CLASS II ONGOING Device recall · Reported 20 Sep 2023 · FDA Enforcement Report

Acuson Redwood 2.0 ultrasound systems recalled

Siemens Medical Solutions USA is recalling Acuson Redwood 2.0 ultrasound systems, distributed nationwide in the US. The recall was initiated on 13 Jul 2023 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesUDI-DI: 04056869251264. Systems with software version prefix VA20.
Quantity recalled796

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-2582-2023
FDA event ID
92750
Recalling firm
Siemens Medical Solutions USA, Issaquah, WA
Classification
Class II
Status
Ongoing
Recall initiated
13 Jul 2023
FDA classified
13 Sep 2023
Reported
20 Sep 2023
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Foreign Material

Timeline

13 Jul 2023Recall initiated by company
13 Sep 2023Classified by FDA+62 days
20 Sep 2023Published in enforcement report+7 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

If a user-generated preset for an 18L6 transducer created on a 1.0 ultrasound system is used with a 2.0 ultrasound system, the 2.0 ultrasound system will display underestimated measurement results when using an 18L6 transducer and viewing in the Dual format visualization function, which may lead to misdiagnosis of a patient's condition or influence patient management decisions in a negative way.

Foreign material recalls cover pieces of metal, plastic, glass, wood or other objects that got into the product during processing and could cause injury if eaten.

Product as listed by the FDA

ACUSON Redwood 2.0 ultrasound systems, REF: 11503314

Where it was distributed

Worldwide - US Nationwide distribution including in the states of IL, WI, CA, NM, TX, NC, MI, MD, NY, OH, IN, NE, PA, LA, SC, MO, OK, VA, ME, AZ, PR, IA, NH, NJ, AR, FL, SD and the countries of GR, ES, SG, TH, KR, IN, DE, TR, AT, IT, AU, JP, FR, CA, PT, SK, BR, SA, GB, NZ, RO, SE, AE, CN, HR, PL, CH, CR, IQ, NL, MX, ZA, BO, PH, MA, PA, EG, FJ, JO, DZ, TW, DK, CZ, CO, UG, CL, VN, CM, BG, QA, AZ, EC, KW, MY, PE, GG, SV, FO, ID, NP, BA, IL, LT, DO, IE.

Nationwide ArkansasArizonaCaliforniaColoradoDelawareFloridaIowaIdahoIllinoisIndianaLouisianaMassachusettsMarylandMaineMichiganMissouriNorth CarolinaNebraskaNew HampshireNew Jersey

Questions about this recall

Why was it recalled?

If a user-generated preset for an 18L6 transducer created on a 1.0 ultrasound system is used with a 2.0 ultrasound system, the 2.0 ultrasound system will display underestimated measurement results when using an 18L6 transducer and viewing in the Dual format visualization function, which may lead to misdiagnosis of a patient's condition or influence patient management decisions in a negative way.

Which lots are affected?

UDI-DI: 04056869251264. Systems with software version prefix VA20.

Where was it sold?

Worldwide - US Nationwide distribution including in the states of IL, WI, CA, NM, TX, NC, MI, MD, NY, OH, IN, NE, PA, LA, SC, MO, OK, VA, ME, AZ, PR, IA, NH, NJ, AR, FL, SD and the countries of GR, ES, SG, TH, KR, IN, DE, TR, AT, IT, AU, JP, FR, CA, PT, SK, BR, SA, GB, NZ, RO, SE, AE, CN, HR, PL, CH, CR, IQ, NL, MX, ZA, BO, PH, MA, PA, EG, FJ, JO, DZ, TW, DK, CZ, CO, UG, CL, VN, CM, BG, QA, AZ, EC, KW, MY, PE, GG, SV, FO, ID, NP, BA, IL, LT, DO, IE.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.