CLASS II Device recall · Reported 8 Apr 2015 · FDA Enforcement Report
Accu-chek Connect Diabetes Management App recalled
Roche Diabetes Care is recalling Accu-chek Connect Diabetes Management App. The recall was initiated on 30 Oct 2014 and the FDA has classified it as Class II.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-1369-2015
- FDA event ID
- 70775
- Recalling firm
- Roche Diabetes Care, Indianapolis, IN
- Classification
- Class II
- Status
- Terminated
- Recall initiated
- 30 Oct 2014
- FDA classified
- 2 Apr 2015
- Reported
- 8 Apr 2015
- Type
- Voluntary: Firm initiated
- Customers notified by
- Letter
- Reason category
- Labeling Errors
- Device type
- Medical Software
Timeline
| 30 Oct 2014 | Recall initiated by company | |
| 2 Apr 2015 | Classified by FDA | +154 days |
| 8 Apr 2015 | Published in enforcement report | +6 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
Roche Diabetes Care has become aware of an issue with the Accu-Chek Connect diabetes management app that could potentially lead to inaccurate bolus advice being provided to the user. A thorough investigation of the situation revealed that this issue may occur if the user changes the screen orientation of the phone from portrait to landscape or vise versa while looking at the Bolus Advisor or Carb
Product as listed by the FDA
ACCU-CHEK Connect Diabetes Management App; Instruction Manual Designed to transfer data for diabetes management.
Where it was distributed
There are no US distributors as the affected device is not marketed in the US. The affected device was distributed in Germany, Italy and South Africa. The application has been downloaded total of 644 times (Italy-219, South Africa-24, Germany-401) and there are 113 bolus advice activations.
Questions about this recall
Why was it recalled?
Roche Diabetes Care has become aware of an issue with the Accu-Chek Connect diabetes management app that could potentially lead to inaccurate bolus advice being provided to the user. A thorough investigation of the situation revealed that this issue may occur if the user changes the screen orientation of the phone from portrait to landscape or vise versa while looking at the Bolus Advisor or Carb
Which lots are affected?
N/A
Where was it sold?
There are no US distributors as the affected device is not marketed in the US. The affected device was distributed in Germany, Italy and South Africa. The application has been downloaded total of 644 times (Italy-219, South Africa-24, Germany-401) and there are 113 bolus advice activations.
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class II mean?
Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.
Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
More recalls from Roche Diabetes Care
All 29| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Accu-Chek Guide (SC) Kit -Intended to quantitatively recalled | Roche Diabetes Care | Other | Nationwide |
| Device | CLASS II | RocheDiabetes Care Platform recalled | Roche Diabetes Care | Device Malfunction | Nationwide |
| Device | CLASS II | Accu-Chek Inform II Test Strip, Whole Blood Glucose Test System recalled | Roche Diabetes Care | Labeling Errors | Nationwide |
| Device | CLASS II | Accu-Chek SmartView Test Strip, Blood Glucose Monitoring System recalled | Roche Diabetes Care | Labeling Errors | Nationwide |
| Device | CLASS II | Accu-Chek Aviva Plus Test Strip, Blood Glucose Monitoring System recalled | Roche Diabetes Care | Labeling Errors | Nationwide |
Other Medical Software recalls
All 903| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Philips Brilliance recalled by Philips North America | Philips North America | Device Malfunction | MA |
| Device | CLASS II | Philips IQon Spectral CT recalled by Philips North America | Philips North America | Device Malfunction | MA |
| Device | CLASS II | Philips Ingenuity CT recalled by Philips North America | Philips North America | Device Malfunction | MA |
| Device | CLASS II | Philips Brilliance CT 64: Brilliance CT 64 Channel recalled | Philips North America | Device Malfunction | MA |
| Device | ANNOUNCED | BMC Luna G3 APAP recalled by BMC MedicalBMC | BMC Medical | Device Malfunction | Nationwide |