CLASS II ONGOING Device recall · Reported 12 Aug 2026 · FDA Enforcement Report
Anti-Torque, Vector, Part: LS-749 part of Vector Pedicle recalled
Innovasis is recalling Anti-Torque, Vector, Part: LS-749 part of Vector Pedicle Screw System, distributed nationwide in the US. The recall was initiated on 22 May 2026 and the FDA has classified it as Class II.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-2859-2026
- FDA event ID
- 99188
- Recalling firm
- Innovasis, Salt Lake City, UT
- Classification
- Class II
- Status
- Ongoing
- Recall initiated
- 22 May 2026
- FDA classified
- 3 Aug 2026
- Reported
- 12 Aug 2026
- Type
- Voluntary: Firm initiated
- Customers notified by
- Letter
- Reason category
- Other Reasons
- Device type
- Spinal Implants
Timeline
| 22 May 2026 | Recall initiated by company | |
| 3 Aug 2026 | Classified by FDA | +73 days |
| 12 Aug 2026 | Published in enforcement report | +9 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
Pedicle screw system intended to provide immobilization/stabilization of spinal segments as adjunct to thoracic/lumbar/sacral spine fusion may have anti-torque instrument with undersized condition which could prevent mating tip from engaging with the pedicle screw tulip and rod during tightening, which may lead to surgical delay, intervention/alternative methods to perform final tightening step.
Product as listed by the FDA
Anti-Torque, Vector, Part: LS-749 part of Vector Pedicle Screw System
Where it was distributed
US Nationwide distribution in the states of TX, CA.
Questions about this recall
Why was it recalled?
Pedicle screw system intended to provide immobilization/stabilization of spinal segments as adjunct to thoracic/lumbar/sacral spine fusion may have anti-torque instrument with undersized condition which could prevent mating tip from engaging with the pedicle screw tulip and rod during tightening, which may lead to surgical delay, intervention/alternative methods to perform final tightening step.
Which lots are affected?
LS-749: UDI-DI: M711BTA2020, M711LS7490, Serial Numbers: 002, 003, 007, 012, 014
Where was it sold?
US Nationwide distribution in the states of TX, CA.
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class II mean?
Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.
Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
More recalls from Innovasis
All 7| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | TxHA PEEK IBF/TxTiHA IBF System recalled by Innovasis | Innovasis | Other | Nationwide |
| Device | CLASS II | Oryx Cervical Screw Caddy recalled by Innovasis | Innovasis | Device Malfunction | Nationwide |
| Device | CLASS II | Excella II Polyaxial Pedicle Screw used as a pedicle screw recalled | Innovasis | Device Malfunction | UT |
| Device | CLASS II | Opteryx Variable Rescue Screw, Part #CP2214 recalled by Innovasis | Innovasis | Foreign Material | Nationwide |
| Device | CLASS II | Excella II Rod recalled by Innovasis | Innovasis | Packaging Defects | 4 states |
Other Spinal Implants recalls
All 1,159| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Extended Polyaxial Head, Attached Gimbal recalled | Seaspine Orthopedics | Other | Nationwide |
| Device | CLASS II | Medline Convenience Kits: Anterior Spine Pack recalled | Medline Industries | Other | Nationwide |
| Device | CLASS II | Neuro; Neuro; Crani; Ortho Spine Pack; Bmhm Neuro Pack recalled | Avid Medical | Sterility | Nationwide |
| Device | CLASS II | Halyard medical convenience kits labeled as: ENT recalled | Avid Medical | Sterility | Nationwide |
| Device | CLASS II | Halyard medical convenience kits labeled as: A-P PACK recalled | Avid Medical | Sterility | Nationwide |