CLASS II ONGOING Device recall · Reported 17 Jan 2024 · FDA Enforcement Report

BD Recanalization System recalled by Bard Peripheral Vascular

Bard Peripheral Vascular is recalling BD Recanalization System, distributed in 34 states. The recall was initiated on 14 Nov 2023 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesUDI-DI: 00801741202421. Serial Numbers: BSX000001, BSX000002, BSX000003, BSX000004, BSX000005, BSX000006, BSX000007, BSX000008, BSX000009, BSX000010, BSX000011, BSX000012, BSX000013, BSX000014, BSX000015, BSX000016, BSX000017, BSX000018, BSX000019, BSX000020, BSX000021, BSX000022, BSX000023, BSX000024, BSX000025, BSX000026, BSX000027, BSX000028, BSX000029, BSX000030, BSX000031, BSX000032, BSX000033, BSX000034, BSX000035, BSX000036, BSX000037, BSX000038, BSX000039, BSX000040, BSX000041, BSX000042, BSX000043, BSX000044, BSX000045, BSX000046, BSX000047, BSX000048, BSX000049, BSX000050, BSX000051, BSX000052, BSX000053, BSX000054, BSX000055, BSX000056, BSX000057, BSX000058, BSX000059, BSX000060, BSX000061, BSX000062, BSX000063, BSX000064, BSX000065, BSX000066, BSX000067, BSX000068, BSX000109, BSX000110, BSX000111, BSX000112, BSX000113, BSX000114, BSX000115, BSX000116, BSX000117, BSX000118, BSX000119, BSX000120, BSX000121, BSX000122, BSX000123, BSX000124, BSX000125, BSX000126, BSX000127, BSX000128, BSX000129, BSX000130, BSX000131, BSX000132, BSX000133, BSX000134, BSX000135, BSX000136, BSX000137, BSX000138, BSX000139, BSX000140, BSX000141, BSX000142, BSX000143, BSX000144, BSX000145, BSX000146, BSX000147, BSX000148, BSX000149, BSX000150, BSX000151, BSX000152, BSX000153, BSX000154, BSX000155, BSX000156, BSX000157, BSX000158, BSX000159, BSX000160, BSX000161, BSX000162, BSX000163, BSX000164, BSX000165, BSX000166, BSX000167, BSX000168, BSX000169, BSX000170, BSX000171, BSX000172, BSX000173, BSX000174, BSX000175, BSX000176, BSX000177, BSX000178, BSX000179, BSX000180, BSX000181, BSX000182, BSX000183, BSX000184, BSX000185, BSX000186, BSX000187, BSX000188, BSX000189, BSX000190, BSX000191, BSX000192, BSX000193, BSX000194, BSX000195, BSX000196, BSX000197, BSX000198, BSX000199, BSX000200, BSX000201, BSX000202, BSX000203, BSX000204, BSX000205, BSX000206, BSX000207, BSX000208, BSX000209, BSX000210, BSX000211, BSX000212, BSX000213, BSX000214, BSX000215, BSX000216, BSX000217, BSX000218, BSX000219, BSX000220, BSX000221, BSX000222, BSX000223, BSX000224, BSX000225, BSX000226, BSX000227, BSX000228, BSX000229, BSX000230, BSX000231, BSX000232, BSX000233, BSX000234, BSX000235, BSX000236, BSX000237, BSX000238, BSX000239, BSX000240, BSX000241, BSX000242, BSX000243, BSX000244, BSX000245, BSX000246, BSX000247, BSX000248, BSX000249, BSX000250, BSX000251, BSX000252, BSX000253, BSX000254, BSX000255, BSX000256, BSX000257, BSX000258, BSX000259, BSX000260, BSX000261, BSX000262, BSX000263, BSX000264, BSX000265, BSX000266, BSX000267, BSX000268, BSX000269, BSX000270, BSX000271, BSX000272, BSX000273, BSX000274, BSX000275, BSX000276, BSX000277, BSX000278, BSX000279, BSX000280, BSX000281, BSX000282, BSX000283, BSX000284, BSX000285, BSX000286, BSX000287, BSX000288, BSX000289, BSX000290, BSX000291, BSX000292, BSX000293, BSX000294, BSX000295, BSX000296, BSX000297, BSX000298, BSX000299, BSX000300, BSX000301, BSX000302, BSX000303, BSX000304, BSX000305, BSX000306, BSX000307, BSX000308, BSX000309, BSX000310, BSX000311, BSX000312, BSX000313, BSX000314, BSX000315, BSX000316, BSX000317, BSX000318, BSX000319, BSX000320, BSX000321, BSX000322, BSX000323, BSX000324, BSX000325, BSX000326, BSX000327, BSX000328, BSX000330, BSX000331, BSX000332, BSX000333, BSX000334, BSX000335, BSX000336, BSX000337, BSX000338, BSX000339, BSX000340, BSX000341, BSX000342, BSX000343, BSX000344, BSX000345, BSX000346, BSX000347, BSX000348, BSX000349, BSX000350, BSX000351, BSX000352, BSX000353, BSX000354, BSX000355, BSX000356, BSX000357, BSX000358, BSX000359, BSX000360, BSX000361, BSX000362, BSX000363, BSX000364, BSX000365, BSX000366, BSX000367, BSX000368, BSX000369, BSX000370, BSX000371, BSX000372, BSX000373, BSX000374, BSX000375, BSX000376, BSX000377, BSX000378, BSX000379, BSX000380, BSX000381, BSX000382, BSX000383, BSX000384, BSX000385, BSX000386, BSX000387, BSX000388, BSX000389, BSX000390, BSX000391, BSX000392, BSX000393, BSX000394, BSX000395, BSX000396, BSX000397, BSX000398, BSX000399, BSX000400, BSX000401, BSX000402, BSX000403, BSX000404, BSX000405, BSX000406, BSX000407, BSX000408, BSX000409, BSX000410, BSX000411, BSX000412, BSX000413, BSX000414, BSX000415, BSX000416, BSX000417, BSX000418, BSX000419, BSX000420, BSX000421, BSX000422, BSX000423, BSX000424, BSX000425, BSX000426, BSX000427, BSX000428, BSX000429, BSX000430, BSX000431, BSX000432, BSX000433, BSX000434, BSX000435, BSX000436, BSX000437, BSX000438, BSX000439, BSX000440, BSX000441, BSX000442, BSX000443, BSX000444
Quantity recalled403

