CLASS I ONGOING Device recall · Reported 23 Apr 2025 · FDA Enforcement Report

AutoPulse NXT Resuscitation System AutoPulse NXT Platform recalled

Zoll Circulation is recalling AutoPulse NXT Resuscitation System AutoPulse NXT Platform, distributed in 18 states. The recall was initiated on 4 Mar 2025 and the FDA has classified it as Class I.

Check your package

Lot, batch or date codesUDI: (01) 00849111003384/Serial number: 1179 1168 1178 1184 1188 1197 1109 1110 1126 1166 1169 1152 1199 1202 1111 1185 1198 1203 1206 1208 1211 1212 1172 1182 1130 1128 1129 1189 1183 1131 1134 1191 1186 1192 1170 1173 1175 1180 1174 1176 1187 1177 1116 1165 1118 1119 1193 1138 1139 1145 1146 1147 1148 1149 1150 1153 1155 1159 1161 1162 1163 1164 1181 1141 1108 1190 1195 1120 1121 1122 1123 1107 1112 1113 1114 1115 1137 1140 1196
Quantity recalled91

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-1533-2025
FDA event ID
96479
Recalling firm
Zoll Circulation, San Jose, CA
Classification
Class I
Status
Ongoing
Recall initiated
4 Mar 2025
FDA classified
11 Apr 2025
Reported
23 Apr 2025
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Device Malfunction
Device type
Medical Software

Timeline

4 Mar 2025Recall initiated by company
11 Apr 2025Classified by FDA+38 days
23 Apr 2025Published in enforcement report+12 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

AutoPulse NXT Resuscitation System may not provide adequate chest compressions or may stop compressions due to code FC1060.

Device recalls are issued when a component, software or design fault could cause the device to fail or give wrong readings. Many are corrections applied in the field rather than physical returns.

Product as listed by the FDA

Brand Name: AutoPulse NXT Resuscitation System Product Name: AutoPulse NXT Platform Model/Catalog Number: Model 200 (Catalog Number: 8700-001070-1) Software Version: Version 2.0.1 Product Description: The AutoPulse Resuscitation System Model 200 is an automated, portable, battery powered device that compresses the chest of an adult human as an adjunct to manual CPR. Component: N/A

Where it was distributed

US: AL, CA, CT, FL, IL, KS, LA, MN, MO, MT, NC, ND, NM, NY, OH, OK< PA, TX, WI OUS: Canada, Hong Kong,

AlabamaCaliforniaConnecticutFloridaIllinoisKansasLouisianaMinnesotaMissouriMontanaNorth CarolinaNorth DakotaNew MexicoNew YorkOhioPennsylvaniaTexasWisconsin

Questions about this recall

Why was it recalled?

AutoPulse NXT Resuscitation System may not provide adequate chest compressions or may stop compressions due to code FC1060.

Which lots are affected?

UDI: (01) 00849111003384/Serial number: 1179 1168 1178 1184 1188 1197 1109 1110 1126 1166 1169 1152 1199 1202 1111 1185 1198 1203 1206 1208 1211 1212 1172 1182 1130 1128 1129 1189 1183 1131 1134 1191 1186 1192 1170 1173 1175 1180 1174 1176 1187 1177 1116 1165 1118 1119 1193 1138 1139 1145 1146 1147 1148 1149 1150 1153 1155 1159 1161 1162 1163 1164 1181 1141 1108 1190 1195 1120 1121 1122 1123 1107 1112 1113 1114 1115 1137 1140 1196

Where was it sold?

US: AL, CA, CT, FL, IL, KS, LA, MN, MO, MT, NC, ND, NM, NY, OH, OK< PA, TX, WI OUS: Canada, Hong Kong,

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class I mean?

Class I is the most serious FDA recall category: there is a reasonable probability that using the product will cause serious health consequences or death.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

More recalls from Zoll Circulation

All 33

Other Medical Software recalls

All 938