CLASS II ONGOING Device recall · Reported 15 Mar 2023 · FDA Enforcement Report

Biophen Protein C 5 kit recalled by Aniara Diagnostica

Aniara Diagnostica is recalling Biophen Protein C 5 kit is for in vitro quantitative determination of Protein C activity, distributed nationwide in the US. The recall was initiated on 1 Feb 2023 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesUDI-DI: 366353700063BD Lot FA083327 (And all future lots)
Quantity recalled2 units

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-1227-2023
FDA event ID
91618
Recalling firm
Aniara Diagnostica, West Chester, OH
Classification
Class II
Status
Ongoing
Recall initiated
1 Feb 2023
FDA classified
6 Mar 2023
Reported
15 Mar 2023
Type
Voluntary: Firm initiated
Customers notified by
E-Mail
Reason category
Other Reasons

Timeline

1 Feb 2023Recall initiated by company
6 Mar 2023Classified by FDA+33 days
15 Mar 2023Published in enforcement report+9 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

New warnings and precautionary statements on product labelling: Reagent 2 (R2) of BIOPHEN Protein C 5 contains the chemical substance - Cesium Chloride which changed classification and has become hazardous according to the CLP regulation ((EC) No 1272/2008) is suspected of damaging fertility

Product as listed by the FDA

BIOPHEN Protein C 5 kit is for in vitro quantitative determination of Protein C activity on human citrated plasma. Protein C is a glycoprotein, vitamin K dependent, which inhibits coagulation. Protein C in plasma is measured following a specific activation with Protac (Agkistrodom Contortrix), an enzyme extracted from snake venom. Activated protein C hydrolysis the chromogenic substrate (SaPC-21) which release para-nitroaniline (pNA). The amount of pNA released (measured by absorbance at 405 nm) is directly proportional to the concentration of Protein C in the specimen.

Where it was distributed

US Nationwide distribution in the state of NC.

Nationwide North Carolina

Questions about this recall

Why was it recalled?

New warnings and precautionary statements on product labelling: Reagent 2 (R2) of BIOPHEN Protein C 5 contains the chemical substance - Cesium Chloride which changed classification and has become hazardous according to the CLP regulation ((EC) No 1272/2008) is suspected of damaging fertility

Which lots are affected?

UDI-DI: 366353700063BD Lot FA083327 (And all future lots)

Where was it sold?

US Nationwide distribution in the state of NC.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

More recalls from Aniara Diagnostica

All 11

Other Test Kits & Diagnostics recalls

All 2,663