CLASS II ONGOING Device recall · Reported 2 Oct 2024 · FDA Enforcement Report
PT301US Airvo 3 Respiratory Support Device recalled
Fisher & Paykel Healthcare is recalling PT301US Airvo 3 Respiratory Support Device, distributed in 29 states. The recall was initiated on 7 Aug 2024 and the FDA has classified it as Class II.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-3262-2024
- FDA event ID
- 95173
- Recalling firm
- Fisher & Paykel Healthcare, Auckland
- Classification
- Class II
- Status
- Ongoing
- Recall initiated
- 7 Aug 2024
- FDA classified
- 25 Sep 2024
- Reported
- 2 Oct 2024
- Type
- Voluntary: Firm initiated
- Customers notified by
- Letter
- Reason category
- Device Malfunction
- Device type
- CPAP & Ventilators
Timeline
| 7 Aug 2024 | Recall initiated by company | |
| 25 Sep 2024 | Classified by FDA | +49 days |
| 2 Oct 2024 | Published in enforcement report | +7 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
Due to a software issue, affected devices that are set up with High Pressure Oxygen (HPO), if the flow alignment alarm occurs, the device will deliver room air only. If this happens, a patient my experience oxygen desaturation that could lead to hypoxia.
Device recalls are issued when a component, software or design fault could cause the device to fail or give wrong readings. Many are corrections applied in the field rather than physical returns.
Product as listed by the FDA
PT301US Airvo 3 Respiratory Support Device with software version 1.2.0 and/or 1.5.1 The Airvo 3 is intended to provide high flow warmed and humidified respiratory gases for administration to spontaneously breathing infant, child, adolescent and adult patients in hospitals and sub-acute facilities.
Where it was distributed
U.S.: AZ, CA, CO, CT, DC, FL, GA, IA, IL, KY, MA, MD, MI, MN, MO, MT, ND, NH, NJ, NV, NY, OH, OR, PA, SD, TX, UT, VA and WA. O.U.S.: Not provided
ArizonaCaliforniaColoradoConnecticutDistrict of ColumbiaFloridaGeorgiaIowaIllinoisKentuckyMassachusettsMarylandMichiganMinnesotaMissouriMontanaNorth DakotaNew HampshireNew JerseyNevada
Questions about this recall
Why was it recalled?
Due to a software issue, affected devices that are set up with High Pressure Oxygen (HPO), if the flow alignment alarm occurs, the device will deliver room air only. If this happens, a patient my experience oxygen desaturation that could lead to hypoxia.
Which lots are affected?
Model Number: PT301US UDI-DI code: 09420012466662 Lot/Serial Numbers:
Where was it sold?
U.S.: AZ, CA, CO, CT, DC, FL, GA, IA, IL, KY, MA, MD, MI, MN, MO, MT, ND, NH, NJ, NV, NY, OH, OR, PA, SD, TX, UT, VA and WA. O.U.S.: Not provided
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class II mean?
Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.
Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
More recalls from Fisher & Paykel Healthcare
All 7| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Airvo 2, REF: PT101US, and myAirvo 2, REF: PT100US, used recalled | Fisher & Paykel Healthcare | Other | Nationwide |
| Device | CLASS II | Airvo 2 Humidifier recalled by Fisher & Paykel Healthcare | Fisher & Paykel Healthcare | Device Malfunction | Nationwide |
| Device | CLASS II | (FPH) IW900-Series Infant Warmer Fisher and Paykel recalled | Fisher & Paykel Healthcare | Device Malfunction | Nationwide |
| Device | CLASS I | Infant Nasal CPAP (continuous positive airway pressure) Prong, Nasal recalled | Fisher & Paykel Healthcare | Other | Nationwide |
| Device | CLASS II | Icon CPAP (continuous positive airway pressure) recalled | Fisher & Paykel Healthcare | Other | Nationwide |
Other CPAP & Ventilators recalls
All 554| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS I | Inca (Infant Nasal Cannulae Assy) Infant Nasal CPAP, Complete Set recalled | CooperSurgical | Other | Nationwide |
| Device | CLASS I | Inca (Infant Nasal Cannulae Assy) Neonatal Nasal CPAP recalled | CooperSurgical | Other | Nationwide |
| Device | CLASS II | Medline Hudson Rci, Dbd-circuit,Vent,Inspiratory,Heated,Neo recalled | Medline Industries | Device Malfunction | Nationwide |
| Device | CLASS II | Getinge Y Sensor Module, Servo-u, Servo recalled | Maquet Critical Care AB | Device Malfunction | Nationwide |
| Device | CLASS II | Biofire Respiratory Panel 2.1 (RP2.1) recalled by BioFire Diagnostics | BioFire Diagnostics | Other | Nationwide |