CLASS II Device recall · Reported 19 Jun 2013 · FDA Enforcement Report

Applied Medicals Epix¿ and Direct Drive Laparoscopic Graspers recalled

Applied Medical Resources is recalling Applied Medicals Epix¿ and Direct Drive Laparoscopic Graspers, distributed nationwide in the US. The recall was initiated on 1 May 2013 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesModel Numbers C4130 and C4140. MODEL NUMBERS C4130- 5mm x 35cm Epix/Direct Drive Laparoscopic Grasper LOT NUMBERS 1145332, 1145943, 1145944, 1145945, 1145992, 1145993, 1145994, 1146371, 1147900, 1147901, 1147955, 1148172, 1148173, 1148174, 1148180, 1148181, 1148184, 1148190, 1148191, 1148192, 1148194, 1148201, 1148203, 1148677, 1148765, 1148766, 1148767, 1148835, 1149040, 1149041, 1149690, 1149757, 1150168, 1150169, 1150170, 1150490, 1150634, 1151246, 1151247, 1151248, 1152274, 1152728, 1153332, 1154394, 1154637, 1155135, 1155473, 1155948, 1156004, 1156010, 1156011, 1156012, 1156270, 1156360, 1156452, 1156577, 1156857, 1157218, 1157679, 1157950, 1157951, 1157952, 1157953, 1157954, 1158052, 1158131, 1158216, 1158217, 1158425, 1159491, 1160167, 1161136, 1161137, 1161398, 1161638, 1162112, 1162503, 1162504, 1162588, 1163399, 1163558, 1163559, 1163575, 1164580, 1164581, 1164582, 1164583, 1168124, 1168637, 1168638, 1169117, 1171168, 1171631, 1172170, 1172668, 1173153, 1174995, 1174996, 1176616, 1177407, 1177408, 1178109, 1180295, 1181554, 1182940, 1183774, 1184416, 1184959, 1187241, 1188110, 1188745, 1189773, 1191368, 1191809. C4140- 5mm x 45cm Epix/Direct Drive Laparoscopic Grasper 1145559, 1147851, 1147852, 1147960, 1148210, 1148211, 1149038, 1149039, 1149689, 1149936, 1150488, 1150489, 1151249, 1152594, 1153330, 1155645, 1157217, 1157949, 1158644, 1160033, 1160170, 1162111, 1167871, 1169855, 1171497, 1172009, 1173587, 1174121, 1177578, 1179347, 1180753, 1181346, 1182924, 1183769, 1184961, 1188111, 1189182 Various Kits Containing Either C4130 or C4140 1148079, 1148139, 1148145, 1148250, 1148514, 1148770, 1148780, 1149100, 1149605, 1150455, 1151831, 1153074, 1153702, 1154931, 1154997, 1155332, 1156581, 1156812, 1157090, 1158535, 1160152, 1160227, 1160388, 1161766, 1162189, 1162343, 1163348, 1163881, 1164590, 1164796, 1164831, 1165443, 1166520, 1168227, 1169190, 1169456, 1169620, 1169772, 1170151, 1170378, 1171280, 1171878, 1171879, 1172330, 1173319, 1173903, 1174093, 1175987, 1175991, 1176171, 1178210, 1178498, 1178581, 1179376, 1182679, 1182972, 1183303, 1184673, 1185599, 1185762, 1186257, 1186868, 1187082, 1188827, 1189581, 1190574, 1191055, 1193112, 1193147, 1193648, 1193688, 1194964, 1194965
Quantity recalled127,305 units

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-1521-2013
FDA event ID
65217
Recalling firm
Applied Medical Resources, Rancho Santa Margarita, CA
Classification
Class II
Status
Terminated
Recall initiated
1 May 2013
FDA classified
7 Jun 2013
Reported
19 Jun 2013
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Other Reasons

Timeline

1 May 2013Recall initiated by company
7 Jun 2013Classified by FDA+37 days
19 Jun 2013Published in enforcement report+12 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

Applied Medical is conducting a voluntary recall on specific lot numbers of its Epix¿ and Direct Drive Laparoscopic Graspers. When the ratchet trigger is activated, if a great enough force is applied, there is a potential for the trigger to fracture and the jaws to remain in the closed position. The likelihood of this situation to occur and result in permanent patient injury is highly unlikely;

Product as listed by the FDA

Applied Medicals Epix¿ and Direct Drive Laparoscopic Graspers, Model Numbers C4130 and C4140. Used for grasping and manipulating tissue during general or laparoscopic surgery.

Where it was distributed

Worldwide Distribution: USA Nationwide and countries of: Argentina, Austria, Australia, Bahrain, Belgium, Canada, Chile, Columbia, Denmark, Ecuador, Finland, France, Germany, Greece, India, Ireland, Israel, Italy, Jordan, Kuwait, Malaysia, Lebanon, Lybia, Mexico, Netherlands, New Zealand, Pakistan, Peru, Poland, Portugal, Qatar, Saudi Arabia, Serbia, Singapore, Spain, South Africa, Switzerland, Taiwan-Province of China,Turkey, United Arabs Emirates United Kingdom, Uruguay, and Venezula-Bolivarian Republic.

