CLASS II ONGOING Device recall · Reported 9 Apr 2025 · FDA Enforcement Report
Boppli: Bedside Device Kit recalled by Pyrames
Pyrames is recalling Boppli: Bedside Device Kit, distributed in California, Massachusetts. The recall was initiated on 24 Feb 2025 and the FDA has classified it as Class II.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-1508-2025
- FDA event ID
- 96414
- Recalling firm
- Pyrames, Cupertino, CA
- Classification
- Class II
- Status
- Ongoing
- Recall initiated
- 24 Feb 2025
- FDA classified
- 2 Apr 2025
- Reported
- 9 Apr 2025
- Type
- Voluntary: Firm initiated
- Customers notified by
- Visit
- Reason category
- Device Malfunction
- Device type
- Other Medical Devices
Timeline
| 24 Feb 2025 | Recall initiated by company | |
| 2 Apr 2025 | Classified by FDA | +37 days |
| 9 Apr 2025 | Published in enforcement report | +7 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
Single-use battery-powered blood pressure monitor, for use on neonates and infants, due to vibratory noise, may not adequately detect changes in blood pressure per specification. Containment actions: 1) Ventilator lines not to touch the isolette (e.g. mattress, side rails, pillow) or patient, 2) Move affected monitor to patient's foot.
Device recalls are issued when a component, software or design fault could cause the device to fail or give wrong readings. Many are corrections applied in the field rather than physical returns.
Product as listed by the FDA
Boppli: Bedside Device Kit, REF: 54-0002; Sensor Band, REF: 51-0007
Where it was distributed
US distribution to states of: MA and CA
Questions about this recall
Why was it recalled?
Single-use battery-powered blood pressure monitor, for use on neonates and infants, due to vibratory noise, may not adequately detect changes in blood pressure per specification. Containment actions: 1) Ventilator lines not to touch the isolette (e.g. mattress, side rails, pillow) or patient, 2) Move affected monitor to patient's foot.
Which lots are affected?
Software version: 2.2.3 Part/UFI-DI/Lot: 54-0002/10860007282925/24020902, 24040302, 23103002; 51-0007/10860007282901/24011702, 24111301
Where was it sold?
US distribution to states of: MA and CA
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class II mean?
Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.
Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
Other Other Medical Devices recalls
All 13,986| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | ANNOUNCED | Luna Bilevel Positive Airway Pressure (BPAP) System recalledLuna | React Health | Other | VT |
| Device | ANNOUNCED | Beds, Rails, Mattresses, and accessories recalled by MedlineMedline | Medline | Other | |
| Device | CLASS II | Hemochron Jr., Whole Blood Microcoagulation System recalled | Accriva Diagnostics | Other | Nationwide |
| Device | CLASS II | Senographe Pristina 3D recalled by GE Medical Systems | GE Medical Systems | Other | WI |
| Device | CLASS II | Mobile column TruSystem recalled by Baxter Healthcare | Baxter Healthcare | Device Malfunction | Nationwide |