CLASS II ONGOING Device recall · Reported 27 Jul 2022 · FDA Enforcement Report

Attune Measured Sizing and Rotation Guide-to size the femur recalled

DePuy Orthopaedics is recalling Attune Measured Sizing and Rotation Guide-to size the femur and align rotation of the, distributed nationwide in the US. The recall was initiated on 29 Jun 2022 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesGTIN: 10603295434061 Lots: AB4536235, AB4733296, AB4841493, AB4810808, AB4841495, AB4904565, AB4967214, AB5004557, and AB5034303
Quantity recalled872 units

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-1419-2022
FDA event ID
90557
Recalling firm
DePuy Orthopaedics, Warsaw, IN
Classification
Class II
Status
Ongoing
Recall initiated
29 Jun 2022
FDA classified
20 Jul 2022
Reported
27 Jul 2022
Type
Voluntary: Firm initiated
Customers notified by
Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Reason category
Foreign Material

Timeline

29 Jun 2022Recall initiated by company
20 Jul 2022Classified by FDA+21 days
27 Jul 2022Published in enforcement report+7 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

Manufactured with a Galvanized High Carbon Steel spring instead of the correct 302 Stainless steel spring. has the potential to oxidize and corrode in the ATTUNE Measured Sizing & Rotation Guide, resulting in surgical delay, debris in surgical field, Osteolysis, Poor Joint Mechanics, Soft Tissue Irritation, Pain

Foreign material recalls cover pieces of metal, plastic, glass, wood or other objects that got into the product during processing and could cause injury if eaten.

Product as listed by the FDA

ATTUNE Measured Sizing and Rotation Guide-to size the femur and align rotation of the bone cuts in preparation for the femoral implant Part Number: 254400525

Where it was distributed

Nationwide Foreign: Argentina, Australia, Austria, Bahrain, Belgium, Brazil, Canada, Chile, China, Croatia, Czech Republic, Denmark, Finland, France, Germany, Greece, Hong Kong, India, Ireland, Israel, Italy, Japan, Kuwait, Lithuania, Mauritius, Mexico, Morocco, Netherlands, New Zealand, Norway, Peru, Poland, Portugal, Puerto Rico, Romania, Saudi Arabia, Serbia, Singapore, South Africa, South Korea, Spain, Switzerland, Taiwan, Thailand, United Arab Emirates, United Kingdom, Vietnam

Nationwide Puerto Rico

Questions about this recall

Is Attune Measured Sizing and Rotation Guide-to size the femur and align rotation of the recalled?

Yes. DePuy Orthopaedics recalled Attune Measured Sizing and Rotation Guide-to size the femur and align rotation of the on 29 Jun 2022. The recall is Class II and its status is Ongoing. Recall number Z-1419-2022.

Why was it recalled?

Manufactured with a Galvanized High Carbon Steel spring instead of the correct 302 Stainless steel spring. has the potential to oxidize and corrode in the ATTUNE Measured Sizing & Rotation Guide, resulting in surgical delay, debris in surgical field, Osteolysis, Poor Joint Mechanics, Soft Tissue Irritation, Pain

Which lots are affected?

GTIN: 10603295434061 Lots: AB4536235, AB4733296, AB4841493, AB4810808, AB4841495, AB4904565, AB4967214, AB5004557, and AB5034303

Where was it sold?

Nationwide Foreign: Argentina, Australia, Austria, Bahrain, Belgium, Brazil, Canada, Chile, China, Croatia, Czech Republic, Denmark, Finland, France, Germany, Greece, Hong Kong, India, Ireland, Israel, Italy, Japan, Kuwait, Lithuania, Mauritius, Mexico, Morocco, Netherlands, New Zealand, Norway, Peru, Poland, Portugal, Puerto Rico, Romania, Saudi Arabia, Serbia, Singapore, South Africa, South Korea, Spain, Switzerland, Taiwan, Thailand, United Arab Emirates, United Kingdom, Vietnam

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 21 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

More recalls from DePuy Orthopaedics

All 153

Other Hip & Knee Implants recalls

All 3,097