CLASS II ONGOING Device recall · Reported 12 Jun 2024 · FDA Enforcement Report

Alaris Pump Infusion Set Low Sorbing Tubing recalled

BD Switzerland Sarl is recalling Alaris Pump Infusion Set Low Sorbing Tubing (PE Lined) Back Check Valve 2 SmartSite Y, distributed nationwide in the US. The recall was initiated on 18 Apr 2024 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesCatalog Number: 10013072 UDI-DI code:07613203021210 Lot Number: 22115368
Quantity recalled11.300 sets

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-1944-2024
FDA event ID
94479
Recalling firm
BD Switzerland Sarl, Eysins, N/A
Classification
Class II
Status
Ongoing
Recall initiated
18 Apr 2024
FDA classified
31 May 2024
Reported
12 Jun 2024
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Other Reasons
Device type
Infusion Pumps

Timeline

18 Apr 2024Recall initiated by company
31 May 2024Classified by FDA+43 days
12 Jun 2024Published in enforcement report+12 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

Due to infusion set tubing drip chamber becoming detached from the tubing potentially causing delays in treatment, or unintentionally exposing users to harsh infusates.

Product as listed by the FDA

Alaris Pump Infusion Set Low Sorbing Tubing (PE Lined) Back Check Valve 2 SmartSite Y Sites REF 10013072 Tubing set used for delivery of fluids, medication, blood and blood products.

Where it was distributed

Worldwide - US Nationwide distribution in the states of AR, CA, CO, IL, KS, MA, ME, NJ, NY, PA, SD, WA and WI. The country of Canada.

Nationwide ArkansasCaliforniaColoradoIllinoisKansasMassachusettsMaineNew JerseyNew YorkPennsylvaniaSouth DakotaWashingtonWisconsin

Questions about this recall

Why was it recalled?

Due to infusion set tubing drip chamber becoming detached from the tubing potentially causing delays in treatment, or unintentionally exposing users to harsh infusates.

Which lots are affected?

Catalog Number: 10013072 UDI-DI code:07613203021210 Lot Number: 22115368

Where was it sold?

Worldwide - US Nationwide distribution in the states of AR, CA, CO, IL, KS, MA, ME, NJ, NY, PA, SD, WA and WI. The country of Canada.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

More recalls from BD Switzerland Sarl

All 15

Other Infusion Pumps recalls

All 549