CLASS II ONGOING Device recall · Reported 15 Jun 2022 · FDA Enforcement Report

Artis icono biplane -angiography systems developed recalled

Siemens Medical Solutions USA is recalling Artis icono biplane -angiography systems developed for single and biplane diagnostic, distributed nationwide in the US. The recall was initiated on 11 Apr 2022 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesUDI-DI: 04056869063317 Serial Number: 180062 180054 180019 180065 180056 180057 180041 180317 180034 180072 180060 180070 180048 180083 180315 180080 180301 180305 180067 180330 180040 180333 180029 180336 180316 180084 180334 180018 180046 180047 180335 180085 180035 180081 180331 180021 180321 180309 180323 180329 180043 180049 180077 180086 180053 180055 180015 180014 180066 180082 180037
Quantity recalled51 units

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-1245-2022
FDA event ID
90184
Recalling firm
Siemens Medical Solutions USA, Malvern, PA
Classification
Class II
Status
Ongoing
Recall initiated
11 Apr 2022
FDA classified
7 Jun 2022
Reported
15 Jun 2022
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Device Malfunction

Timeline

11 Apr 2022Recall initiated by company
7 Jun 2022Classified by FDA+57 days
15 Jun 2022Published in enforcement report+8 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

if the protective earth is interrupted and an additional fault leads to a fault current, the detection of the fault current may not be ensured may cause an electric shock if a person is in contact with metallic parts of the Carm

Device recalls are issued when a component, software or design fault could cause the device to fail or give wrong readings. Many are corrections applied in the field rather than physical returns.

Product as listed by the FDA

Artis icono biplane -angiography systems developed for single and biplane diagnostic imaging and interventional procedures SMN: 11327600

Where it was distributed

US Nationwide distribution.

Nationwide

Questions about this recall

Is Artis icono biplane -angiography systems developed for single and biplane diagnostic recalled?

Yes. Siemens Medical Solutions USA recalled Artis icono biplane -angiography systems developed for single and biplane diagnostic on 11 Apr 2022. The recall is Class II and its status is Ongoing. Recall number Z-1245-2022.

Why was it recalled?

if the protective earth is interrupted and an additional fault leads to a fault current, the detection of the fault current may not be ensured may cause an electric shock if a person is in contact with metallic parts of the Carm

Which lots are affected?

UDI-DI: 04056869063317 Serial Number: 180062 180054 180019 180065 180056 180057 180041 180317 180034 180072 180060 180070 180048 180083 180315 180080 180301 180305 180067 180330 180040 180333 180029 180336 180316 180084 180334 180018 180046 180047 180335 180085 180035 180081 180331 180021 180321 180309 180323 180329 180043 180049 180077 180086 180053 180055 180015 180014 180066 180082 180037

Where was it sold?

US Nationwide distribution.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 21 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.