CLASS II ONGOING Device recall · Reported 11 Nov 2026 · FDA Enforcement Report

Mobile column TruSystem recalled by Baxter Healthcare

Baxter Healthcare is recalling Mobile column TruSystem 7500 U, distributed nationwide in the US. The recall was initiated on 14 Sep 2026 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesUDI-DI: 00887761974388; Serial number: 110183858
Quantity recalled1 unit

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-0333-2027
FDA event ID
99826
Recalling firm
Baxter Healthcare, Round Lake, IL
Classification
Class II
Status
Ongoing
Recall initiated
14 Sep 2026
FDA classified
3 Nov 2026
Reported
11 Nov 2026
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Device Malfunction

Timeline

14 Sep 2026Recall initiated by company
3 Nov 2026Classified by FDA+50 days
11 Nov 2026Published in enforcement report+8 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

The main lift may unexpectedly drop from its set position to its lowest position during use. The tabletop may not be able to be raised again after the drop occurs. An uncontrolled drop of the surgical table may result in an abrupt patient displacement during surgery. Serious or critical adverse health consequences may occur, including traumatic injury, airway compromise, and surgical procedure complications, among others.

Device recalls are issued when a component, software or design fault could cause the device to fail or give wrong readings. Many are corrections applied in the field rather than physical returns.

Product as listed by the FDA

Mobile column TruSystem 7500 U, Product Code 1730720

Where it was distributed

US Nationwide distribution in the states of IL, MO, TX.

Nationwide IllinoisMissouriTexas

Questions about this recall

Why was it recalled?

The main lift may unexpectedly drop from its set position to its lowest position during use. The tabletop may not be able to be raised again after the drop occurs. An uncontrolled drop of the surgical table may result in an abrupt patient displacement during surgery. Serious or critical adverse health consequences may occur, including traumatic injury, airway compromise, and surgical procedure complications, among others.

Which lots are affected?

UDI-DI: 00887761974388; Serial number: 110183858

Where was it sold?

US Nationwide distribution in the states of IL, MO, TX.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 17 Nov 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

Other products in this recall event

1 other product · FDA event 99826

This product is part of a recall event; the main page for the event is TruSystem 7000 U Operating Table (dV) recalled by Baxter Healthcare.

More recalls from Baxter Healthcare

All 585

Other Other Medical Devices recalls

All 13,985