CLASS II Device recall · Reported 12 Feb 2020 · FDA Enforcement Report
Bander Ureteral Diversion Open-End Stent Set recalled by Cook
Cook is recalling Bander Ureteral Diversion Open-End Stent Set, Global Product No. G14822 G14823, distributed nationwide in the US. The recall was initiated on 4 Jan 2019 and the FDA has classified it as Class II.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-0972-2020
- FDA event ID
- 82026
- Recalling firm
- Cook, Bloomington, IN
- Classification
- Class II
- Status
- Terminated
- Recall initiated
- 4 Jan 2019
- FDA classified
- 4 Feb 2020
- Reported
- 12 Feb 2020
- Type
- Voluntary: Firm initiated
- Customers notified by
- Letter
- Reason category
- Other Reasons
- Device type
- Heart & Vascular Devices
Timeline
| 4 Jan 2019 | Recall initiated by company | |
| 4 Feb 2020 | Classified by FDA | +396 days |
| 12 Feb 2020 | Published in enforcement report | +8 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
The wire guide may be incorrectly loaded into the wire guide holder, leading to the stiff tip of the wire guide being introduced into the patient instead of the flexible tip.
Product as listed by the FDA
Bander Ureteral Diversion Open-End Stent Set, Global Product No. G14822 G14823
Where it was distributed
Domestic distribution to nationwide US. Foreign distribution to Australia, Belgium, Bolivia, Canada, Canary Islands, Chile, China, Colombia, Croatia, Czech Republic, Denmark, France, French Polynesia, Germany, Guatemala, Honduras, Hungary, India, Ireland, Italy, Japan, Korea, Luxembourg, Malaysia, Malta, Mexico, Mongolia, Netherlands, New Caledonia, New Zealand, Oman, Panama, Paraguay, Poland, Puerto Rico, Romania, Russian Federation, South Africa, Spain, Sweden, Switzerland, Taiwan, Trinidad and Tobago, United Arab Emirates, and the United Kingdom.
Questions about this recall
Is Bander Ureteral Diversion Open-End Stent Set, Global Product No. G14822 G14823 recalled?
Yes. Cook recalled Bander Ureteral Diversion Open-End Stent Set, Global Product No. G14822 G14823 on 4 Jan 2019. The recall is Class II and its status is Terminated. Recall number Z-0972-2020.
Why was it recalled?
The wire guide may be incorrectly loaded into the wire guide holder, leading to the stiff tip of the wire guide being introduced into the patient instead of the flexible tip.
Which lots are affected?
8011030 8033568 8063430 8063431 8150491 8417631 8417632 8587162 8587163 8590532 8590533 7990830 8003084 8033571 8063428 8590520 8590522 8629475 NS8629473
Where was it sold?
Domestic distribution to nationwide US. Foreign distribution to Australia, Belgium, Bolivia, Canada, Canary Islands, Chile, China, Colombia, Croatia, Czech Republic, Denmark, France, French Polynesia, Germany, Guatemala, Honduras, Hungary, India, Ireland, Italy, Japan, Korea, Luxembourg, Malaysia, Malta, Mexico, Mongolia, Netherlands, New Caledonia, New Zealand, Oman, Panama, Paraguay, Poland, Puerto Rico, Romania, Russian Federation, South Africa, Spain, Sweden, Switzerland, Taiwan, Trinidad and Tobago, United Arab Emirates, and the United Kingdom.
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class II mean?
Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.
Read the official FDA notice · Source record retrieved 21 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
More recalls from Cook
All 264| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Skinny Needle with Chiba Tip recalled by Cook | Cook | Other | Nationwide |
| Device | CLASS II | Flexor Check-Flo Introducer Ansel Modification recalled by Cook | Cook | Labeling Errors | 14 states |
| Device | CLASS II | Flexor Check-Flo Introducer Ansel Modification recalled by Cook | Cook | Labeling Errors | 14 states |
| Device | CLASS II | Transseptal Needle, Trocar recalled by Cook | Cook | Device Malfunction | Nationwide |
| Device | CLASS II | Transseptal Needle with Catheter recalled by Cook | Cook | Foreign Material | Nationwide |
Other Heart & Vascular Devices recalls
All 6,199| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Getinge PVC Thoracic Catheter, Soft PVC Right Angle recalled | Atrium Medical | Potency & Assay | Nationwide |
| Device | CLASS II | Products that contain the Pericare Wipes: *Bardex¿ I.C recalled | C.R. Bard | Other | Nationwide |
| Device | CLASS II | Products that contain the Pericare Wipes recalled by C.R. Bard | C.R. Bard | Other | Nationwide |
| Device | CLASS II | Products that contain the Pericare Wipes recalled by C.R. Bard | C.R. Bard | Other | Nationwide |
| Device | CLASS II | Getinge PVC Thoracic Catheter, Firm PVC Straight recalled | Atrium Medical | Potency & Assay | Nationwide |