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-0676-2024
FDA event ID
93607
Recalling firm
Bard Peripheral Vascular, Tempe, AZ
Classification
Class II
Status
Ongoing
Recall initiated
14 Nov 2023
FDA classified
5 Jan 2024
Reported
17 Jan 2024
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Device Malfunction

Timeline

14 Nov 2023Recall initiated by company
5 Jan 2024Classified by FDA+52 days
17 Jan 2024Published in enforcement report+12 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

Recanalization systems may be in a manufacturing mode that allows the system to continuously run without the typical 30 second pauses, and 4-minute warning/5-minute restart required messages to indicate the ten-minute maximum catheter use time has been exceeded, which may increase the risk of the device overheating or breaking.

Device recalls are issued when a component, software or design fault could cause the device to fail or give wrong readings. Many are corrections applied in the field rather than physical returns.

Product as listed by the FDA

BD Recanalization System, REF: BDRECANSYSTEM

Where it was distributed

US: CA, OH, PA, NY, TX, NJ, MS, IL, GA, AL, AZ, NC, NV, PR, IN, MA, WI, KS, AR, FL, MD, IA, NE, LA, HI, MO, DE, DC, VA, NM, CO, KY, ME, MN

AlabamaArkansasArizonaCaliforniaColoradoDistrict of ColumbiaDelawareFloridaGeorgiaHawaiiIowaIllinoisIndianaKansasKentuckyLouisianaMassachusettsMarylandMaineMinnesota

Questions about this recall

Is BD Recanalization System recalled?

Yes. Bard Peripheral Vascular recalled BD Recanalization System on 14 Nov 2023. The recall is Class II and its status is Ongoing. Recall number Z-0676-2024.

Why was it recalled?

Recanalization systems may be in a manufacturing mode that allows the system to continuously run without the typical 30 second pauses, and 4-minute warning/5-minute restart required messages to indicate the ten-minute maximum catheter use time has been exceeded, which may increase the risk of the device overheating or breaking.

Which lots are affected?

