CLASS I Device recall · Reported 31 Jan 2018 · FDA Enforcement Report

Adult AirLife dual-limb, dual-heat circuit recalled by Vyaire Medical

Vyaire Medical is recalling Adult AirLife dual-limb, dual-heat circuit (, distributed nationwide in the US. The recall was initiated on 29 Nov 2014 and the FDA has classified it as Class I.

Check your package

Lot, batch or date codes0001100127 0001116871 0001116872 0001124936 0001129065 0001129066 0001129067 0001129069 0001129070 0001131426 0001131428 0001134350 0001135400 0001136027 0001136028 0001136029 0001145151 0001145152 0001145153 0001157125
Quantity recalled16,670 units in total

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-0346-2018
FDA event ID
78636
Recalling firm
Vyaire Medical, Vernon Hills, IL
Classification
Class I
Status
Terminated
Recall initiated
29 Nov 2014
FDA classified
24 Jan 2018
Reported
31 Jan 2018
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Other Reasons

Timeline

29 Nov 2014Recall initiated by company
24 Jan 2018Classified by FDA+1152 days
31 Jan 2018Published in enforcement report+7 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

Manufacturing error in humidification chamber may cause an overfill/overflow condition of the chamber. The may case a water back-up into the patient circuit resulting in an excessive amount of water entering the airway/lungs of the ventilated patient.

Product as listed by the FDA

Adult AirLife dual-limb, dual-heat circuit (Model AH280) with AirLife Humidification Chamber, Model Number AH290. The AH290 Humidification Chamber and Heated Breathing Circuit Kit(s) operate as part of a system for delivering health and humidified respirator gasses to patients requiring invasive and non-invasive ventilation.

Where it was distributed

US Nationwide Distribution

Nationwide

Questions about this recall

Why was it recalled?

Manufacturing error in humidification chamber may cause an overfill/overflow condition of the chamber. The may case a water back-up into the patient circuit resulting in an excessive amount of water entering the airway/lungs of the ventilated patient.

Which lots are affected?

0001100127 0001116871 0001116872 0001124936 0001129065 0001129066 0001129067 0001129069 0001129070 0001131426 0001131428 0001134350 0001135400 0001136027 0001136028 0001136029 0001145151 0001145152 0001145153 0001157125

Where was it sold?

US Nationwide Distribution

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class I mean?

Class I is the most serious FDA recall category: there is a reasonable probability that using the product will cause serious health consequences or death.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.