CLASS I ONGOING Device recall · Reported 26 Nov 2025 · FDA Enforcement Report
BD Alaris infusion Pump Module recalled by CareFusion 303
CareFusion 303 is recalling BD Alaris infusion Pump Module 8100, distributed nationwide in the US. The recall was initiated on 17 Oct 2025 and the FDA has classified it as Class I.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-0430-2026
- FDA event ID
- 97669
- Recalling firm
- CareFusion 303, San Diego, CA
- Classification
- Class I
- Status
- Ongoing
- Recall initiated
- 17 Oct 2025
- FDA classified
- 20 Nov 2025
- Reported
- 26 Nov 2025
- Type
- Voluntary: Firm initiated
- Reason category
- Device Malfunction
- Device type
- Infusion Pumps
Timeline
| 17 Oct 2025 | Recall initiated by company | |
| 20 Nov 2025 | Classified by FDA | +34 days |
| 26 Nov 2025 | Published in enforcement report | +6 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
If infusion pump is dropped or severely jarred this may damage the pump module bezel assembly, which can cause under-infusion, over-infusion, unregulated flow, or pump module failure to calibrate; so dropped/jarred pumps should be removed from use, tested, and inspected by qualified service personnel prior to reuse; tip sheet, cleaning and disinfecting procedure and reference guide updated.
Device recalls are issued when a component, software or design fault could cause the device to fail or give wrong readings. Many are corrections applied in the field rather than physical returns.
Product as listed by the FDA
BD Alaris infusion Pump Module 8100, REF: 8100; KIT, BEZEL ASSY 8100BD, REF: 49000437; KIT ASSY ONE PIECE BEZEL 8100, REF: 49000270. Used with the following new and updated documents: BD Alaris Infusion System Inspection Safety Reminder, Part: BD-163124, Revision: 00; Cleaning and Disinfecting Procedure, Part: P00000864; Cleaning and Disinfecting Quick Reference Guide
Where it was distributed
Worldwide distribution. US nationwide including Puerto Rico, Guam, American Samoa, Northern Mariana Islands; and countries of SG, BH, AE, JP, CA, IT, QA, PH, DE, TW, PK, KR, BW, SA, MY, IL, NZ, AU, IN, GB, CN, NL, FR, ZA, BE, MX, CH, GI, KW, and TR.
Questions about this recall
Why was it recalled?
If infusion pump is dropped or severely jarred this may damage the pump module bezel assembly, which can cause under-infusion, over-infusion, unregulated flow, or pump module failure to calibrate; so dropped/jarred pumps should be removed from use, tested, and inspected by qualified service personnel prior to reuse; tip sheet, cleaning and disinfecting procedure and reference guide updated.
Which lots are affected?
All Alaris Pump Modules distributed after June 19, 2004. REF/UDI-DI: 8100/10885403810015, 10885403810046, 10885403810039, 10885403222054, 10885403517723, 10885403213212, 10885403211423; 49000437/10885403471018; 49000270/10885403247743
Where was it sold?
Worldwide distribution. US nationwide including Puerto Rico, Guam, American Samoa, Northern Mariana Islands; and countries of SG, BH, AE, JP, CA, IT, QA, PH, DE, TW, PK, KR, BW, SA, MY, IL, NZ, AU, IN, GB, CN, NL, FR, ZA, BE, MX, CH, GI, KW, and TR.
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class I mean?
Class I is the most serious FDA recall category: there is a reasonable probability that using the product will cause serious health consequences or death.
Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
More recalls from CareFusion 303
All 143| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | BD Pyxis MedStation ES Tower recalled by CareFusion 303 | CareFusion 303 | Temperature Abuse | Nationwide |
| Device | CLASS II | BD Pyxis MedBank Mini recalled by CareFusion 303 | CareFusion 303 | Temperature Abuse | Nationwide |
| Device | CLASS II | BD Pyxis MedStation recalled by CareFusion 303 | CareFusion 303 | Temperature Abuse | Nationwide |
| Device | CLASS II | BD Pyxis Anesthesia Station ES recalled by CareFusion 303 | CareFusion 303 | Temperature Abuse | Nationwide |
| Device | CLASS II | BD Pyxis MedStation ES recalled by CareFusion 303 | CareFusion 303 | Temperature Abuse | Nationwide |
Other Infusion Pumps recalls
All 549| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS I | BD Alaris Pump Infusion Sets 10561554* recalled by BD Switzerland Sarl | BD Switzerland Sarl | Other | Nationwide |
| Device | CLASS II | IV Non-Dedicated Gravity Administration Sets with drip chambers recalled | ICU Medical | Foreign Material | Nationwide |
| Device | CLASS II | Plum administration sets recalled by ICU Medical | ICU Medical | Particulate Matter | Nationwide |
| Device | CLASS II | Baxter Novum IQ Syringe Pump, Loaner recalled by Baxter Healthcare | Baxter Healthcare | Device Malfunction | Nationwide |
| Device | CLASS II | Baxter Novum IQ Syringe Pump recalled by Baxter Healthcare | Baxter Healthcare | Device Malfunction | Nationwide |