CLASS II ONGOING Device recall · Reported 27 Mar 2024 · FDA Enforcement Report

Assembly, Array Guide Tmini recalled by Think Surgical

Think Surgical is recalling Assembly, Array Guide Tmini, distributed nationwide in the US. The recall was initiated on 12 Feb 2024 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesUDI-DI: M9461091010. Lot Numbers: 2316701, 2232601, 23030, 2228501, 2327802, 22152
Quantity recalled16

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-1318-2024
FDA event ID
94086
Recalling firm
Think Surgical, Fremont, CA
Classification
Class II
Status
Ongoing
Recall initiated
12 Feb 2024
FDA classified
20 Mar 2024
Reported
27 Mar 2024
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Other Reasons

Timeline

12 Feb 2024Recall initiated by company
20 Mar 2024Classified by FDA+37 days
27 Mar 2024Published in enforcement report+7 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

There is a galling effect (cold-weld) caused by friction between the array fixation pins and the TMINI array guide holes due to insufficient clearance between the outer diameter of the fixation pins and the inner diameter of the TMINI array guide holes. As a result, the array guide has the possibility of binding to the array fixation pins during drilling.

Product as listed by the FDA

Assembly, Array Guide TMINI, REF: 109101, part of the TMINI Knee Instrument Set, used with the TMINI Miniature Robotic System

Where it was distributed

US Nationwide distribution in the states of AL, OH, NV, TX, UT, CA, MO.

Nationwide AlabamaCaliforniaMissouriNevadaOhioTexasUtah

Questions about this recall

Why was it recalled?

There is a galling effect (cold-weld) caused by friction between the array fixation pins and the TMINI array guide holes due to insufficient clearance between the outer diameter of the fixation pins and the inner diameter of the TMINI array guide holes. As a result, the array guide has the possibility of binding to the array fixation pins during drilling.

Which lots are affected?

UDI-DI: M9461091010. Lot Numbers: 2316701, 2232601, 23030, 2228501, 2327802, 22152

Where was it sold?

US Nationwide distribution in the states of AL, OH, NV, TX, UT, CA, MO.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

More recalls from Think Surgical