CLASS II ONGOING Device recall · Reported 23 Sep 2026 · FDA Enforcement Report

Scorpio Recessed Metal Backed Patella 3045 recalled

Howmedica Osteonics is recalling Scorpio Recessed Metal Backed Patella 3045-0026 3045-0028 3045-0030 3045-0032 Orthopaedic, distributed nationwide in the US. The recall was initiated on 30 Jul 2026 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codes3045-0026 No UDI Lots 12WAHA 12WAHB HOHE HOVX 3045-0028 No UDI Lots 13VAW 13VAWA 3045-0030 No UDI Lots BIYIE HNZY 3045-0032 No UDI Lots BIYIH HOYK
Quantity recalled12,156 US

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-3203-2026
FDA event ID
99558
Recalling firm
Howmedica Osteonics, Mahwah, NJ
Classification
Class II
Status
Ongoing
Recall initiated
30 Jul 2026
FDA classified
16 Sep 2026
Reported
23 Sep 2026
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Foreign Material

Timeline

30 Jul 2026Recall initiated by company
16 Sep 2026Classified by FDA+48 days
23 Sep 2026Published in enforcement report+7 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

Affected devices were manufactured with Ultra High Molecular Weight Polyethylene (UHMWPE) raw material aged in excess of five years. Product has the potential for elevated levels of oxidation, which may result in increased wear, leading to hazards such as excessive wear debris, material fragments, fractured device, or disassociated device. Potential harms include revision surgery, pain, and inflammation.

Foreign material recalls cover pieces of metal, plastic, glass, wood or other objects that got into the product during processing and could cause injury if eaten.

Product as listed by the FDA

SCORPIO RECESSED METAL BACKED PATELLA 3045-0026 3045-0028 3045-0030 3045-0032 Orthopaedic implant components manufactured from UHMWPE (polyethylene)

Where it was distributed

Worldwide distribution - US Nationwide and the countries of Canada, LATAM, Europe, Asia Pacific, and South Africa.

Nationwide

Questions about this recall

Why was it recalled?

Affected devices were manufactured with Ultra High Molecular Weight Polyethylene (UHMWPE) raw material aged in excess of five years. Product has the potential for elevated levels of oxidation, which may result in increased wear, leading to hazards such as excessive wear debris, material fragments, fractured device, or disassociated device. Potential harms include revision surgery, pain, and inflammation.

Which lots are affected?

3045-0026 No UDI Lots 12WAHA 12WAHB HOHE HOVX 3045-0028 No UDI Lots 13VAW 13VAWA 3045-0030 No UDI Lots BIYIE HNZY 3045-0032 No UDI Lots BIYIH HOYK

Where was it sold?

Worldwide distribution - US Nationwide and the countries of Canada, LATAM, Europe, Asia Pacific, and South Africa.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 29 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

Other products in this recall event

11 other products · FDA event 99558

This product is part of a recall event; the main page for the event is Scorpio C-dome Patella 73-2710 73-2910 Scorpio U-dome recalled.

More recalls from Howmedica Osteonics

All 244

Other Hip & Knee Implants recalls

All 3,109