CLASS I Device recall · Reported 23 Nov 2022 · FDA Enforcement Report

18239: ASM Omnipod Dash PDM, insulin delivery system recalled by Insulet

Insulet is recalling 18239: ASM Omnipod Dash PDM, insulin delivery system, distributed nationwide in the US. The recall was initiated on 17 Oct 2022 and the FDA has classified it as Class I.

Check your package

Lot, batch or date codesPart Number/UDI-DI 18239 : 10385082000009 PT 000030: 10385082000122 PT 000010: 10385082000139 PT 000011: 10385082000146 All Serial Numbers and Lots
Quantity recalled248,288 US; 167,802 OUS

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-0154-2023
FDA event ID
90987
Recalling firm
Insulet, Acton, MA
Classification
Class I
Status
Terminated
Recall initiated
17 Oct 2022
FDA classified
15 Nov 2022
Reported
23 Nov 2022
Type
Voluntary: Firm initiated
Customers notified by
Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Reason category
Packaging Defects

Timeline

17 Oct 2022Recall initiated by company
15 Nov 2022Classified by FDA+29 days
23 Nov 2022Published in enforcement report+8 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

The firm has become aware of PDM battery issues, including battery swelling, fluid leaking from the battery, and extreme overheating, which may pose a fire hazard. The firm has received a complaint alleging fire as a result of thermal energy from a DASH PDM.

Packaging recalls include wrong labels, wrong products in the wrong bottle, missing leaflets, short-filled or empty capsules and seals that fail.

Product as listed by the FDA

18239: ASM Omnipod Dash PDM, insulin delivery system. PT-000010: Assembly, Dash Final PDM U100, mg/dL PT-000011: Assembly, Dash Final PDM U100, mmol/L PT-000030: Assembly, DASH PDM, Canada Remote Controller (Personal Diabetes Manager or PDM) associated with Insulin Pump (Pod) *The part numbers above may be incorporated into higher level assemblies such as DASH Starter Kits.

Where it was distributed

Worldwide Distribution: US nationwide, Canada, Europe, Australia, and the Middle East.

Nationwide

Questions about this recall

Why was it recalled?

The firm has become aware of PDM battery issues, including battery swelling, fluid leaking from the battery, and extreme overheating, which may pose a fire hazard. The firm has received a complaint alleging fire as a result of thermal energy from a DASH PDM.

Which lots are affected?

Part Number/UDI-DI 18239 : 10385082000009 PT 000030: 10385082000122 PT 000010: 10385082000139 PT 000011: 10385082000146 All Serial Numbers and Lots

Where was it sold?

Worldwide Distribution: US nationwide, Canada, Europe, Australia, and the Middle East.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class I mean?

Class I is the most serious FDA recall category: there is a reasonable probability that using the product will cause serious health consequences or death.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

More recalls from Insulet

All 21

Other Insulin Pumps & Glucose Monitors recalls

All 467