CLASS II ONGOING Device recall · Reported 22 Oct 2025 · FDA Enforcement Report
Airvo 2, REF: PT101US, and myAirvo 2, REF: PT100US, used recalled
Fisher & Paykel Healthcare is recalling Airvo 2, REF: PT101US, and myAirvo 2, REF: PT100US, used with, distributed nationwide in the US. The recall was initiated on 18 Sep 2025 and the FDA has classified it as Class II.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-0158-2026
- FDA event ID
- 97591
- Recalling firm
- Fisher & Paykel Healthcare, Auckland
- Classification
- Class II
- Status
- Ongoing
- Recall initiated
- 18 Sep 2025
- FDA classified
- 15 Oct 2025
- Reported
- 22 Oct 2025
- Type
- Voluntary: Firm initiated
- Customers notified by
- Letter
- Reason category
- Other Reasons
- Device type
- Other Medical Devices
Timeline
| 18 Sep 2025 | Recall initiated by company | |
| 15 Oct 2025 | Classified by FDA | +27 days |
| 22 Oct 2025 | Published in enforcement report | +7 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
When unintentionally disconnected from power source, humidifier device (used to deliver high flow warmed and humidified respiratory gases) may sound audible alert for less than 120 seconds, which may delay user awareness to reconnect power, which could lead to oxygen desaturation, hypoxia. Disinfection kit user manual updated to include checking the audible alert prior to use on a new patient.
Product as listed by the FDA
Airvo 2, REF: PT101US, and myAirvo 2, REF: PT100US, used with Disinfection Kit User Manual, REF: UI-185043723
Where it was distributed
US Nationwide distribution including in the states of TX, MS, LA, NC, TN, NY, FL, IL, UT, MT, OH, MI, NJ, AR, GA, WI, CA, IN, OK, CO, AL, SC, AZ, MD, WA, MO, PA, MA, NH, KY, KS, VA, CT, MN, ND, NM, NE, IA, WV, ID, NV, OR, HI, VT, RI, AL, SD, WY, PR, ME.
Nationwide AlabamaArkansasArizonaCaliforniaColoradoConnecticutFloridaGeorgiaHawaiiIowaIdahoIllinoisIndianaKansasKentuckyLouisianaMassachusettsMarylandMaineMichigan
Questions about this recall
Why was it recalled?
When unintentionally disconnected from power source, humidifier device (used to deliver high flow warmed and humidified respiratory gases) may sound audible alert for less than 120 seconds, which may delay user awareness to reconnect power, which could lead to oxygen desaturation, hypoxia. Disinfection kit user manual updated to include checking the audible alert prior to use on a new patient.
Which lots are affected?
REF/UDI-DI: PT101US/9420012422347; PT100US/9420012422248. Disinfection Kit User Manual all revisions before UI-185043723 rev P
Where was it sold?
US Nationwide distribution including in the states of TX, MS, LA, NC, TN, NY, FL, IL, UT, MT, OH, MI, NJ, AR, GA, WI, CA, IN, OK, CO, AL, SC, AZ, MD, WA, MO, PA, MA, NH, KY, KS, VA, CT, MN, ND, NM, NE, IA, WV, ID, NV, OR, HI, VT, RI, AL, SD, WY, PR, ME.
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class II mean?
Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.
Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
More recalls from Fisher & Paykel Healthcare
All 7| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | PT301US Airvo 3 Respiratory Support Device recalled | Fisher & Paykel Healthcare | Device Malfunction | 29 states |
| Device | CLASS II | Airvo 2 Humidifier recalled by Fisher & Paykel Healthcare | Fisher & Paykel Healthcare | Device Malfunction | Nationwide |
| Device | CLASS II | (FPH) IW900-Series Infant Warmer Fisher and Paykel recalled | Fisher & Paykel Healthcare | Device Malfunction | Nationwide |
| Device | CLASS I | Infant Nasal CPAP (continuous positive airway pressure) Prong, Nasal recalled | Fisher & Paykel Healthcare | Other | Nationwide |
| Device | CLASS II | Icon CPAP (continuous positive airway pressure) recalled | Fisher & Paykel Healthcare | Other | Nationwide |
Other Other Medical Devices recalls
All 13,778| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Citric Complete Dry Citric Acid Concentrate recalled | Nipro Renal Soultions USA | Other | Nationwide |
| Device | CLASS I | Kit: Epidural/Pain TQC 20/Cs recalled by Medical Action Industries | Medical Action Industries Inc. 306 | Sterility | 6 states |
| Device | CLASS II | Halyard IV Start Kit Non-Sterile recalled by Avid Medical | Avid Medical | Sterility | Nationwide |
| Device | CLASS I | Kit: Single Shot Epidural 20/Cs recalled by Medical Action Industries | Medical Action Industries Inc. 306 | Sterility | 6 states |
| Device | CLASS II | AMS Convenience Kit: Dsaek Pack - RX recalled | Windstone Medical Packaging | Device Malfunction | Nationwide |