UDI-DI: 00801741202421. Serial Numbers: BSX000001, BSX000002, BSX000003, BSX000004, BSX000005, BSX000006, BSX000007, BSX000008, BSX000009, BSX000010, BSX000011, BSX000012, BSX000013, BSX000014, BSX000015, BSX000016, BSX000017, BSX000018, BSX000019, BSX000020, BSX000021, BSX000022, BSX000023, BSX000024, BSX000025, BSX000026, BSX000027, BSX000028, BSX000029, BSX000030, BSX000031, BSX000032, BSX000033, BSX000034, BSX000035, BSX000036, BSX000037, BSX000038, BSX000039, BSX000040, BSX000041, BSX000042, BSX000043, BSX000044, BSX000045, BSX000046, BSX000047, BSX000048, BSX000049, BSX000050, BSX000051, BSX000052, BSX000053, BSX000054, BSX000055, BSX000056, BSX000057, BSX000058, BSX000059, BSX000060, BSX000061, BSX000062, BSX000063, BSX000064, BSX000065, BSX000066, BSX000067, BSX000068, BSX000109, BSX000110, BSX000111, BSX000112, BSX000113, BSX000114, BSX000115, BSX000116, BSX000117, BSX000118, BSX000119, BSX000120, BSX000121, BSX000122, BSX000123, BSX000124, BSX000125, BSX000126, BSX000127, BSX000128, BSX000129, BSX000130, BSX000131, BSX000132, BSX000133, BSX000134, BSX000135, BSX000136, BSX000137, BSX000138, BSX000139, BSX000140, BSX000141, BSX000142, BSX000143, BSX000144, BSX000145, BSX000146, BSX000147, BSX000148, BSX000149, BSX000150, BSX000151, BSX000152, BSX000153, BSX000154, BSX000155, BSX000156, BSX000157, BSX000158, BSX000159, BSX000160, BSX000161, BSX000162, BSX000163, BSX000164, BSX000165, BSX000166, BSX000167, BSX000168, BSX000169, BSX000170, BSX000171, BSX000172, BSX000173, BSX000174, BSX000175, BSX000176, BSX000177, BSX000178, BSX000179, BSX000180, BSX000181, BSX000182, BSX000183, BSX000184, BSX000185, BSX000186, BSX000187, BSX000188, BSX000189, BSX000190, BSX000191, BSX000192, BSX000193, BSX000194, BSX000195, BSX000196, BSX000197, BSX000198, BSX000199, BSX000200, BSX000201, BSX000202, BSX000203, BSX000204, BSX000205, BSX000206, BSX000207, BSX000208, BSX000209, BSX000210, BSX000211, BSX000212, BSX000213, BSX000214, BSX000215, BSX000216, BSX000217, BSX000218, BSX000219, BSX000220, BSX000221, BSX000222, BSX000223, BSX000224, BSX000225, BSX000226, BSX000227, BSX000228, BSX000229, BSX000230, BSX000231, BSX000232, BSX000233, BSX000234, BSX000235, BSX000236, BSX000237, BSX000238, BSX000239, BSX000240, BSX000241, BSX000242, BSX000243, BSX000244, BSX000245, BSX000246, BSX000247, BSX000248, BSX000249, BSX000250, BSX000251, BSX000252, BSX000253, BSX000254, BSX000255, BSX000256, BSX000257, BSX000258, BSX000259, BSX000260, BSX000261, BSX000262, BSX000263, BSX000264, BSX000265, BSX000266, BSX000267, BSX000268, BSX000269, BSX000270, BSX000271, BSX000272, BSX000273, BSX000274, BSX000275, BSX000276, BSX000277, BSX000278, BSX000279, BSX000280, BSX000281, BSX000282, BSX000283, BSX000284, BSX000285, BSX000286, BSX000287, BSX000288, BSX000289, BSX000290, BSX000291, BSX000292, BSX000293, BSX000294, BSX000295, BSX000296, BSX000297, BSX000298, BSX000299, BSX000300, BSX000301, BSX000302, BSX000303, BSX000304, BSX000305, BSX000306, BSX000307, BSX000308, BSX000309, BSX000310, BSX000311, BSX000312, BSX000313, BSX000314, BSX000315, BSX000316, BSX000317, BSX000318, BSX000319, BSX000320, BSX000321, BSX000322, BSX000323, BSX000324, BSX000325, BSX000326, BSX000327, BSX000328, BSX000330, BSX000331, BSX000332, BSX000333, BSX000334, BSX000335, BSX000336, BSX000337, BSX000338, BSX000339, BSX000340, BSX000341, BSX000342, BSX000343, BSX000344, BSX000345, BSX000346, BSX000347, BSX000348, BSX000349, BSX000350, BSX000351, BSX000352, BSX000353, BSX000354, BSX000355, BSX000356, BSX000357, BSX000358, BSX000359, BSX000360, BSX000361, BSX000362, BSX000363, BSX000364, BSX000365, BSX000366, BSX000367, BSX000368, BSX000369, BSX000370, BSX000371, BSX000372, BSX000373, BSX000374, BSX000375, BSX000376, BSX000377, BSX000378, BSX000379, BSX000380, BSX000381, BSX000382, BSX000383, BSX000384, BSX000385, BSX000386, BSX000387, BSX000388, BSX000389, BSX000390, BSX000391, BSX000392, BSX000393, BSX000394, BSX000395, BSX000396, BSX000397, BSX000398, BSX000399, BSX000400, BSX000401, BSX000402, BSX000403, BSX000404, BSX000405, BSX000406, BSX000407, BSX000408, BSX000409, BSX000410, BSX000411, BSX000412, BSX000413, BSX000414, BSX000415, BSX000416, BSX000417, BSX000418, BSX000419, BSX000420, BSX000421, BSX000422, BSX000423, BSX000424, BSX000425, BSX000426, BSX000427, BSX000428, BSX000429, BSX000430, BSX000431, BSX000432, BSX000433, BSX000434, BSX000435, BSX000436, BSX000437, BSX000438, BSX000439, BSX000440, BSX000441, BSX000442, BSX000443, BSX000444

Where was it sold?

US: CA, OH, PA, NY, TX, NJ, MS, IL, GA, AL, AZ, NC, NV, PR, IN, MA, WI, KS, AR, FL, MD, IA, NE, LA, HI, MO, DE, DC, VA, NM, CO, KY, ME, MN

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 21 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

More recalls from Bard Peripheral Vascular

All 138

Other Other Medical Devices recalls

All 13,778