CLASS II ONGOING Device recall · Reported 30 Aug 2023 · FDA Enforcement Report

The Air/Water Valve is provided/used recalled

Olympus Corporation of the Americas is recalling The Air/Water Valve is provided/used with the following Endoscopes: Ultrasonic, distributed nationwide in the US. The recall was initiated on 12 Jul 2023 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesUDI-DI: 04953170355929. All lot numbers
Quantity recalled29590 units

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-2447-2023
FDA event ID
92728
Recalling firm
Olympus Corporation of the Americas, Center Valley, PA
Classification
Class II
Status
Ongoing
Recall initiated
12 Jul 2023
FDA classified
23 Aug 2023
Reported
30 Aug 2023
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Other Reasons

Timeline

12 Jul 2023Recall initiated by company
23 Aug 2023Classified by FDA+42 days
30 Aug 2023Published in enforcement report+7 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

The air/water valve MAJ-1444 used with OER-Pro and OER-Elite may become damaged and result in loss of the one-way valve functionality by repeated automated endoscope reprocessing in Olympus OER machines, causing body fluid could backflow into the air/water channel of the ultrasonic endoscopes during the procedure.

Product as listed by the FDA

The Air/Water Valve is provided/used with the following Endoscopes: ULTRASONIC GASTROFIBERSCOPE, ULTRASONIC GASTROVIDEOSCOPE, ULTRASONIC COLONOVIDEOSCOPE: GF-UC140P-AL5, GF-UCT140-AL5, GF-UE160-AL5, GF-UCT180, GF-UM20, GF-UM130, GF-UMQ130, GF-UM160, GF-UC160P-OL5, GF-UCT160-OL5

Where it was distributed

US Nationwide distribution.

Nationwide

Questions about this recall

Why was it recalled?

The air/water valve MAJ-1444 used with OER-Pro and OER-Elite may become damaged and result in loss of the one-way valve functionality by repeated automated endoscope reprocessing in Olympus OER machines, causing body fluid could backflow into the air/water channel of the ultrasonic endoscopes during the procedure.

Which lots are affected?

UDI-DI: 04953170355929. All lot numbers

Where was it sold?

US Nationwide distribution.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

More recalls from Olympus Corporation of the Americas

All 415

Other Surgical Instruments recalls

All 3,809