CLASS II Device recall · Reported 7 Nov 2012 · FDA Enforcement Report

Aviator Drill Bit recalled by Stryker Spine

Stryker Spine is recalling Aviator Drill Bit, distributed nationwide in the US. The recall was initiated on 5 Jul 2012 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesE207315 Catalog number 48770610, 48770612, 48770616 All lots
Quantity recalled6154 units

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-0163-2013
FDA event ID
63272
Recalling firm
Stryker Spine, Allendale, NJ
Classification
Class II
Status
Terminated
Recall initiated
5 Jul 2012
FDA classified
29 Oct 2012
Reported
7 Nov 2012
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Other Reasons
Device type
Spinal Implants

Timeline

5 Jul 2012Recall initiated by company
29 Oct 2012Classified by FDA+116 days
7 Nov 2012Published in enforcement report+9 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

Aviator drill bits are 10 mm longer than the Reflex Hybrid drill bits and are not interchangeable which could lead to over-drilling by as much as 10 mm.

Product as listed by the FDA

Aviator Drill Bit (10 mm, 12 mm, 14 mm, 16 mm, Manufactured by: Stryker Spine SAS Z.1 Marticot -33610- Cestas France +33 (0) 5.57.97.06.30 http://www.stryker.com Distributed in the USA by Stryker Spine 2 Pearl Ct., Allendale NJ 07401-1677 USA +1-201-750-8000 Drill bits are reusable instruments, sterilized prior to use, used in conjunction with an Aviator Drill Guide to provide a pathway for screw insertion. The aviator Anterior Cervical Plating (ACP) System is intended for use as an aid in cervical spinal fusion and is intended for unilateral fixation.

Where it was distributed

Nationwide distribution.

Nationwide

Questions about this recall

Why was it recalled?

Aviator drill bits are 10 mm longer than the Reflex Hybrid drill bits and are not interchangeable which could lead to over-drilling by as much as 10 mm.

Which lots are affected?

E207315 Catalog number 48770610, 48770612, 48770616 All lots

Where was it sold?

Nationwide distribution.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

More recalls from Stryker Spine

All 29

Other Spinal Implants recalls

All 1,159