CLASS II Device recall · Reported 14 Jun 2017 · FDA Enforcement Report

Berchtold Chromophare F300 Exam Light Chromophare recalled

Stryker Communications is recalling Berchtold Chromophare F300 Exam Light Chromophare examination lights are medical lights. The recall was initiated on 21 Apr 2017 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesSerial Numbers: 7735170-X68377, 7735170-X68397, 7735170-X14846, 7735170-X15688, 7735170-X14492, 7735170-X14495, 7735170-X14497, 7735170-X14501, 7735170-X14502, 7735170-X14504, 7735170-X14506, 7735170-X14507, 7735170-X14508, 7735170-X14510, 7735170-X68383, 7735170-X68384, 7735170-X68385, 7735170-X68386, 7735170-X68393, 7735170-X68394, 7735170-X68395, 7735170-X68396, 7735170-X68392, 7735170-X68400, 7735170-X68401, 7735170-X15607, 7735170-X15608, 7735170-X15609, 7735170-X15610, 7735170-X15611, 7735170-X15612, 7735170-X15613, 7735170-X15614, 7735170-X15615
Quantity recalled57 units

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-2248-2017
FDA event ID
77261
Recalling firm
Stryker Communications, Flower Mound, TX
Classification
Class II
Status
Terminated
Recall initiated
21 Apr 2017
FDA classified
5 Jun 2017
Reported
14 Jun 2017
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Other Reasons

Timeline

21 Apr 2017Recall initiated by company
5 Jun 2017Classified by FDA+45 days
14 Jun 2017Published in enforcement report+9 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

The exam light may have a tolerance issue with the adapter assembly, which could potentially not allow the snap ring to be seated correctly. If this is the case, there could be insufficient mount force that may cause the equipment to fall, resulting in serious injury.

Product as listed by the FDA

Berchtold Chromophare F300 Exam Light CHROMOPHARE examination lights are medical lights used for local illumination of the patients' body so that illnesses, injuries and disabilities can be diagnosed and treated

Where it was distributed

Worldwide Distribution - US and Canada

Questions about this recall

Why was it recalled?

The exam light may have a tolerance issue with the adapter assembly, which could potentially not allow the snap ring to be seated correctly. If this is the case, there could be insufficient mount force that may cause the equipment to fall, resulting in serious injury.

Which lots are affected?

Serial Numbers: 7735170-X68377, 7735170-X68397, 7735170-X14846, 7735170-X15688, 7735170-X14492, 7735170-X14495, 7735170-X14497, 7735170-X14501, 7735170-X14502, 7735170-X14504, 7735170-X14506, 7735170-X14507, 7735170-X14508, 7735170-X14510, 7735170-X68383, 7735170-X68384, 7735170-X68385, 7735170-X68386, 7735170-X68393, 7735170-X68394, 7735170-X68395, 7735170-X68396, 7735170-X68392, 7735170-X68400, 7735170-X68401, 7735170-X15607, 7735170-X15608, 7735170-X15609, 7735170-X15610, 7735170-X15611, 7735170-X15612, 7735170-X15613, 7735170-X15614, 7735170-X15615

Where was it sold?

Worldwide Distribution - US and Canada

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

More recalls from Stryker Communications

All 20

Other Other Medical Devices recalls

All 13,778