CLASS II Device recall · Reported 29 Apr 2015 · FDA Enforcement Report

Advia Chemistry Lipase Reagent recalled

Siemens Healthcare Diagnostics is recalling Advia Chemistry Lipase Reagent, distributed nationwide in the US. The recall was initiated on 23 Mar 2015 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesCatalog Number: B01-4840-01; Material Number (SMN)10311896, Lot Numbers: 318706, 332463, expiration dates 2015-06-30, 2015-08-31
Quantity recalled7814

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-1457-2015
FDA event ID
70835
Recalling firm
Siemens Healthcare Diagnostics, Tarrytown, NY
Classification
Class II
Status
Terminated
Recall initiated
23 Mar 2015
FDA classified
20 Apr 2015
Reported
29 Apr 2015
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Other Reasons

Timeline

23 Mar 2015Recall initiated by company
20 Apr 2015Classified by FDA+28 days
29 Apr 2015Published in enforcement report+9 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

Siemens internal investigation confirmed that current contamination avoidance settings do not prevent carryover resulting in falsely elevated ADVIA Chemistry Lipase outlier results when running with Triglycerides, Triglycerides_2, Triglycerides_c, Cholesterol concentrated, and DLDL reagents.

Product as listed by the FDA

ADVIA Chemistry Lipase Reagent , Catalog Number B01-4840-01 REF: 01984894.

Where it was distributed

Worldwide Distribution - US (nationwide) Distribution and to the countries of : Thailand, Pakistan, Vatican, Guadeloupe, Reunion, French Polynesia , Serbia, Martinique, India, China, Singapore, Brazil, South Africa, Mexico, Canada, Israel, Australia, Egypt, Russian Federation, Indonesia, Rep. Korea (S) ,Unit.Arab Emir., Taiwan ,Argentina ,Chile, Malaysia ,Colombia, Peru, Kazakhstan, Paraguay, Vietnam, Algeria, Austria, Belgium, Switzerland, Cyprus, Czech Republic, Germany, Estonia, Spain, Finland, France, United Kingdom, Greece, Hungary, Italy, Lithuania, Luxembourg, Latvia, Netherlands, Norway, Poland, Portugal, Romania, Slovenia, Slovakia and Turkey.

Nationwide

Questions about this recall

Why was it recalled?

Siemens internal investigation confirmed that current contamination avoidance settings do not prevent carryover resulting in falsely elevated ADVIA Chemistry Lipase outlier results when running with Triglycerides, Triglycerides_2, Triglycerides_c, Cholesterol concentrated, and DLDL reagents.

Which lots are affected?

Catalog Number: B01-4840-01; Material Number (SMN)10311896, Lot Numbers: 318706, 332463, expiration dates 2015-06-30, 2015-08-31

Where was it sold?

Worldwide Distribution - US (nationwide) Distribution and to the countries of : Thailand, Pakistan, Vatican, Guadeloupe, Reunion, French Polynesia , Serbia, Martinique, India, China, Singapore, Brazil, South Africa, Mexico, Canada, Israel, Australia, Egypt, Russian Federation, Indonesia, Rep. Korea (S) ,Unit.Arab Emir., Taiwan ,Argentina ,Chile, Malaysia ,Colombia, Peru, Kazakhstan, Paraguay, Vietnam, Algeria, Austria, Belgium, Switzerland, Cyprus, Czech Republic, Germany, Estonia, Spain, Finland, France, United Kingdom, Greece, Hungary, Italy, Lithuania, Luxembourg, Latvia, Netherlands, Norway, Poland, Portugal, Romania, Slovenia, Slovakia and Turkey.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.