Nationwide

Questions about this recall

Is Applied Medicals Epix¿ and Direct Drive Laparoscopic Graspers recalled?

Yes. Applied Medical Resources recalled Applied Medicals Epix¿ and Direct Drive Laparoscopic Graspers on 1 May 2013. The recall is Class II and its status is Terminated. Recall number Z-1521-2013.

Why was it recalled?

Applied Medical is conducting a voluntary recall on specific lot numbers of its Epix¿ and Direct Drive Laparoscopic Graspers. When the ratchet trigger is activated, if a great enough force is applied, there is a potential for the trigger to fracture and the jaws to remain in the closed position. The likelihood of this situation to occur and result in permanent patient injury is highly unlikely;

Which lots are affected?

Model Numbers C4130 and C4140. MODEL NUMBERS C4130- 5mm x 35cm Epix/Direct Drive Laparoscopic Grasper LOT NUMBERS 1145332, 1145943, 1145944, 1145945, 1145992, 1145993, 1145994, 1146371, 1147900, 1147901, 1147955, 1148172, 1148173, 1148174, 1148180, 1148181, 1148184, 1148190, 1148191, 1148192, 1148194, 1148201, 1148203, 1148677, 1148765, 1148766, 1148767, 1148835, 1149040, 1149041, 1149690, 1149757, 1150168, 1150169, 1150170, 1150490, 1150634, 1151246, 1151247, 1151248, 1152274, 1152728, 1153332, 1154394, 1154637, 1155135, 1155473, 1155948, 1156004, 1156010, 1156011, 1156012, 1156270, 1156360, 1156452, 1156577, 1156857, 1157218, 1157679, 1157950, 1157951, 1157952, 1157953, 1157954, 1158052, 1158131, 1158216, 1158217, 1158425, 1159491, 1160167, 1161136, 1161137, 1161398, 1161638, 1162112, 1162503, 1162504, 1162588, 1163399, 1163558, 1163559, 1163575, 1164580, 1164581, 1164582, 1164583, 1168124, 1168637, 1168638, 1169117, 1171168, 1171631, 1172170, 1172668, 1173153, 1174995, 1174996, 1176616, 1177407, 1177408, 1178109, 1180295, 1181554, 1182940, 1183774, 1184416, 1184959, 1187241, 1188110, 1188745, 1189773, 1191368, 1191809. C4140- 5mm x 45cm Epix/Direct Drive Laparoscopic Grasper 1145559, 1147851, 1147852, 1147960, 1148210, 1148211, 1149038, 1149039, 1149689, 1149936, 1150488, 1150489, 1151249, 1152594, 1153330, 1155645, 1157217, 1157949, 1158644, 1160033, 1160170, 1162111, 1167871, 1169855, 1171497, 1172009, 1173587, 1174121, 1177578, 1179347, 1180753, 1181346, 1182924, 1183769, 1184961, 1188111, 1189182 Various Kits Containing Either C4130 or C4140 1148079, 1148139, 1148145, 1148250, 1148514, 1148770, 1148780, 1149100, 1149605, 1150455, 1151831, 1153074, 1153702, 1154931, 1154997, 1155332, 1156581, 1156812, 1157090, 1158535, 1160152, 1160227, 1160388, 1161766, 1162189, 1162343, 1163348, 1163881, 1164590, 1164796, 1164831, 1165443, 1166520, 1168227, 1169190, 1169456, 1169620, 1169772, 1170151, 1170378, 1171280, 1171878, 1171879, 1172330, 1173319, 1173903, 1174093, 1175987, 1175991, 1176171, 1178210, 1178498, 1178581, 1179376, 1182679, 1182972, 1183303, 1184673, 1185599, 1185762, 1186257, 1186868, 1187082, 1188827, 1189581, 1190574, 1191055, 1193112, 1193147, 1193648, 1193688, 1194964, 1194965

Where was it sold?

Worldwide Distribution: USA Nationwide and countries of: Argentina, Austria, Australia, Bahrain, Belgium, Canada, Chile, Columbia, Denmark, Ecuador, Finland, France, Germany, Greece, India, Ireland, Israel, Italy, Jordan, Kuwait, Malaysia, Lebanon, Lybia, Mexico, Netherlands, New Zealand, Pakistan, Peru, Poland, Portugal, Qatar, Saudi Arabia, Serbia, Singapore, Spain, South Africa, Switzerland, Taiwan-Province of China,Turkey, United Arabs Emirates United Kingdom, Uruguay, and Venezula-Bolivarian Republic.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 21 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

More recalls from Applied Medical Resources

All 30

Other Surgical Instruments recalls

All 3,809