defect that can cause rotors to break into two or more pieces (bifurcate) during a centrifugation run, resulting in destruction of samples being processed. When used for in vitro diagnostic (IVD) sample preparation, rotor breakage may delay patient test results. Additionally, sample destruction may expose laboratory personnel to biohazardous materials.

Device recalls are issued when a component, software or design fault could cause the device to fail or give wrong readings. Many are corrections applied in the field rather than physical returns.

Product as listed by the FDA

ROTOR ASSY-SW40, Reference Number 331302, when the rotors are installed in the following centrifuges: Optima XPN-100 (Model Numbers A99846 and A94469), Optima XPN-90 (Model Numbers A99842 and A94468), Optima XPN-80 (Model Numbers A99839 and A95765), Optima XE-100 (Model Numbers A99836 and A94516), and Optima XE-90 (Model Numbers A99833 and A94471)

Where it was distributed

US Nationwide distribution.

Nationwide

Questions about this recall

Why was it recalled?

Optima XE XPN ultracentrifuge rotors may contain a microstructure defect that can cause rotors to break into two or more pieces (bifurcate) during a centrifugation run, resulting in destruction of samples being processed. When used for in vitro diagnostic (IVD) sample preparation, rotor breakage may delay patient test results. Additionally, sample destruction may expose laboratory personnel to biohazardous materials.

Which lots are affected?

Serial Numbers: 24U15260 to 26U15764; Optima UDI-DI: A99846 - 15099590416324, A99842 - 15099590415754, A99839 - 15099590415822, A99836 - 15099590415846, A99833 - 15099590415860

Where was it sold?

US Nationwide distribution.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 3 Nov 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

Other products in this recall event

1 other product · FDA event 99882
ReportedClassRecallCompanyReasonWhere
CLASS IIRotor Assy-SW41, Reference Number recalled by Beckman CoulterBeckman CoulterDevice MalfunctionNationwide

More recalls from Beckman Coulter

All 340

Other Other Medical Devices recalls

All 13,950