CLASS II ONGOING Device recall · Reported 11 Nov 2026 · FDA Enforcement Report
Hemochron Jr., Whole Blood Microcoagulation System recalled
Accriva Diagnostics is recalling Hemochron Jr., Whole Blood Microcoagulation System, Activated Partial Thromboplastin Time, distributed nationwide in the US. The recall was initiated on 14 Sep 2026 and the FDA has classified it as Class II.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-0317-2027
- FDA event ID
- 99863
- Recalling firm
- Accriva Diagnostics, San Diego, CA
- Classification
- Class II
- Status
- Ongoing
- Recall initiated
- 14 Sep 2026
- FDA classified
- 30 Oct 2026
- Reported
- 11 Nov 2026
- Type
- Voluntary: Firm initiated
- Customers notified by
- Letter
- Reason category
- Other Reasons
- Device type
- Other Medical Devices
Timeline
| 14 Sep 2026 | Recall initiated by company | |
| 30 Oct 2026 | Classified by FDA | +46 days |
| 11 Nov 2026 | Published in enforcement report | +12 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
Microcoagulation test used to evaluate low dose heparin anticoagulation has a mispositioned or absent clot activating reagent, which may result in out-of-range high error code, or erroneously high results, which may lead retest, or overestimation of patient's anticoagulation, leading to inadequate anticoagulant therapy, which could result in thrombosis.
Product as listed by the FDA
Hemochron Jr., Whole Blood Microcoagulation System, Activated Partial Thromboplastin Time (APTT) Test Cuvette, REF: J103
Where it was distributed
Worldwide - US Nationwide distribution in the states of MA, MO, AK, WV, NJ, TX, VA, IL, NY and the countries of China, Colombia.
Nationwide AlaskaIllinoisMassachusettsMissouriNew JerseyNew YorkTexasVirginiaWest Virginia
Questions about this recall
Why was it recalled?
Microcoagulation test used to evaluate low dose heparin anticoagulation has a mispositioned or absent clot activating reagent, which may result in out-of-range high error code, or erroneously high results, which may lead retest, or overestimation of patient's anticoagulation, leading to inadequate anticoagulant therapy, which could result in thrombosis.
Which lots are affected?
UDI-DI: 10711234103012, Lot(Expiration): B6JCA005(02/28/2027), C6JCA008(03/31/2027), D6JCA011(04/30/2027), K5JCA041(10/31/2026), C6JCA007(03/31/2027)
Where was it sold?
Worldwide - US Nationwide distribution in the states of MA, MO, AK, WV, NJ, TX, VA, IL, NY and the countries of China, Colombia.
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class II mean?
Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.
Read the official FDA notice · Source record retrieved 17 Nov 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
More recalls from Accriva Diagnostics
All 7| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | directCHECK Act-lr whole blood quality control, level recalled | Accriva Diagnostics | Potency & Assay | Nationwide |
| Device | CLASS II | Hemochron Activated Clotting Time Low-Range Test Cuvette recalled | Accriva Diagnostics | Foreign Material | Nationwide |
| Device | CLASS III | VerifyNow PRUTest Platelet Reactivity Test recalled | Accriva Diagnostics | Packaging Defects | Nationwide |
| Device | CLASS II | VerifyNow P2Y12 Platelet Reactivity Test recalled by Accriva Diagnostics | Accriva Diagnostics | Other | Nationwide |
| Device | CLASS II | AVOXimeter 1000E, Product Number: AVOX1000E recalled | Accriva Diagnostics | Device Malfunction | AZ |
Other Other Medical Devices recalls
All 13,985| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | ANNOUNCED | Beds, Rails, Mattresses, and accessories recalled by MedlineMedline | Medline | Other | |
| Device | CLASS II | Senographe Pristina 3D recalled by GE Medical Systems | GE Medical Systems | Other | WI |
| Device | CLASS II | Mobile column TruSystem recalled by Baxter Healthcare | Baxter Healthcare | Device Malfunction | Nationwide |
| Device | CLASS II | Halyard Minor Pack recalled by Avid Medical | Avid Medical | Sterility | Nationwide |
| Device | CLASS II | Dragen Server for NextSeq 550Dx recalled by Illumina | Illumina | Other | Nationwide |