CLASS II ONGOING Device recall · Reported 3 Jul 2024 · FDA Enforcement Report

Accu-Chek Guide (SC) Kit -Intended to quantitatively recalled

Roche Diabetes Care is recalling Accu-Chek Guide (SC) Kit -Intended to quantitatively measure glucose in fresh capillary, distributed nationwide in the US. The recall was initiated on 13 May 2024 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesGTIN: 00365702729100 Serial Numbers: US: 92339920445, 92340117408, 92339920116, 92339955415, 92340120006, 92339094787, 92340116052,92340120057, 92339744998. Canada:93040305146
Quantity recalled48 units

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-2174-2024
FDA event ID
94655
Recalling firm
Roche Diabetes Care, Indianapolis, IN
Classification
Class II
Status
Ongoing
Recall initiated
13 May 2024
FDA classified
21 Jun 2024
Reported
3 Jul 2024
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Other Reasons

Timeline

13 May 2024Recall initiated by company
21 Jun 2024Classified by FDA+39 days
3 Jul 2024Published in enforcement report+12 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

The reason for recall is the meters may show an incorrect measurement unit (mmol/L rather than mg/dL) and may result in the wrong unit of measure. Which may subsequently cause an error in interpreting a result as approximately 18 times lower that the actual blood glucose. Potential immediate health consequences of the glucose meter providing a reading in the wrong units of measurement (UOM) may result in inappropriate rescue therapy for presumed hypoglycemia including urgent and repeat administration of carbohydrates. This in turn could lead to frequent episodes of hyperglycemia which is unrecognized and undertreated.Long range consequences of continued use of the affected meters could include mismanagement of the patient s diabetes for longer periods of time potentially resulting in a major health event (in general persistent and/or severe hyperglycemia) that may lead them to seek medical attention, and depending on the duration of hyperglycemia, could contribute to the development of microvascular (i.e., retinopathy, neuropathy, nephropathy) and/or macrovascular (i.e., myocardial infarction, ischemic stroke) complications of diabetes and could lead to fetal damage (abnormal fetal growth and altered organ development and maturation, fetal hypoxemia, spontaneous abortion, and congenital anomalies) in pregnant women. This issue was identified by three initial consumer complaints, with reports that the meters were presenting the incorrect unit of measure.

Product as listed by the FDA

Accu-Chek Guide (SC) Kit -Intended to quantitatively measure glucose in fresh capillary whole blood from the fingertip, palm, and upper arm as an aid in monitoring the effectiveness of glucose control. Product REF Number: 08453071001

Where it was distributed

Nationwide and Worldwide

Nationwide

Questions about this recall

Why was it recalled?

The reason for recall is the meters may show an incorrect measurement unit (mmol/L rather than mg/dL) and may result in the wrong unit of measure. Which may subsequently cause an error in interpreting a result as approximately 18 times lower that the actual blood glucose. Potential immediate health consequences of the glucose meter providing a reading in the wrong units of measurement (UOM) may result in inappropriate rescue therapy for presumed hypoglycemia including urgent and repeat administration of carbohydrates. This in turn could lead to frequent episodes of hyperglycemia which is unrecognized and undertreated.Long range consequences of continued use of the affected meters could include mismanagement of the patient s diabetes for longer periods of time potentially resulting in a major health event (in general persistent and/or severe hyperglycemia) that may lead them to seek medical attention, and depending on the duration of hyperglycemia, could contribute to the development of microvascular (i.e., retinopathy, neuropathy, nephropathy) and/or macrovascular (i.e., myocardial infarction, ischemic stroke) complications of diabetes and could lead to fetal damage (abnormal fetal growth and altered organ development and maturation, fetal hypoxemia, spontaneous abortion, and congenital anomalies) in pregnant women. This issue was identified by three initial consumer complaints, with reports that the meters were presenting the incorrect unit of measure.

Which lots are affected?

GTIN: 00365702729100 Serial Numbers: US: 92339920445, 92340117408, 92339920116, 92339955415, 92340120006, 92339094787, 92340116052,92340120057, 92339744998. Canada:93040305146

Where was it sold?

Nationwide and Worldwide

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.