CLASS II Device recall · Reported 3 Feb 2021 · FDA Enforcement Report
Brain Heart Infusion Agar with 6 ug/ml Vancomycin recalled
Hardy Diagnostics is recalling Brain Heart Infusion Agar with 6 ug/ml Vancomycin, distributed in 11 states. The recall was initiated on 21 Dec 2020 and the FDA has classified it as Class II.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-0919-2021
- FDA event ID
- 87091
- Recalling firm
- Hardy Diagnostics, Santa Maria, CA
- Classification
- Class II
- Status
- Terminated
- Recall initiated
- 21 Dec 2020
- FDA classified
- 25 Jan 2021
- Reported
- 3 Feb 2021
- Type
- Voluntary: Firm initiated
- Customers notified by
- Letter
- Reason category
- Other Bacterial Contamination
- Device type
- Other Medical Devices
Timeline
| 21 Dec 2020 | Recall initiated by company | |
| 25 Jan 2021 | Classified by FDA | +35 days |
| 3 Feb 2021 | Published in enforcement report | +9 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
Complaints were received reporting of Staphylococcus aureus (ATCC 29213) breakthrough when the strain should have been inhibited.
Product as listed by the FDA
Brain Heart Infusion Agar with 6 ug/ml Vancomycin Cat no: G14 Lot no: 472698 Manufactured: 2020-10-29 Expires: 2020-12-28 Container Type: 15x100mm plate, 18 ml
Where it was distributed
U.S.: CA, UT, MI, Puerto Rico, IN, CT, AZ, GA, MA, TN, WA O.U.S. None
ArizonaCaliforniaConnecticutGeorgiaIndianaMassachusettsMichiganPuerto RicoTennesseeUtahWashington
Questions about this recall
Why was it recalled?
Complaints were received reporting of Staphylococcus aureus (ATCC 29213) breakthrough when the strain should have been inhibited.
Which lots are affected?
Catalog Number: G14 UDI Code: 00816576020102 Lot Number: 472698
Where was it sold?
U.S.: CA, UT, MI, Puerto Rico, IN, CT, AZ, GA, MA, TN, WA O.U.S. None
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class II mean?
Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.
Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
More recalls from Hardy Diagnostics
All 20| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | HardyDisk AST Cefiderocol 30¿g (FDC30) recalled by Hardy Diagnostics | Hardy Diagnostics | Other | Nationwide |
| Device | CLASS II | HardyCHROM CRE, selective and differential culture medium recalled | Hardy Diagnostics | Device Malfunction | 22 states |
| Device | CLASS II | Quickslide GramPro 1 Automated Gram Stanier is a microscope recalled | Hardy Diagnostics | Device Malfunction | Nationwide |
| Device | CLASS II | Viral Transport Medium recalled by Hardy Diagnostics | Hardy Diagnostics | Foreign Material | Nationwide |
| Device | CLASS III | Strep B Carrot Broth One-Step recalled by Hardy Diagnostics | Hardy Diagnostics | Other | 21 states |
Other Other Medical Devices recalls
All 13,908| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Plastics Tray kits recalled by Avid Medical | Avid Medical | Foreign Material | Nationwide |
| Device | CLASS II | Heart Prep Tray kits recalled by Avid Medical | Avid Medical | Other | Nationwide |
| Device | CLASS II | Pack Open Heart Cfh kits recalled by Avid Medical | Avid Medical | Other | Nationwide |
| Device | CLASS II | Sperm Ab Elisa (96 well plate test) recalled by DRG International | DRG International | Other | Nationwide |
| Device | CLASS III | BeneView Series recalled by Mindray DS USA | Mindray DS USA | Device Malfunction | Nationwide |