CLASS II Device recall · Reported 9 Feb 2022 · FDA Enforcement Report
Aries HSV 1&2 Assay recalled by Luminex
Luminex is recalling Aries HSV 1&2 Assay, distributed nationwide in the US. The recall was initiated on 17 Nov 2021 and the FDA has classified it as Class II.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-0553-2022
- FDA event ID
- 89244
- Recalling firm
- Luminex, Austin, TX
- Classification
- Class II
- Status
- Terminated
- Recall initiated
- 17 Nov 2021
- FDA classified
- 2 Feb 2022
- Reported
- 9 Feb 2022
- Type
- Voluntary: Firm initiated
- Customers notified by
- Reason category
- Packaging Defects
- Device type
- Test Kits & Diagnostics
Timeline
| 17 Nov 2021 | Recall initiated by company | |
| 2 Feb 2022 | Classified by FDA | +77 days |
| 9 Feb 2022 | Published in enforcement report | +7 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
Impacted lot may leak inside the ARIES instrument. Testing confirmed a leak rate of 0.9%. Leaking cassettes could potentially lead to incorrect results or a delay in reporting results due to an Invalid.
Packaging recalls include wrong labels, wrong products in the wrong bottle, missing leaflets, short-filled or empty capsules and seals that fail.
Product as listed by the FDA
ARIES HSV 1&2 Assay, Catalog/REF# 50-10017
Where it was distributed
Worldwide distribution - US Nationwide distribution in the states of New Jersey, California, Connecticut, Minnesota, Missouri, New Jersey, New Mexico, Pennsylvania, Tennessee, Washington, and Wisconsin) and the country of Malaysia.
Nationwide CaliforniaConnecticutMinnesotaMissouriNew JerseyNew MexicoPennsylvaniaTennesseeWashingtonWisconsin
Questions about this recall
Why was it recalled?
Impacted lot may leak inside the ARIES instrument. Testing confirmed a leak rate of 0.9%. Leaking cassettes could potentially lead to incorrect results or a delay in reporting results due to an Invalid.
Which lots are affected?
Lot Numbers AB3600A and AB3560A
Where was it sold?
Worldwide distribution - US Nationwide distribution in the states of New Jersey, California, Connecticut, Minnesota, Missouri, New Jersey, New Mexico, Pennsylvania, Tennessee, Washington, and Wisconsin) and the country of Malaysia.
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class II mean?
Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.
Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
More recalls from Luminex
All 39| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Verigene Gram Negative Blood Culture (Bc-gn) Nucleic Acid Test recalled | Luminex | Other | Nationwide |
| Device | CLASS II | Verigene Gram Positive Blood Culture (Bc-gp) Nucleic Acid Test recalled | Luminex | Other Bacterial Contamination | Nationwide |
| Device | CLASS II | Verigene Bc-gp Utility Tray Carrier) recalled by Luminex | Luminex | Device Malfunction | Nationwide |
| Device | CLASS II | Verigene Enteric Pathogens Nucleic Acid Test (EP) recalled by Luminex | Luminex | Other | Nationwide |
| Device | CLASS II | Verigene Clostridium difficile Nucleic Acid Test recalled by Luminex | Luminex | Other Bacterial Contamination | Nationwide |
Other Test Kits & Diagnostics recalls
All 2,661| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | McKesson Consult hCG Urine Cassette recalled by Alere San Diego | Alere San Diego | Potency & Assay | Nationwide |
| Device | CLASS II | Systems/units containing affected recalled | Beckman Coulter Mishima K.K | Device Malfunction | Nationwide |
| Device | CLASS II | Fisher Scientific Sure-Vue Urine hCG recalled by Alere San Diego | Alere San Diego | Potency & Assay | Nationwide |
| Device | CLASS II | Vitek 2 Ast-N461 Test Kit, Gram Negative Susceptibility recalled | Biomerieux | Other | MO |
| Device | CLASS II | DASH Rapid PCR Instrument recalled by Nuclein | Nuclein | Device Malfunction | Nationwide |