CLASS II ONGOING Device recall · Reported 28 Jan 2026 · FDA Enforcement Report
The LDL-Cholesterol assay is comprised of two distinct phases recalled
Beckman Coulter Ireland is recalling The LDL-Cholesterol assay is comprised of two distinct phases, distributed in 51 states. The recall was initiated on 8 Dec 2025 and the FDA has classified it as Class II.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-1135-2026
- FDA event ID
- 98094
- Recalling firm
- Beckman Coulter Ireland, O'Callaghan'S Mills
- Classification
- Class II
- Status
- Ongoing
- Recall initiated
- 8 Dec 2025
- FDA classified
- 21 Jan 2026
- Reported
- 28 Jan 2026
- Type
- Voluntary: Firm initiated
- Customers notified by
- Letter
- Reason category
- Particulate Matter
- Device type
- Test Kits & Diagnostics
Timeline
| 8 Dec 2025 | Recall initiated by company | |
| 21 Jan 2026 | Classified by FDA | +44 days |
| 28 Jan 2026 | Published in enforcement report | +7 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
Beckman Coulter identified that LDL Cholesterol OSR6x96 is not meeting the performance claim listed in the Instructions for Use (IFU). LDL Cholesterol reagent showed clinically significant interference due to triglyceride concentration up to 1500 mg/dL may cause a maximum positive bias of up to 28.8% in low LDL Cholesterol samples (concentration 80 mg/dL) and up to 11.02% in high LDL Cholesterol samples (concentration 160mg/dL).
Product as listed by the FDA
The LDL-Cholesterol assay is comprised of two distinct phases. In phase one a unique detergent solubilizes cholesterol from non-LDL- lipoprotein particles. This cholesterol is consumed by cholesterol esterase, cholesterol oxidase, peroxidase and 4- aminoantipyrine to generate a colorless end product. In phase two a second detergent in reagent 2 releases cholesterol from the LDL lipoproteins. This cholesterol reacts with cholesterol esterase, cholesterol oxidase and a chromogen system to yield a blue color complex which can be measured bichromatically at 540/660nm. The resulting increase in absorbance is directly proportional to the LDL-C concentration in the sample.
Where it was distributed
US: MD, NJ, OH, FL, CA, TX, NC, WA, MA, NV, GA, CO, PA, NM, UT, VA, KS, MT, OK, NY, AZ, LA, OR, AL, MO, IN, IL, KY, MS, NH, AK, SD, MN, TN, PR, MI, IA, WI, WV, ME, SC, HI, RI, ND, NE, ID, AR, WY, DC, GU, DE, CT. OUS: Brazil, South Africa, Argentina, Mexico, Canada, Costa Rica, Peru, El Salvador, Ecuador, Panama, Singapore, Ireland, United Kingdom of Great Britain and Northern Ireland, Germany, China, Korea, Republic of
AlaskaAlabamaArkansasArizonaCaliforniaColoradoConnecticutDistrict of ColumbiaDelawareFloridaGeorgiaHawaiiIowaIdahoIllinoisIndianaKansasKentuckyLouisianaMassachusetts
Questions about this recall
Why was it recalled?
Beckman Coulter identified that LDL Cholesterol OSR6x96 is not meeting the performance claim listed in the Instructions for Use (IFU). LDL Cholesterol reagent showed clinically significant interference due to triglyceride concentration up to 1500 mg/dL may cause a maximum positive bias of up to 28.8% in low LDL Cholesterol samples (concentration 80 mg/dL) and up to 11.02% in high LDL Cholesterol samples (concentration 160mg/dL).
Which lots are affected?
REF/UDI-DI: OSR6196/15099590010928, OSR6296/15099590020361. All Lots and Expiration Dates. IFU Version BAOSR6X96-10 and prior.
Where was it sold?
US: MD, NJ, OH, FL, CA, TX, NC, WA, MA, NV, GA, CO, PA, NM, UT, VA, KS, MT, OK, NY, AZ, LA, OR, AL, MO, IN, IL, KY, MS, NH, AK, SD, MN, TN, PR, MI, IA, WI, WV, ME, SC, HI, RI, ND, NE, ID, AR, WY, DC, GU, DE, CT. OUS: Brazil, South Africa, Argentina, Mexico, Canada, Costa Rica, Peru, El Salvador, Ecuador, Panama, Singapore, Ireland, United Kingdom of Great Britain and Northern Ireland, Germany, China, Korea, Republic of
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class II mean?
Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.
Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
More recalls from Beckman Coulter Ireland
All 3| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Beckman Coulter Bicarbonate recalled | Beckman Coulter Ireland | Other | Nationwide |
| Device | CLASS III | Apo B Reagent recalled by Beckman Coulter Ireland | Beckman Coulter Ireland | Potency & Assay | Nationwide |
Other Test Kits & Diagnostics recalls
All 2,656| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | STA Staclot Protein recalled by Diagnostica Stago | Diagnostica Stago | Potency & Assay | Nationwide |
| Device | CLASS II | DxC 500i Clinical Analyzer Modules using recalled by Beckman Coulter | Beckman Coulter | Packaging Defects | Nationwide |
| Device | CLASS II | DxC 500 AU Clinical Analyzer Modules using recalled by Beckman Coulter | Beckman Coulter | Packaging Defects | Nationwide |
| Device | CLASS II | Syva EMIT 2000 Quinidine Assay recalled | Siemens Healthcare Diagnostics | Potency & Assay | Nationwide |
| Device | CLASS II | Beckman Coulter, Emit 2000 Quinidine Assay recalled | Siemens Healthcare Diagnostics | Potency & Assay | Nationwide |