CLASS III ONGOING Device recall · Reported 4 Jun 2025 · FDA Enforcement Report

Apo B Reagent recalled by Beckman Coulter Ireland

Beckman Coulter Ireland is recalling Apo B Reagent, distributed nationwide in the US. The recall was initiated on 23 Apr 2025 and the FDA has classified it as Class III.

Check your package

Lot, batch or date codesLot # 2631/UDI: 15099590010409
Quantity recalled2747 units

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-1815-2025
FDA event ID
96746
Recalling firm
Beckman Coulter Ireland, O'Callaghan'S Mills, N/A
Classification
Class III
Status
Ongoing
Recall initiated
23 Apr 2025
FDA classified
23 May 2025
Reported
4 Jun 2025
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Potency & Assay

Timeline

23 Apr 2025Recall initiated by company
23 May 2025Classified by FDA+30 days
4 Jun 2025Published in enforcement report+12 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

The product Value Assignment Sheet (VAS) is used in calibration of the device. An erroneous value was included in the VAS which has the potential to generate an incorrect APO B patient result which could exceed the total allowable error for the APO B assay (11.6%). The error results in results with a positive bias up to 42%. However, based on the mitigating factors the risk to health is considered to be negligible or limited.

Potency and specification recalls happen when a batch tests outside its approved limits for strength, purity or dissolution, which can mean the dose delivered is higher or lower than labelled.

Product as listed by the FDA

Apo B Reagent, REF: OSR6143

Where it was distributed

Worldwide - US Nationwide distribution in the states of AZ, CA, CO, FL, GA, IL, IN, MD, MI, NJ, NY, OH, OK, OR, PA, TX, VA, WA, WV and the countries of Argentina, Australia, Belarus, Belgium, Bulgaria, Canada, China, Czechia, Egypt, Estonia, France, Germany, Ghana, Greece, Hong Kong, India, Ireland, Israel, Italy, Korea, Republic of, Malaysia, Netherlands, New Zealand, Russian Federation, Saudi Arabia, Slovakia, Spain, Switzerland, Taiwan, Province of China, United Kingdom of Great Britain and Northern Ireland, United States of America, Viet Nam.

Nationwide ArizonaCaliforniaColoradoFloridaGeorgiaIllinoisIndianaMarylandMichiganNew JerseyNew YorkOhioOklahomaOregonPennsylvaniaTexasVirginiaWashingtonWest Virginia

Questions about this recall

Why was it recalled?

The product Value Assignment Sheet (VAS) is used in calibration of the device. An erroneous value was included in the VAS which has the potential to generate an incorrect APO B patient result which could exceed the total allowable error for the APO B assay (11.6%). The error results in results with a positive bias up to 42%. However, based on the mitigating factors the risk to health is considered to be negligible or limited.

Which lots are affected?

Lot # 2631/UDI: 15099590010409

Where was it sold?

Worldwide - US Nationwide distribution in the states of AZ, CA, CO, FL, GA, IL, IN, MD, MI, NJ, NY, OH, OK, OR, PA, TX, VA, WA, WV and the countries of Argentina, Australia, Belarus, Belgium, Bulgaria, Canada, China, Czechia, Egypt, Estonia, France, Germany, Ghana, Greece, Hong Kong, India, Ireland, Israel, Italy, Korea, Republic of, Malaysia, Netherlands, New Zealand, Russian Federation, Saudi Arabia, Slovakia, Spain, Switzerland, Taiwan, Province of China, United Kingdom of Great Britain and Northern Ireland, United States of America, Viet Nam.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class III mean?

Class III recalls are for products unlikely to cause any adverse health consequence, usually labelling or quality issues.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

More recalls from Beckman Coulter Ireland

All 3

Other Test Kits & Diagnostics recalls

All 2,663