CLASS II ONGOING Device recall · Reported 2 Oct 2024 · FDA Enforcement Report

PT301US Airvo 3 Respiratory Support Device recalled

Fisher & Paykel Healthcare is recalling PT301US Airvo 3 Respiratory Support Device, distributed in 29 states. The recall was initiated on 7 Aug 2024 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesModel Number: PT301US UDI-DI code: 09420012466662 Lot/Serial Numbers:
Quantity recalled944 units

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-3262-2024
FDA event ID
95173
Recalling firm
Fisher & Paykel Healthcare, Auckland
Classification
Class II
Status
Ongoing
Recall initiated
7 Aug 2024
FDA classified
25 Sep 2024
Reported
2 Oct 2024
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Device Malfunction

Timeline

7 Aug 2024Recall initiated by company
25 Sep 2024Classified by FDA+49 days
2 Oct 2024Published in enforcement report+7 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

Due to a software issue, affected devices that are set up with High Pressure Oxygen (HPO), if the flow alignment alarm occurs, the device will deliver room air only. If this happens, a patient my experience oxygen desaturation that could lead to hypoxia.

Device recalls are issued when a component, software or design fault could cause the device to fail or give wrong readings. Many are corrections applied in the field rather than physical returns.

Product as listed by the FDA

PT301US Airvo 3 Respiratory Support Device with software version 1.2.0 and/or 1.5.1 The Airvo 3 is intended to provide high flow warmed and humidified respiratory gases for administration to spontaneously breathing infant, child, adolescent and adult patients in hospitals and sub-acute facilities.

Where it was distributed

U.S.: AZ, CA, CO, CT, DC, FL, GA, IA, IL, KY, MA, MD, MI, MN, MO, MT, ND, NH, NJ, NV, NY, OH, OR, PA, SD, TX, UT, VA and WA. O.U.S.: Not provided

ArizonaCaliforniaColoradoConnecticutDistrict of ColumbiaFloridaGeorgiaIowaIllinoisKentuckyMassachusettsMarylandMichiganMinnesotaMissouriMontanaNorth DakotaNew HampshireNew JerseyNevada

Questions about this recall

Why was it recalled?

Due to a software issue, affected devices that are set up with High Pressure Oxygen (HPO), if the flow alignment alarm occurs, the device will deliver room air only. If this happens, a patient my experience oxygen desaturation that could lead to hypoxia.

Which lots are affected?

Model Number: PT301US UDI-DI code: 09420012466662 Lot/Serial Numbers:

Where was it sold?

U.S.: AZ, CA, CO, CT, DC, FL, GA, IA, IL, KY, MA, MD, MI, MN, MO, MT, ND, NH, NJ, NV, NY, OH, OR, PA, SD, TX, UT, VA and WA. O.U.S.: Not provided

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

More recalls from Fisher & Paykel Healthcare

All 7

Other CPAP & Ventilators recalls

All 554