CLASS II ONGOING Device recall · Reported 20 Sep 2023 · FDA Enforcement Report
Acuson Redwood 2.0 ultrasound systems recalled
Siemens Medical Solutions USA is recalling Acuson Redwood 2.0 ultrasound systems, distributed nationwide in the US. The recall was initiated on 13 Jul 2023 and the FDA has classified it as Class II.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-2582-2023
- FDA event ID
- 92750
- Recalling firm
- Siemens Medical Solutions USA, Issaquah, WA
- Classification
- Class II
- Status
- Ongoing
- Recall initiated
- 13 Jul 2023
- FDA classified
- 13 Sep 2023
- Reported
- 20 Sep 2023
- Type
- Voluntary: Firm initiated
- Customers notified by
- Letter
- Reason category
- Foreign Material
- Device type
- Imaging Equipment
Timeline
| 13 Jul 2023 | Recall initiated by company | |
| 13 Sep 2023 | Classified by FDA | +62 days |
| 20 Sep 2023 | Published in enforcement report | +7 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
If a user-generated preset for an 18L6 transducer created on a 1.0 ultrasound system is used with a 2.0 ultrasound system, the 2.0 ultrasound system will display underestimated measurement results when using an 18L6 transducer and viewing in the Dual format visualization function, which may lead to misdiagnosis of a patient's condition or influence patient management decisions in a negative way.
Foreign material recalls cover pieces of metal, plastic, glass, wood or other objects that got into the product during processing and could cause injury if eaten.
Product as listed by the FDA
ACUSON Redwood 2.0 ultrasound systems, REF: 11503314
Where it was distributed
Worldwide - US Nationwide distribution including in the states of IL, WI, CA, NM, TX, NC, MI, MD, NY, OH, IN, NE, PA, LA, SC, MO, OK, VA, ME, AZ, PR, IA, NH, NJ, AR, FL, SD and the countries of GR, ES, SG, TH, KR, IN, DE, TR, AT, IT, AU, JP, FR, CA, PT, SK, BR, SA, GB, NZ, RO, SE, AE, CN, HR, PL, CH, CR, IQ, NL, MX, ZA, BO, PH, MA, PA, EG, FJ, JO, DZ, TW, DK, CZ, CO, UG, CL, VN, CM, BG, QA, AZ, EC, KW, MY, PE, GG, SV, FO, ID, NP, BA, IL, LT, DO, IE.
Nationwide ArkansasArizonaCaliforniaColoradoDelawareFloridaIowaIdahoIllinoisIndianaLouisianaMassachusettsMarylandMaineMichiganMissouriNorth CarolinaNebraskaNew HampshireNew Jersey
Questions about this recall
Why was it recalled?
If a user-generated preset for an 18L6 transducer created on a 1.0 ultrasound system is used with a 2.0 ultrasound system, the 2.0 ultrasound system will display underestimated measurement results when using an 18L6 transducer and viewing in the Dual format visualization function, which may lead to misdiagnosis of a patient's condition or influence patient management decisions in a negative way.
Which lots are affected?
UDI-DI: 04056869251264. Systems with software version prefix VA20.
Where was it sold?
Worldwide - US Nationwide distribution including in the states of IL, WI, CA, NM, TX, NC, MI, MD, NY, OH, IN, NE, PA, LA, SC, MO, OK, VA, ME, AZ, PR, IA, NH, NJ, AR, FL, SD and the countries of GR, ES, SG, TH, KR, IN, DE, TR, AT, IT, AU, JP, FR, CA, PT, SK, BR, SA, GB, NZ, RO, SE, AE, CN, HR, PL, CH, CR, IQ, NL, MX, ZA, BO, PH, MA, PA, EG, FJ, JO, DZ, TW, DK, CZ, CO, UG, CL, VN, CM, BG, QA, AZ, EC, KW, MY, PE, GG, SV, FO, ID, NP, BA, IL, LT, DO, IE.
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class II mean?
Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.
Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
More recalls from Siemens Medical Solutions USA
All 796| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Artis icono ceiling recalled by Siemens Medical Solutions USA | Siemens Medical Solutions USA | Other | Nationwide |
| Device | CLASS II | Artis icono biplane recalled by Siemens Medical Solutions USA | Siemens Medical Solutions USA | Other | Nationwide |
| Device | CLASS II | Artis icono floor recalled by Siemens Medical Solutions USA | Siemens Medical Solutions USA | Other | Nationwide |
| Device | CLASS II | Artis icono ceiling recalled by Siemens Medical Solutions USA | Siemens Medical Solutions USA | Other | Nationwide |
| Device | CLASS II | Artis pheno recalled by Siemens Medical Solutions USA | Siemens Medical Solutions USA | Other | PA |
Other Imaging Equipment recalls
All 2,336| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Medline Convenience Kits containing BD ChloraPrep Clear recalled | Medline Industries | Packaging Defects | Nationwide |
| Device | CLASS II | Medline Convenience Kits: Ultrasound Breast Pack recalled | Medline Industries | Sterility | Nationwide |
| Device | CLASS II | Alphenix Models Infx-8000C Infx-8000F Infx-8000V This device recalled | Canon Medical System USA | Other | Nationwide |
| Device | CLASS II | Merge PACS and Merge OrthoPACS recalled by Merge Healthcare | Merge Healthcare | Labeling Errors | Nationwide |
| Device | CLASS II | Philips Azurion recalled by Philips Medical Systems | Philips Medical Systems Nederl | Potency & Assay |