CLASS II ONGOING Device recall · Reported 30 Dec 2020 · FDA Enforcement Report
(30") Appx, Admin Set, 2 Spiros w/Red Cap recalled by ICU Medical
ICU Medical is recalling (30") Appx, Admin Set, 2 Spiros w/Red Cap, 20 Drop In-Line Drip Chamber w/15 Micron, distributed nationwide in the US. The recall was initiated on 31 Aug 2020 and the FDA has classified it as Class II.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-0708-2021
- FDA event ID
- 86426
- Recalling firm
- ICU Medical, San Clemente, CA
- Classification
- Class II
- Status
- Ongoing
- Recall initiated
- 31 Aug 2020
- FDA classified
- 23 Dec 2020
- Reported
- 30 Dec 2020
- Type
- Voluntary: Firm initiated
- Customers notified by
- Letter
- Reason category
- Packaging Defects
- Device type
- Masks, Gloves & PPE
Timeline
| 31 Aug 2020 | Recall initiated by company | |
| 23 Dec 2020 | Classified by FDA | +114 days |
| 30 Dec 2020 | Published in enforcement report | +7 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
Potential for spinning versions of male luers to exhibit leaks due to a molding defect. Fluid leakage may potentially cause delay of infusion, contamination of the fluid path, exposure to hazardous medications, blood loss, exposure to patient blood, or fluid path air-in-line.
Packaging recalls include wrong labels, wrong products in the wrong bottle, missing leaflets, short-filled or empty capsules and seals that fail.
Product as listed by the FDA
76 cm (30") Appx 3.3 ml, Admin Set, 2 Spiros w/Red Cap, 20 Drop In-Line Drip Chamber w/15 Micron Filter, Bag Hanger, REF: 011-CH3261; 44 cm(17 IN) Appx 2.9 ml,PUR YELLOW TRANS SET,ChemoClave,CK VLV,Spiros,PURPLE CAP, REF: 011-CH3568; Oncology Kit, 5 Gang Stopcock w/5 ChemoClave, Spiros w/Red Cap; 10cm Smallbore Bifuse Ext w/2 ChemoClave, Red Ring, BCV; 297cm Ext w/ChemoClave, REF: 011-CH3600; Oncology Kit, 5 Gang Stopcock w/5 ChemoClave, Spiros w/Red Cap; 10cm Smallbore Bifuse Ext, w/2 ChemoClave, Red Ring, BCV; 597 cm Ext w/ChemoClave, REF: 011-CH3601; Kit, 2 5 Gang Stopcocks w/5 MicroClave Clear, Spiros; 2 Smallbore Bifuse Ext Sets, 2 MicroClave Clear, Red Ring, Check Valve, RMLL; 2 Ext Sets, MicroClave Clear Clipped Together, RMLL, REF: 011-CH3610; Admin Set w/Clave, Spiros, REF: 011-CH3772; Admin Set w/Clave, Spiros, REF: 011-CH3773; Admin Set w/Clave, Spiros, REF: 011-CH3967; 76" (193 cm) 50 ml, Diluent Set for Channel 2 w/20 Drop Drip Chamber, Spiros, Drop-In Red Cap, REF: 011-CH4011
Where it was distributed
Worldwide distribution - U.S. Nationwide distribution including in the states of FL, MD, LA, PA, IN, TX, NY, MO, VA, OH, GA, NC, IL, CA, AZ, AK, NE, DE, WI, CO, UT, ND, MS, MA, MN, WA, SC, OR, KS, MI, NH, NJ, KY, TN, ID, AR, CT, OK, AL, SD, ME, IA, VT, WY, DC, NV, RI, MT, NM, HI, WV and the countries of Italy, Belgium, United Kingdom, Canada, Germany, Argentina, Australia, Poland, Greece, Switzerland, Israel, France, Finland, Spain, Malaysia, Taiwan, Ireland, Luxembourg, Hong Kong , Brazil, Netherlands, Austria, Sweden, Turkey, Singapore, Slovakia, Croatia, South Africa, Romania, Thailand.
Nationwide AlaskaAlabamaArkansasArizonaCaliforniaColoradoConnecticutDistrict of ColumbiaDelawareFloridaGeorgiaHawaiiIowaIdahoIllinoisIndianaKansasKentuckyLouisianaMassachusetts
Questions about this recall
Why was it recalled?
Potential for spinning versions of male luers to exhibit leaks due to a molding defect. Fluid leakage may potentially cause delay of infusion, contamination of the fluid path, exposure to hazardous medications, blood loss, exposure to patient blood, or fluid path air-in-line.
Which lots are affected?
Lots: 4784390, 4784404, 4865931, 4865006, 4590283, 4774210, 4865952, 4711264, 4605516, 4723103, 4882016, 4806569 ,
Where was it sold?
Worldwide distribution - U.S. Nationwide distribution including in the states of FL, MD, LA, PA, IN, TX, NY, MO, VA, OH, GA, NC, IL, CA, AZ, AK, NE, DE, WI, CO, UT, ND, MS, MA, MN, WA, SC, OR, KS, MI, NH, NJ, KY, TN, ID, AR, CT, OK, AL, SD, ME, IA, VT, WY, DC, NV, RI, MT, NM, HI, WV and the countries of Italy, Belgium, United Kingdom, Canada, Germany, Argentina, Australia, Poland, Greece, Switzerland, Israel, France, Finland, Spain, Malaysia, Taiwan, Ireland, Luxembourg, Hong Kong , Brazil, Netherlands, Austria, Sweden, Turkey, Singapore, Slovakia, Croatia, South Africa, Romania, Thailand.
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class II mean?
Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.
Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
Other products in this recall event
12 other products · FDA event 86426This product is part of a recall event; the main page for the event is 30" Appx, 20 Drop Admin Set w/Integrated ChemoLock Drip recalled.
More recalls from ICU Medical
All 663| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | CADD Yellow Medication Cassette Reservoir with NRFit connector US recalled | ICU Medical | Packaging Defects | Nationwide |
| Device | CLASS II | IV Sets with drip chambers recalled by ICU Medical | ICU Medical | Particulate Matter | Nationwide |
| Device | CLASS II | IV Sets with drip chambers recalled by ICU Medical | ICU Medical | Particulate Matter | Nationwide |
| Device | CLASS II | IV Non-Dedicated Gravity Administration Sets with drip chambers recalled | ICU Medical | Foreign Material | Nationwide |
| Device | CLASS II | IV Sets with drip chambers recalled by ICU Medical | ICU Medical | Particulate Matter | Nationwide |
Other Masks, Gloves & PPE recalls
All 450| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Opportunity Pro recalled by American Nitrile Operations | American Nitrile Operations | Device Malfunction | Nationwide |
| Device | CLASS II | Opportunity Flex recalled by American Nitrile Operations | American Nitrile Operations | Device Malfunction | Nationwide |
| Device | CLASS II | Performance Pro recalled by American Nitrile Operations | American Nitrile Operations | Device Malfunction | Nationwide |
| Device | CLASS II | P28 Phanton Intramedullary Nail System recalled by Paragon 28 | Paragon 28 | Sterility | Nationwide |
| Device | CLASS II | Cyberknife Treatment Deliver System equipped recalled | Accuray | Device Malfunction | Nationwide |