CLASS II ONGOING Device recall · Reported 12 Aug 2026 · FDA Enforcement Report

Anti-Torque, Vector, Part: LS-749 part of Vector Pedicle recalled

Innovasis is recalling Anti-Torque, Vector, Part: LS-749 part of Vector Pedicle Screw System, distributed nationwide in the US. The recall was initiated on 22 May 2026 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesLS-749: UDI-DI: M711BTA2020, M711LS7490, Serial Numbers: 002, 003, 007, 012, 014
Quantity recalled5

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-2859-2026
FDA event ID
99188
Recalling firm
Innovasis, Salt Lake City, UT
Classification
Class II
Status
Ongoing
Recall initiated
22 May 2026
FDA classified
3 Aug 2026
Reported
12 Aug 2026
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Other Reasons
Device type
Spinal Implants

Timeline

22 May 2026Recall initiated by company
3 Aug 2026Classified by FDA+73 days
12 Aug 2026Published in enforcement report+9 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

Pedicle screw system intended to provide immobilization/stabilization of spinal segments as adjunct to thoracic/lumbar/sacral spine fusion may have anti-torque instrument with undersized condition which could prevent mating tip from engaging with the pedicle screw tulip and rod during tightening, which may lead to surgical delay, intervention/alternative methods to perform final tightening step.

Product as listed by the FDA

Anti-Torque, Vector, Part: LS-749 part of Vector Pedicle Screw System

Where it was distributed

US Nationwide distribution in the states of TX, CA.

Nationwide CaliforniaTexas

Questions about this recall

Why was it recalled?

Pedicle screw system intended to provide immobilization/stabilization of spinal segments as adjunct to thoracic/lumbar/sacral spine fusion may have anti-torque instrument with undersized condition which could prevent mating tip from engaging with the pedicle screw tulip and rod during tightening, which may lead to surgical delay, intervention/alternative methods to perform final tightening step.

Which lots are affected?

LS-749: UDI-DI: M711BTA2020, M711LS7490, Serial Numbers: 002, 003, 007, 012, 014

Where was it sold?

US Nationwide distribution in the states of TX, CA.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

More recalls from Innovasis

All 7

Other Spinal Implants recalls

All 1,159