CLASS II ONGOING Device recall · Reported 27 Jul 2022 · FDA Enforcement Report
Attune Measured Sizing and Rotation Guide-to size the femur recalled
DePuy Orthopaedics is recalling Attune Measured Sizing and Rotation Guide-to size the femur and align rotation of the, distributed nationwide in the US. The recall was initiated on 29 Jun 2022 and the FDA has classified it as Class II.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-1419-2022
- FDA event ID
- 90557
- Recalling firm
- DePuy Orthopaedics, Warsaw, IN
- Classification
- Class II
- Status
- Ongoing
- Recall initiated
- 29 Jun 2022
- FDA classified
- 20 Jul 2022
- Reported
- 27 Jul 2022
- Type
- Voluntary: Firm initiated
- Customers notified by
- Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
- Reason category
- Foreign Material
- Device type
- Hip & Knee Implants
Timeline
| 29 Jun 2022 | Recall initiated by company | |
| 20 Jul 2022 | Classified by FDA | +21 days |
| 27 Jul 2022 | Published in enforcement report | +7 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
Manufactured with a Galvanized High Carbon Steel spring instead of the correct 302 Stainless steel spring. has the potential to oxidize and corrode in the ATTUNE Measured Sizing & Rotation Guide, resulting in surgical delay, debris in surgical field, Osteolysis, Poor Joint Mechanics, Soft Tissue Irritation, Pain
Foreign material recalls cover pieces of metal, plastic, glass, wood or other objects that got into the product during processing and could cause injury if eaten.
Product as listed by the FDA
ATTUNE Measured Sizing and Rotation Guide-to size the femur and align rotation of the bone cuts in preparation for the femoral implant Part Number: 254400525
Where it was distributed
Nationwide Foreign: Argentina, Australia, Austria, Bahrain, Belgium, Brazil, Canada, Chile, China, Croatia, Czech Republic, Denmark, Finland, France, Germany, Greece, Hong Kong, India, Ireland, Israel, Italy, Japan, Kuwait, Lithuania, Mauritius, Mexico, Morocco, Netherlands, New Zealand, Norway, Peru, Poland, Portugal, Puerto Rico, Romania, Saudi Arabia, Serbia, Singapore, South Africa, South Korea, Spain, Switzerland, Taiwan, Thailand, United Arab Emirates, United Kingdom, Vietnam
Questions about this recall
Is Attune Measured Sizing and Rotation Guide-to size the femur and align rotation of the recalled?
Yes. DePuy Orthopaedics recalled Attune Measured Sizing and Rotation Guide-to size the femur and align rotation of the on 29 Jun 2022. The recall is Class II and its status is Ongoing. Recall number Z-1419-2022.
Why was it recalled?
Manufactured with a Galvanized High Carbon Steel spring instead of the correct 302 Stainless steel spring. has the potential to oxidize and corrode in the ATTUNE Measured Sizing & Rotation Guide, resulting in surgical delay, debris in surgical field, Osteolysis, Poor Joint Mechanics, Soft Tissue Irritation, Pain
Which lots are affected?
GTIN: 10603295434061 Lots: AB4536235, AB4733296, AB4841493, AB4810808, AB4841495, AB4904565, AB4967214, AB5004557, and AB5034303
Where was it sold?
Nationwide Foreign: Argentina, Australia, Austria, Bahrain, Belgium, Brazil, Canada, Chile, China, Croatia, Czech Republic, Denmark, Finland, France, Germany, Greece, Hong Kong, India, Ireland, Israel, Italy, Japan, Kuwait, Lithuania, Mauritius, Mexico, Morocco, Netherlands, New Zealand, Norway, Peru, Poland, Portugal, Puerto Rico, Romania, Saudi Arabia, Serbia, Singapore, South Africa, South Korea, Spain, Switzerland, Taiwan, Thailand, United Arab Emirates, United Kingdom, Vietnam
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class II mean?
Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.
Read the official FDA notice · Source record retrieved 21 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
More recalls from DePuy Orthopaedics
All 153| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Attune Affixium Cementless Fixed Bearing Knee recalled | DePuy Orthopaedics | Labeling Errors | Nationwide |
| Device | CLASS II | Attune Posterior (PS) Fixed Bearing (FB) Tibial Insert SZ5 recalled | DePuy Orthopaedics | Other | Nationwide |
| Device | CLASS II | Attune Revision Limb Preservation System (LPS) Tibial Insert XSM recalled | DePuy Orthopaedics | Other | Nationwide |
| Device | CLASS II | Attune Revision Limb Preservation System (LPS) Tibial Insert XSM recalled | DePuy Orthopaedics | Other | Nationwide |
| Device | CLASS II | Attune Revision Limb Preservation System (LPS) Tibial Insert SM recalled | DePuy Orthopaedics | Other | Nationwide |
Other Hip & Knee Implants recalls
All 3,097| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Medline convenience kits labeled as: Femoral Pack, Medline Kit recalled | Medline Industries | Other | Nationwide |
| Device | CLASS II | Arvis Surgical Navigation System Gen recalled by Kico Knee Innovation | Kico Knee Innovation | Nitrosamine Impurity | 19 states |
| Device | CLASS II | Veritas Standard Reverse Total Shoulder Veritas Standard recalled | restor3d | Device Malfunction | Nationwide |
| Device | CLASS II | Endo-Model, Replacement Set, Tibial Plateau, right/left, L, Uhmwpe recalled | Waldemar Link GmbH & Co. KG (Mfg Site) | Lead & Heavy Metals | IN |
| Device | CLASS II | U2 Total Knee System Femoral offset bushing recalled | United Orthopedic | Device Malfunction | Nationwide |