CLASS II ONGOING Device recall · Reported 28 Jan 2026 · FDA Enforcement Report

The LDL-Cholesterol assay is comprised of two distinct phases recalled

Beckman Coulter Ireland is recalling The LDL-Cholesterol assay is comprised of two distinct phases, distributed in 51 states. The recall was initiated on 8 Dec 2025 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesREF/UDI-DI: OSR6196/15099590010928, OSR6296/15099590020361. All Lots and Expiration Dates. IFU Version BAOSR6X96-10 and prior.
Quantity recalled34,451

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-1135-2026
FDA event ID
98094
Recalling firm
Beckman Coulter Ireland, O'Callaghan'S Mills
Classification
Class II
Status
Ongoing
Recall initiated
8 Dec 2025
FDA classified
21 Jan 2026
Reported
28 Jan 2026
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Particulate Matter

Timeline

8 Dec 2025Recall initiated by company
21 Jan 2026Classified by FDA+44 days
28 Jan 2026Published in enforcement report+7 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

Beckman Coulter identified that LDL Cholesterol OSR6x96 is not meeting the performance claim listed in the Instructions for Use (IFU). LDL Cholesterol reagent showed clinically significant interference due to triglyceride concentration up to 1500 mg/dL may cause a maximum positive bias of up to 28.8% in low LDL Cholesterol samples (concentration 80 mg/dL) and up to 11.02% in high LDL Cholesterol samples (concentration 160mg/dL).

Product as listed by the FDA

The LDL-Cholesterol assay is comprised of two distinct phases. In phase one a unique detergent solubilizes cholesterol from non-LDL- lipoprotein particles. This cholesterol is consumed by cholesterol esterase, cholesterol oxidase, peroxidase and 4- aminoantipyrine to generate a colorless end product. In phase two a second detergent in reagent 2 releases cholesterol from the LDL lipoproteins. This cholesterol reacts with cholesterol esterase, cholesterol oxidase and a chromogen system to yield a blue color complex which can be measured bichromatically at 540/660nm. The resulting increase in absorbance is directly proportional to the LDL-C concentration in the sample.

Where it was distributed

US: MD, NJ, OH, FL, CA, TX, NC, WA, MA, NV, GA, CO, PA, NM, UT, VA, KS, MT, OK, NY, AZ, LA, OR, AL, MO, IN, IL, KY, MS, NH, AK, SD, MN, TN, PR, MI, IA, WI, WV, ME, SC, HI, RI, ND, NE, ID, AR, WY, DC, GU, DE, CT. OUS: Brazil, South Africa, Argentina, Mexico, Canada, Costa Rica, Peru, El Salvador, Ecuador, Panama, Singapore, Ireland, United Kingdom of Great Britain and Northern Ireland, Germany, China, Korea, Republic of

AlaskaAlabamaArkansasArizonaCaliforniaColoradoConnecticutDistrict of ColumbiaDelawareFloridaGeorgiaHawaiiIowaIdahoIllinoisIndianaKansasKentuckyLouisianaMassachusetts

Questions about this recall

Why was it recalled?

Beckman Coulter identified that LDL Cholesterol OSR6x96 is not meeting the performance claim listed in the Instructions for Use (IFU). LDL Cholesterol reagent showed clinically significant interference due to triglyceride concentration up to 1500 mg/dL may cause a maximum positive bias of up to 28.8% in low LDL Cholesterol samples (concentration 80 mg/dL) and up to 11.02% in high LDL Cholesterol samples (concentration 160mg/dL).

Which lots are affected?

REF/UDI-DI: OSR6196/15099590010928, OSR6296/15099590020361. All Lots and Expiration Dates. IFU Version BAOSR6X96-10 and prior.

Where was it sold?

US: MD, NJ, OH, FL, CA, TX, NC, WA, MA, NV, GA, CO, PA, NM, UT, VA, KS, MT, OK, NY, AZ, LA, OR, AL, MO, IN, IL, KY, MS, NH, AK, SD, MN, TN, PR, MI, IA, WI, WV, ME, SC, HI, RI, ND, NE, ID, AR, WY, DC, GU, DE, CT. OUS: Brazil, South Africa, Argentina, Mexico, Canada, Costa Rica, Peru, El Salvador, Ecuador, Panama, Singapore, Ireland, United Kingdom of Great Britain and Northern Ireland, Germany, China, Korea, Republic of

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.